Distal Radius Volar Locking Plate Design and Associated Vulnerability of the Flexor Pollicis Longus

被引:51
作者
Limthongthang, Roongsak [1 ]
Bachoura, Abdo
Jacoby, Sidney M.
Osterman, A. Lee
机构
[1] Mahidol Univ, Dept Orthopaed Surg, Fac Med, Siriraj Hosp, Bangkok 10700, Thailand
来源
JOURNAL OF HAND SURGERY-AMERICAN VOLUME | 2014年 / 39卷 / 05期
关键词
Distal radius fracture; flexor tendon rupture; plate prominence; volar plating; watershed line; TENDON RUPTURES; FRACTURE; FIXATION; PLACEMENT;
D O I
10.1016/j.jhsa.2014.01.038
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Flexor pollicis longus (FPL) tendon rupture is a well-documented complication related to the use of distal radius volar locking plates (VLPs). The final common pathway of flexor tendon rupture appears to involve implants prominent at the watershed line. We hypothesized that significant differences in VLP prominence exist between various plate designs. Methods Ten fresh frozen specimens were dissected to identify the path of the FPL in relationship to the distal radius at the watershed line. Five VLP designs were fixed to each specimen based on their anatomic fit, and slid distally until the distal edge of the plate reached the watershed line. The position of each fixed plate was evaluated by fluoroscopy. We used a 3-dimensional laser scanner to create computer models. The total surface area of plate prominence volar to the watershed line and the prominent area beneath the FPL were measured in the axial plane using computer software. Results At the watershed line, the FPL was located at 54% of the maximal width of the radius, as measured from its volar-ulnar comer. There were no significant differences in the location of plate fixation on lateral view radiographs according to the classification of Soong et al. The mean total surface area of plate prominence was 36 mm(2). The mean prominent area beneath the FPL was 10 mm(2). Significant differences in plate prominence were noted for various designs. Conclusions Despite optimal plate placement, various VLP designs were observed to have prominent profiles volar to the watershed line, to varying extents. Copyright (C) 2014 by the American Society for Surgery of the Hand. All rights reserved.
引用
收藏
页码:852 / 860
页数:9
相关论文
共 23 条
  • [1] Rupture of flexor pollicis longus tendon - A complication of volar plating of the distal radius
    Bell, JSP
    Wollstein, R
    Citron, ND
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (02): : 225 - 226
  • [2] Precontoured fixed-angle volar distal radius plates: A comparison of anatomic fit
    Buzzell, Jonathan E.
    Weikert, Douglas R.
    Watson, Jeffry T.
    Lee, Donald H.
    [J]. JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2008, 33A (07): : 1144 - 1152
  • [3] FLEXOR POLLICIS LONGUS TENDON RUPTURES AFTER PALMAR PLATE FIXATION OF FRACTURES OF THE DISTAL RADIUS
    Casaletto, J. A.
    Machin, D.
    Leung, R.
    Brown, D. J.
    [J]. JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2009, 34E (04) : 471 - 474
  • [4] Effects of implant material and plate design on tendon function and morphology
    Cohen, Mark S.
    Turner, Thomas M.
    Urban, Robert M.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (445) : 81 - 90
  • [5] Flexor tendon injuries following locked volar plating of distal radius fractures
    Cross, Andrew W.
    Schmidt, Christopher C.
    [J]. JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2008, 33A (02): : 164 - 167
  • [6] Osteosynthesis of distal radial fractures with a volar locking screw plate system
    Drobetz, H
    Kutscha-Lissberg, E
    [J]. INTERNATIONAL ORTHOPAEDICS, 2003, 27 (01) : 1 - 6
  • [7] Distal Radial Fracture Treatment: What You Get May Depend on Your Age and Address
    Fanuele, Jason
    Koval, Kenneth J.
    Lurie, Jon
    Zhou, Weiping
    Tosteson, Anna
    Ring, David
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (06) : 1313 - 1319
  • [8] Fuller D J, 1973, J Bone Joint Surg Br, V55, P173
  • [9] HENRY AK, 1957, EXTENSILE EXPOSURE, P67
  • [10] An Anatomical Study of the Watershed Line on the Volar, Distal Aspect of the Radius: Implications for Plate Placement and Avoidance of Tendon Ruptures
    Imatani, Junya
    Akita, Keiichi
    Yamaguchi, Kumiko
    Shimizu, Hirotaka
    Kondou, Hidenori
    Ozaki, Toshifumi
    [J]. JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2012, 37A (08): : 1550 - 1554