Measurement of intraocular pressure using the Tono-Pen in comparison with Goldmann applanation tonometry - a clinical study in 100 eyes

被引:13
|
作者
Deuter, CME [1 ]
Schlote, T [1 ]
Hahn, GA [1 ]
Bende, T [1 ]
Derse, M [1 ]
机构
[1] Univ Tubingen, Augenklin, Abt Erkrankungen Vorderen & Hinteren Augenabschni, D-72076 Tubingen, Germany
关键词
glaucoma; intraocular pressure; tonometry;
D O I
10.1055/s-2002-26718
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: In clinical practice ophthalmologists often need a tonometer which is independent of a slit lamp. Such a hand-held device is the Tono-Pen. We compared the precision of two equal Tono-Pens with Goldmann applanation tonometry. Material and methods: Measurement of intraocular pressure (IOP) was done in 100 eyes of 51 patients (mean age 63 15 years) suffering from ocular hypertension or glaucoma. According to a random table either the right or left eye was measured using Goldmann tonometer first and the Torio-Pen second. For the other eye the measurement was reversed. One of the two equal Tono-Pens (So-Ian/USA) was used according to a second random table. Three measurements were obtained with each instrument on both eyes within 15 minutes subsequently. Patients were placed in an upright position for all measurements. Results: Even for well-trained ophthalmologists a learning curve of approximately 10 measurements was observed using the Tono-Pen. The Tono-Pen measured an average IOP of 16.9 mm Hg in all 100 eyes. The Goldmann tonometer measured an average IOP of 17.7 mm Hg. The difference was not statistically significant. The standard deviation for all measurements was better for the Tono-Pen (4.7 mm Hg vs 5.8 mm Hg for Goldmann tonometer). No reduction of the IOP after Tono-Pen measurement was observed (in contrast to the Goldmann tonometer). The reproducibility of the Tono-Pen on the same eye was inferior to the Goldmann tonometer by a factor of 2. There was an almost significant difference in reproducibility between two equal Tono-Pens. Conclusions: Measurement of IOP with the Tono-Pen is comparable to Goldmann applanation tonometry if an average of 3 measurements is used. The difference between two equal Tono-Pens indicates the need for improvement of the quality check during production.
引用
收藏
页码:138 / 142
页数:5
相关论文
共 50 条
  • [31] Intraocular pressure measurement - Comparison of dynamic contour tonometry and Goldmann applanation tonometry
    Schneider, E
    Grehn, F
    JOURNAL OF GLAUCOMA, 2006, 15 (01) : 2 - 6
  • [32] Measurement of mouse intraocular pressure with the Tono-Pen
    Pang, IH
    Wang, WH
    Millar, JC
    Clark, AF
    EXPERIMENTAL EYE RESEARCH, 2005, 81 (03) : 359 - 360
  • [33] INTRAOCULAR-PRESSURE MEASUREMENT AFTER PENETRATING KERATOPLASTY - MINIFIED GOLDMANN APPLANATION TONOMETER, PNEUMATONOMETER, AND TONO-PEN VERSUS MANOMETRY
    MENAGE, MJ
    KAUFMAN, PL
    CROFT, MA
    LANDAY, SP
    BRITISH JOURNAL OF OPHTHALMOLOGY, 1994, 78 (09) : 671 - 676
  • [34] Comparison of Goldmann Applanation Tonometry and Dynamic Contour Tonometry in the Measurement of Intraocular Pressure in Eyes with Different Corneal Thicknesses
    Roszkowska, A. M.
    De Grazia, L.
    Cirone, M.
    Ferreri, G.
    OPHTHALMOLOGICA, 2009, 223 (04) : 244 - 249
  • [35] INTRAOCULAR-PRESSURE MEASUREMENT AFTER PENETRATING KERATOPLASTY - MINIFIED GOLDMANN APPLANATION TONOMETER, PNEUMATONOMETER AND TONO-PEN VERSUS MANOMETRY
    KAUFMAN, PL
    MENAGE, M
    CROFT, MA
    LANDAY, S
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1994, 35 (04) : 1874 - 1874
  • [36] Comparison of Tono-Pen Avia and Handheld Applanation Tonometry in Primary Congenital Glaucoma
    Cronemberger, Sebastiao
    Veloso, Artur W.
    JOURNAL OF GLAUCOMA, 2021, 30 (05) : E227 - E230
  • [37] Rebound tonometry by iCare 200 (IC200): comparison with Tono-Pen in the operating room and Goldmann applanation in the clinic
    Umfress, Allison C.
    Glaser, Tanya S.
    Ploysangam, Pimpiroon
    Freedman, Sharon F.
    JOURNAL OF AAPOS, 2021, 25 (06):
  • [38] THE EFFECT OF PERKINS, TONO-PEN, AND SCHIOTZ TONOMETRY ON INTRAOCULAR-PRESSURE
    WHITACRE, MM
    EMIG, M
    HASSANEIN, K
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1991, 111 (01) : 59 - 64
  • [39] MEASUREMENT OF RABBIT INTRAOCULAR-PRESSURE WITH THE TONO-PEN
    MERMOUD, A
    BAERVELDT, G
    MINCKLER, DS
    LEE, MB
    RAO, NA
    OPHTHALMOLOGICA, 1995, 209 (05) : 275 - 277
  • [40] Evaluation of the applanation tonometer Tono-Pen Avia® Vet™ for the measurement of the canine and feline intraocular pressure
    Helbig, Stefanie
    Eule, J. Corinna
    TIERAERZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2021, 49 (02): : 86 - 92