Extended collimator model for pencil-beam dose calculation in proton radiotherapy

被引:28
作者
Kanematsu, Nobuyuki
Akagi, Takashi
Takatani, Yasuyuki
Yonai, Shunsuke
Sakamoto, Hidenobu
Yamashita, Haruo
机构
[1] Natl Inst Radiol Sci, Inage Ku, Chiba 2638555, Japan
[2] Hyogo Ion Beam Med Ctr, Tatsuno, Hyogo 6705165, Japan
[3] Mitsubishi Electr Corp, Hyogo Ku, Kobe, Hyogo 6528555, Japan
[4] Shizuoka Canc Ctr, Shizuoka 4118777, Japan
关键词
D O I
10.1088/0031-9155/51/19/006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We have developed a simple collimator model to improve the accuracy of penumbra behaviour in pencil-beam dose calculation for proton radiotherapy. In this model, transmission of particles through a three-dimensionally extended opening of a collimator is calculated in conjunction with phase-space distribution of the particles. Comparison of the dose distributions calculated using the new three-dimensional collimator model and the conventional two-dimensional model to lateral dose profiles experimentally measured with collimated proton beams showed the superiority of the new model over the conventional one.
引用
收藏
页码:4807 / 4817
页数:11
相关论文
共 27 条
[1]   Scatter factors in proton therapy with a broad beam [J].
Akagi, T ;
Kanematsu, N ;
Takatani, Y ;
Sakamoto, H ;
Hishikawa, Y ;
Abe, M .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (07) :1919-1928
[2]   Ridge filter design for proton therapy at Hyogo Ion Beam Medical Center [J].
Akagi, T ;
Higashi, A ;
Tsugami, H ;
Sakamoto, H ;
Masuda, Y ;
Hishikawa, Y .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (22) :N301-N312
[3]  
[Anonymous], JPN J MED PHYS
[4]  
[Anonymous], 1993, 49 ICRU
[5]   ELECTRON-BEAM DOSE PLANNING USING DISCRETE GAUSSIAN BEAMS - MATHEMATICAL BACKGROUND [J].
BRAHME, A ;
LAX, I ;
ANDREO, P .
ACTA RADIOLOGICA ONCOLOGY, 1981, 20 (02) :147-158
[6]   Monte Carlo and analytical calculation of proton pencil beams for computerized treatment plan optimization [J].
Carlsson, AK ;
Andreo, P ;
Brahme, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (06) :1033-1053
[7]   INSTRUMENTATION FOR TREATMENT OF CANCER USING PROTON AND LIGHT-ION BEAMS [J].
CHU, WT ;
LUDEWIGT, BA ;
RENNER, TR .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (08) :2055-2122
[8]   MULTIPLE COULOMB SCATTERING OF 160 MEV PROTONS [J].
GOTTSCHALK, B ;
KOEHLER, AM ;
SCHNEIDER, RJ ;
SISTERSON, JM ;
WAGNER, MS .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1993, 74 (04) :467-490
[9]   ELECTRON-BEAM DOSE CALCULATIONS [J].
HOGSTROM, KR ;
MILLS, MD ;
ALMOND, PR .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03) :445-459
[10]   Influence of multiple scattering and energy loss straggling on the absorbed dose distributions of therapeutic light ion beams: I. Analytical pencil beam model [J].
Hollmark, M ;
Uhrdin, J ;
Belkic, D ;
Gudowska, I ;
Brahme, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (14) :3247-3265