Lead isotope ratio analysis of bullet samples by using quadrupole ICP-MS

被引:3
|
作者
Tamura, Shu-ichi
Hokura, Akiko
Oishi, Masahiro
Nakai, Izumi
机构
[1] Tokyo Univ Sci, Fac Sci, Dept Appl Chem, Shinjuku Ku, Tokyo 1628601, Japan
[2] TDK Co, Ctr Mat Anal, Ichikawa, Chiba 2728558, Japan
关键词
bullet samples; lead Isotope ratio analysis; forensic identification; quadrupole ICP-MS;
D O I
10.2116/bunsekikagaku.55.827
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The measurement conditions for the precise analysis of the lead stable isotope ratio by using an ICP-MS equipped with a quadrupole mass spectrometer were studied in order to apply the technique to the forensic identification of bullet samples. The values of the relative standard deviation obtained for the ratio of Pb-208/Ph-206, Pb-207/Pb-206 and Pb-204/Pb-206 were lower than 0.2% after optimization of the analytical conditions, including the optimum lead concentration of the sample solution to be about 70 ppb and an integration time for I m/z of 15 s. This method was applied to an analysis of lead in bullets for rifles and handguns; a stable isotope ratio of lead was found to be suitable for the identification of bullets. This study has demonstrated that the lead isotope ratio measured by using a quadrupole ICP-MS was useful for a practical analysis of bullet samples in forensic science.
引用
收藏
页码:827 / 834
页数:8
相关论文
共 50 条
  • [41] CAIS correction for blood matrix effect on determination of lead concentration and isotope ratio by ICP-MS
    Zeng Jing
    Wang Xiao-yan
    Liu Hu-sheng
    Dun Zhe
    Zhai Lei
    Wang Jing-yu
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2007, 27 (02) : 374 - 377
  • [42] RHENIUM-OSMIUM ISOTOPE RATIO DETERMINATION BY ICP-MS
    BENETEAU, SB
    RICHARDSON, JM
    ATOMIC SPECTROSCOPY, 1992, 13 (04) : 118 - 122
  • [43] ACCURATE DETERMINATION OF LEAD BY ISOTOPE-DILUTION ICP-MS
    ITO, K
    KUBOYAMA, S
    OGURO, H
    NAKAMURA, S
    MUROZUMI, M
    BUNSEKI KAGAKU, 1989, 38 (12) : 697 - 700
  • [44] Environmental forensic application of lead isotope ratio determination: A case study using laser ablation sector ICP-MS
    Ghazi, AM
    Millette, JR
    ENVIRONMENTAL FORENSICS, 2004, 5 (02) : 97 - 108
  • [45] Determination of lead isotope ratios in environmental matrices by quadrupole ICP-MS working at low sample consumption rates
    Bazzano, Andrea
    Grotti, Marco
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2014, 29 (05) : 926 - 933
  • [46] EDXRF and ICP-MS analysis of environmental samples
    Nguyen, TH
    Boman, J
    Leermakers, M
    X-RAY SPECTROMETRY, 1998, 27 (04) : 265 - 276
  • [47] EDXRF and ICP-MS Analysis of Environmental Samples
    Physics and Engineering Physics, Chalmers University of Technology, S-412 96 Gothenburg, Sweden
    不详
    不详
    X-Ray Spectrom., 4 (265-276):
  • [48] Analysis of leaves by using the laser ICP-MS with isotope dissolution method
    Scholze, H
    Hoffmann, E
    Ludke, C
    Platalla, A
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1996, 355 (7-8): : 892 - 894
  • [49] A Rapid and Precise Determination of Boron Isotope Ratio in Water and Carbonate Samples by Multiple Collector ICP-MS
    Tanimizu, Masaharu
    Nagaishi, Kazuya
    Ishikawa, Tsuyoshi
    ANALYTICAL SCIENCES, 2018, 34 (06) : 667 - 674
  • [50] Effective Removal of Isobaric Interferences on Strontium and Lead Using Triple-Quadrupole ICP-MS
    Kutscher, Daniel
    Lofthouse, Simon
    Nelms, Simon
    Ducos, Shona McSheehy
    SPECTROSCOPY, 2017, 32 (10) : 38 - 41