Computational Study on the Thermal Effects of Implantable Magnetic Stimulation Based on Planar Coils

被引:21
作者
Park, Hee-Jin [1 ]
Seol, Jae Hun [1 ]
Ku, Jeonghun [2 ]
Kim, Sohee [1 ,3 ]
机构
[1] Gwangju Inst Sci & Technol, Sch Mechatron, Gwangju 500712, South Korea
[2] Keimyung Univ, Dept Biomed Engn, Daegu, South Korea
[3] Gwangju Inst Sci & Technol, Dept Med Syst Engn, Gwangju 500712, South Korea
基金
新加坡国家研究基金会;
关键词
Brain temperature; finite-element analysis (FEA); implantable magnetic stimulation; planar coil; temperature increase; DEEP BRAIN-STIMULATION; COMPUTED ELECTRIC-FIELD; MOTOR THRESHOLD; TEMPERATURE; TISSUE; VOLUME; PERFUSION; DENSITY; NEURONS; HUMANS;
D O I
10.1109/TBME.2015.2490244
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Goal: Micromagnetic stimulation using coils sufficiently small to be implanted has been suggested as a potential method to overcome the limitations of electrical stimulation. We investigated the temperature increases in the brain implanted with planar coils. Methods: We conducted computational simulations on the thermal effects of implantable magnetic stimulation in a brain model using finite-element analysis, by varying geometric parameters of planar spiral coils, and repetitive stimulation pulse patterns. Results: It was revealed that the temperature increase in the brain can be controlled by a careful design of coils to implant. The coil diameter greater than 8 mm was required to induce a temperature rise in the brain of less than 1 degrees C. If the coil diameter was larger than 10 mm, the subsequent temperature rises in the brain tissue was maintained consistently at about 0.24 degrees C or lower. Conclusion: Selection of the number of coil turns must rely on tradeoffs between the required current and voltage that the implanted source can generate, as the required voltage increases while the required current decreases with increasing number of coil turns. Additionally, the coil insulation with a thickness of a few micrometers was found to suppress the temperature rise in the brain effectively. Significance: Although these simulations employed only one threshold value of 10 V/m, which is rather on the lower end of stimulation threshold, the simulation results are expected to serve as guidelines for designing planar coils to be implanted in the brain for magnetic stimulation.
引用
收藏
页码:158 / 167
页数:10
相关论文
共 63 条
[1]   An implantable neural interface with electromagnetic stimulation capabilities [J].
Accoto, D. ;
Francomano, M. T. ;
Rainer, A. ;
Trombetta, M. ;
Rossini, P. M. ;
Guglielmelli, E. .
MEDICAL HYPOTHESES, 2013, 81 (02) :322-327
[2]  
Ahlbom A, 1998, HEALTH PHYS, V74, P494
[3]  
[Anonymous], 2013, POWERMAG RES TECHN S
[4]  
[Anonymous], 2007, Transcranial magnetic stimulation in clinical psychiatry
[5]  
[Anonymous], 2011, DUPONT KAPT HN POL F
[6]   Epidural electrical stimulation to improve chronic poststroke aphasia: A 5-year follow-up [J].
Balossier, Anne ;
Etard, Olivier ;
Descat, Chloe ;
Vivien, Denis ;
Emery, Evelyne .
BRAIN STIMULATION, 2012, 5 (03) :364-368
[7]  
Bergman T. L., 2011, FUNDAMENTALS HEAT MA, P183
[8]   Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays [J].
Biran, R ;
Martin, DC ;
Tresco, PA .
EXPERIMENTAL NEUROLOGY, 2005, 195 (01) :115-126
[9]   Microscopic magnetic stimulation of neural tissue [J].
Bonmassar, Giorgio ;
Lee, Seung Woo ;
Freeman, Daniel K. ;
Polasek, Miloslav ;
Fried, Shelley I. ;
Gale, John T. .
NATURE COMMUNICATIONS, 2012, 3
[10]   MODULATION BY APPLIED ELECTRIC-FIELDS OF PURKINJE AND STELLATE CELL-ACTIVITY IN THE ISOLATED TURTLE CEREBELLUM [J].
CHAN, CY ;
NICHOLSON, C .
JOURNAL OF PHYSIOLOGY-LONDON, 1986, 371 :89-114