Use of spherical harmonic deconvolution methods to compensate for nonlinear gradient effects on MRI images

被引:124
作者
Janke, A [1 ]
Zhao, HW [1 ]
Cowin, GJ [1 ]
Galloway, GJ [1 ]
Doddrell, DM [1 ]
机构
[1] Univ Queensland, Ctr Magnet Resonance, St Lucia, Qld 4072, Australia
关键词
MRI; gradients; distortion correction; NMR plotting probe; spherical harmonics;
D O I
10.1002/mrm.20122
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Spatial encoding in MR techniques is achieved by sampling the signal as a function of time in the presence of a magnetic field gradient. The gradients are assumed to generate 6 linear magnetic field gradient, and typical image reconstruction relies upon this approximation. However, high-speed gradients in the current generation of MRI scanners often sacrifice linearity for improvements in speed. Such nonlinearity results in distorted images. The problem is presented in terms of first principles and a correction method based on a gradient field spherical harmonic expansion is proposed. In our case, the amount of distortion measured within a typical field of view (FOV) required for head imaging is sufficiently large that without the use of some distortion correction technique, the images would be of limited use for stereotaxy or longitudinal studies, where precise volumetric information is required. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:115 / 122
页数:8
相关论文
共 22 条
[1]   ANALYSIS OF MACHINE-DEPENDENT AND OBJECT-INDUCED GEOMETRIC DISTORTION IN 2DFT MR IMAGING [J].
BAKKER, CJG ;
MOERLAND, MA ;
BHAGWANDIEN, R ;
BEERSMA, R .
MAGNETIC RESONANCE IMAGING, 1992, 10 (04) :597-608
[2]   A TECHNIQUE FOR ACCURATE MAGNETIC-RESONANCE-IMAGING IN THE PRESENCE OF FIELD INHOMOGENEITIES [J].
CHANG, H ;
FITZPATRICK, JM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (03) :319-329
[3]  
Chen NK, 2001, MAGNET RESON MED, V45, P525, DOI 10.1002/1522-2594(200103)45:3<525::AID-MRM1070>3.0.CO
[4]  
2-S
[5]   Animal: Validation and applications of nonlinear registration-based segmentation [J].
Collins, DL ;
Evans, AC .
INTERNATIONAL JOURNAL OF PATTERN RECOGNITION AND ARTIFICIAL INTELLIGENCE, 1997, 11 (08) :1271-1294
[6]   An evaluation of the use of magnetic field maps to undistort echo-planar images [J].
Cusack, R ;
Brett, M ;
Osswald, K .
NEUROIMAGE, 2003, 18 (01) :127-142
[7]   TEMPORAL SPHERICAL-HARMONIC EXPANSION AND COMPENSATION OF EDDY-CURRENT FIELDS PRODUCED BY GRADIENT PULSES [J].
ECCLES, CD ;
CROZIER, S ;
WESTPHAL, M ;
DODDRELL, DM .
JOURNAL OF MAGNETIC RESONANCE SERIES A, 1993, 103 (02) :135-141
[8]   Aspects of MR image distortions in radiotherapy treatment planning [J].
Fransson, A ;
Andreo, P ;
Pötter, R .
STRAHLENTHERAPIE UND ONKOLOGIE, 2001, 177 (02) :59-73
[9]  
KRIEG R, 2002, Patent No. 6501273
[10]  
Langlois S, 1999, JMRI-J MAGN RESON IM, V9, P821, DOI 10.1002/(SICI)1522-2586(199906)9:6<821::AID-JMRI9>3.0.CO