On-line determination of respiration rates of aquatic organisms in a mono-phase oxystat at steady-state dissolved oxygen tensions

被引:10
|
作者
Eriksen, NT [1 ]
Iversen, JJL [1 ]
机构
[1] ODENSE UNIV, INST BIOCHEM, DK-5230 ODENSE M, DENMARK
关键词
D O I
10.1007/s002270050081
中图分类号
Q17 [水生生物学];
学科分类号
071004 ;
摘要
On-line determination of respiration rates at steady-state dissolved oxygen tensions was performed in a thermostatted measuring chamber equipped with a galvanic oxygen electrode. The measuring chamber is designed to completely exclude gas bubbles from the liquid phase and is named a mono-phase oxystat. Additions of oxygenated water to maintain steady-state dissolved oxygen tensions are controlled by a computer, and the respiration rate is determined from the rate of these additions. An external loop with a flow cell in a spectrophotometer makes it possible to determine and control the concentration of algal cells in the monophase oxystat and allowed simultaneous determination of respiration and filtration rates of filter-feeding invertebrates. Because the system maintains steady-state conditions the experiments do not suffer from time limitations. The response time of the system is negligible compared to the response time of other open systems.