Effect of isoflurane on defibrillation threshold in biphasic active-can defibrillation systems

被引:0
作者
Horvath, G [1 ]
Ilan, M [1 ]
Kadish, A [1 ]
Goldberger, J [1 ]
机构
[1] Northwestern Mem Hosp, Sect Cardiac Electrophysiol, Dept Internal Med, Div Cardiol, Chicago, IL 60611 USA
关键词
anesthesia; defibrillation; drug effect; implantable defibrillator; ventricular fibrillation;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In two patients receiving implantable cardioverter-defibrillators (ICD) for spontaneous and inducible ventricular tarhyarrhythmias, the use of isoflurane appeared to significantly lower defibrillation thresholds (DFT) at implantation. In the first, at initial implantation and subsequent revision using isoflurane, adequate DFTs mere verified by multiple tests, Post-implantation testing using midazolam and fentanyl revealed significantly higher DFTs, necessitating a third operation without isoflurane to obtain an adequate and subsequently verifiable DFT, During the second case, discontinuation of isoflurane intraoperatively resulted in a DFT rise within 25 minutes. In patients receiving general anesthesia for ICD implantation, this possible DFT effect of isoflurane should be recognized.
引用
收藏
页码:700 / 702
页数:3
相关论文
共 4 条
[1]   THE DEFIBRILLATION THRESHOLD - A COMPARISON OF ANESTHETICS AND MEASUREMENT METHODS [J].
GILL, RM ;
SWEENEY, RJ ;
REID, PR .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (04) :708-714
[2]   A PILOT-STUDY - DEFIBRILLATION THRESHOLDS IN DOGS ARE SIMILAR WITH ISOFLURANE, HALOTHANE, AND PENTOBARBITAL [J].
JARVIS, AS ;
LAHTINEN, SP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (03) :280-285
[3]   Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia [J].
Moss, AJ ;
Hall, WJ ;
Cannom, DS ;
Daubert, JP ;
Higgins, SL ;
Klein, H ;
Levine, JH ;
Saksena, S ;
Waldo, AL ;
Wilber, D ;
Brown, MW ;
Heo, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1996, 335 (26) :1933-1940
[4]   SURVIVAL AFTER IMPLANTATION OF THE CARDIOVERTER DEFIBRILLATOR [J].
NEWMAN, D ;
SAUVE, MJ ;
HERRE, J ;
LANGBERG, JJ ;
LEE, MA ;
TITUS, C ;
FRANKLIN, J ;
SCHEINMAN, MM ;
GRIFFIN, JC .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (09) :899-903