The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1: role of the third-body wear process

被引:23
作者
Affatato, S
Bersaglia, G
Foltran, I
Taddei, P
Fini, G
Toni, A
机构
[1] Ist Orteped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, Ctr Studio Spettroscop Raman, Dipartimento Biochim G Moruzzi, Sez Chim & Propedeut Biochim, I-40126 Bologna, Italy
[3] Div Ortoped & Traumatol 1, Bologna, Italy
关键词
morphology; debris; sterilisation; third-body wear process; SEM analysis;
D O I
10.1016/S0142-9612(02)00238-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Due to its excellent combination of properties, ultra-high-molecular-weight-polyethylene has been used for the last 30 years in the replacement of damaged articulating cartilage for total joint replacement surgery. However, in some cases, wear, failure and delamination have been observed. Polyethylene performance may be affected by oxidation during consolidation of the resin, sterilisation of the finished specimens and post-irradiation storage. In order to evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide(EtO)-treatment) and third-body particles on the ultra-high-molecular-weight-polyethylene wear behaviour, gamma- and EtO-sterilised ultra-high-molecular-weight-polyethylene acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5 million cycles in bovine calf serum in the presence of third-body PMMA particles. Weight loss measurements revealed that the gamma-sterilised acetabular cups exhibited a significantly lower wear rate than those EtO-sterilised. Moreover, significant differences were found for each type of sterilisation between the gravimetric wear trends obtained until 2.5 million cycles in the presence and in the absence of PMMA particles. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:4839 / 4846
页数:8
相关论文
共 72 条
  • [1] Mixed oxides prosthetic ceramic ball heads.: Part 1:: effect of the ZrO2 fraction on the wear of ceramic on polyethylene joints
    Affatato, S
    Testoni, M
    Cacciari, GL
    Toni, A
    [J]. BIOMATERIALS, 1999, 20 (10) : 971 - 975
  • [2] Effects of the sterilisation method on the wear of UHMWPE acetabular cups tested in a hip joint simulator
    Affatato, S
    Bordini, B
    Fagnano, C
    Taddei, P
    Tinti, A
    Toni, A
    [J]. BIOMATERIALS, 2002, 23 (06) : 1439 - 1446
  • [3] Mixed-oxides prosthetic ceramic ball heads.: Part 2:: Effect of the ZrO2 fraction on the wear of ceramic on ceramic joints
    Affatato, S
    Testoni, M
    Cacciari, GL
    Toni, A
    [J]. BIOMATERIALS, 1999, 20 (20) : 1925 - 1929
  • [4] Mixed oxides prosthetic ceramic ball heads.: Part 3:: effect of the ZrO2 fraction on the wear of ceramic on ceramic hip joint prostheses.: A long-term in vitro wear study
    Affatato, S
    Goldoni, M
    Testoni, M
    Toni, A
    [J]. BIOMATERIALS, 2001, 22 (07) : 717 - 723
  • [5] AFFATATO S, 2000, P 12 C EUR SOC BIOM, P453
  • [6] Bargmann LS, 1999, CLIN ORTHOP RELAT R, P49
  • [7] THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT
    BARTEL, DL
    BICKNELL, VL
    WRIGHT, TM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) : 1041 - 1051
  • [8] Micro-Raman spectroscopy for the crystallinity characterization of UHMWPE hip cups run on joint simulators
    Bertoluzza, A
    Fagnano, C
    Rossi, M
    Tinti, A
    Cacciari, GL
    [J]. JOURNAL OF MOLECULAR STRUCTURE, 2000, 521 : 89 - 95
  • [9] BLOEBAUM RD, 1991, CLIN ORTHOP RELAT R, V269, P120
  • [10] BOGGAN RS, 1993, P 19 ANN M SOC BIOM, P329