Variable-rate selective excitation for rapid MRI sequences
被引:143
|作者:
Hargreaves, BA
[1
]
Cunningham, CH
[1
]
Nishimura, DG
[1
]
Conolly, SM
[1
]
机构:
[1] Stanford Univ, Dept Elect Engn, Magnet Resonance Syst Res Lab, Stanford, CA USA
关键词:
selective excitation;
VERSE;
SSFP;
rapid imaging;
True-FISP;
D O I:
10.1002/mrm.20168
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Balanced steady-state free precession (SSFP) imaging sequences require short repetition times (TRs) to avoid off-resonance artifacts. The use of slab-selective excitations is common, as this can improve imaging speed by limiting the field of view (FOV). However, the necessarily short-duration excitations have poor slab profiles. This results in unusable slices at the slab edge due to significant flip-angle variations or aliasing in the slab direction. Variable-rate selective excitation (VERSE) is a technique by which a time-varying gradient waveform is combined with a modified RF waveform to provide the same excitation profile with different RF power and duration characteristics. With the use of VERSE, it is possible to design short-duration pulses with dramatically improved slab profiles. These pulses achieve high flip angles with only minor off-resonance sensitivity, while meeting SAR limits at 1.5 T. The improved slab profiles will enable more rapid 3D imaging of limited volumes, with more consistent image contrast across the excited slab. (C) 2004 Wiley-Liss, Inc.