A NEW HIGH-SENSITIVITY PACKAGE-LEAK TESTING METHOD FOR MEMS SENSORS

被引:2
作者
Fujiyoshi, M. [1 ]
Nonomura, Y. [1 ]
Senda, H. [2 ]
机构
[1] Toyota Cent Res & Dev Labs Inc, Obu, Aichi, Japan
[2] Toyota Motor Co Ltd, Aichi, Japan
来源
IEEE 22ND INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS 2009) | 2009年
关键词
D O I
10.1109/MEMSYS.2009.4805488
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A new high-sensitivity package-leak testing method has been developed for micro electro mechanical system (MEMS) sensors such as a vibrating angular-rate sensor housed in a vacuum package. Procedures of the method to obtain high leak-rate resolution are as follows. (1) A package filled with helium gas is kept in a small and closed chamber to accumulate helium gas leaking out of the package. (2) The chamber is chilled with liquid-N-2 after the accumulation to reduce background gases except the helium gas. (3) The helium gas is transported to a mass spectrometer in a short time and recorded as a transient waveform. The leak-rate resolution of 1x10(-17) Pa.m(3)/s was obtained. This resolution was 10(5) times superior to that of the conventional method.