Accuracy of electron density, effective atomic number, and iodine concentration determination with a dual-layer dual-energy computed tomography system

被引:135
作者
Hua, Chia-ho [1 ]
Shapira, Nadav [2 ]
Merchant, Thomas E. [1 ]
Klahr, Paul [3 ]
Yagil, Yoad [2 ]
机构
[1] St Jude Childrens Res Hosp, Dept Radiat Oncol, 332 N Lauderdale St, Memphis, TN 38105 USA
[2] Philips Med Syst, Global Adv Technol, IL-3190500 Haifa, Israel
[3] Philips Healthcare, CT Clin Sci, Cleveland, OH 44143 USA
关键词
dual-layer CT; effective atomic number; electron density; iodine quantitation; material separation; TISSUE SEGMENTATION; CT NUMBERS; QUANTIFICATION; SIMULATION; PRINCIPLES; RADIOTHERAPY; SUBTRACTION; DETECTOR; SINGLE; RATIO;
D O I
10.1002/mp.12903
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This study aimed to quantitate the accuracy of the determination of electron density (ED), effective atomic number (Zeff), and iodine concentration, directed for more accurate radiation therapy planning, with a new dual-layer dual-energy computed tomography (DL-DECT) system. The dependence of the accuracy of these values on the scan and reconstruction parameters, as well as on the phantom size, was also examined. Methods: Measurements were performed on a commercial DECT system with a DL detector (IQon Spectral CT, Philips Healthcare), using phantoms with various tissue-equivalent inserts as well as iodine and calcium inserts of different concentrations. The expected values of ED and Zeff for the insert materials were derived from the chemical compositions provided by the vendors. The nominal scan condition for the accuracy measurements was 120 kVp, 20 mGy CTDIvol, 0.812 pitch, 16 x 0.625 mm collimation, and 0.33-second gantry rotation. Results: The median deviation of ED ranged from -0.1% to 1.1% for all Gammex tissue inserts. The median deviation of Zeff ranged from -2.3% to 1.7% for soft tissue and bone inserts and was <= 7% for lung inserts. The absolute deviations for ED and Zeff in lung inserts were within 1% of the ED of water and 1 a.u., respectively. For two different phantom sizes, the ED values agreed to within 0.7% and the Z(eff) values agreed to within 2%, except for the lung inserts. When the scan parameters were changed from 120 kVp/20 mGy to 140 kVp/30 mGy, the ED differed within [-0.51%, 0.65%] and the Z(eff) differed within [-1.1%, 0.23%] for all materials except lungs, in which Z(eff) increased by 2.4%. The accuracy of ED and Z(eff) measurement at 120 kVp was no worse than that at 140 kVp. For iodine quantitation, the median absolute deviations from the nominal values were up to 0.3 mg/mL for iodine concentrations of 2-20 mg/mL, with an overall median deviation of -0.1 mg/mL. Iodine and calcium were well separated on the ED-Z(eff) scatter plot, even at the lowest concentrations (2 mg/mL for iodine and 50 mg/mL for calcium). Conclusions: The accuracy of ED measurement, Z(eff) determination, and iodine quantitation derived from DL-DECT was demonstrated with phantom measurements. The accuracies were not sensitive to scan and reconstruction parameters, namely tube potential, dose, rotation time, and spectral reconstruction level, especially in the case of electron density. (c) 2018 American Association of Physicists in Medicine
引用
收藏
页码:2486 / 2497
页数:12
相关论文
共 34 条
[1]   Dual-energy CT quantitative imaging: a comparison study between twin-beam and dual-source CT scanners [J].
Almeida, Isabel P. ;
Schyns, Lotte E. J. R. ;
Ollers, Michel C. ;
van Elmpt, Wouter ;
Parodi, Katia ;
Landry, Guillaume ;
Verhaegen, Frank .
MEDICAL PHYSICS, 2017, 44 (01) :171-179
[2]   A Double-Layer Detector, Dual-Energy CT - Principles, Advantages and Applications [J].
Altman, A. ;
Carmi, R. .
MEDICAL PHYSICS, 2009, 36 (06) :2750-+
[3]  
Altman A, 2013, MED PHYS, V40, P450
[4]   ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[5]  
[Anonymous], XCOM: Photon cross sections database
[6]   Dual-Phase Dual-Energy CT in Patients Treated with Erlotinib for Advanced Non-Small Cell Lung Cancer: Possible Benefits of Iodine Quantification in Response Assessment [J].
Baxa, Jan ;
Matouskova, Tana ;
Krakorova, Gabriela ;
Schmidt, Bernhard ;
Flohr, Thomas ;
Sedlmair, Martin ;
Bejcek, Jiri ;
Ferda, Jiri .
EUROPEAN RADIOLOGY, 2016, 26 (08) :2828-2836
[7]   Dual-energy CT-based material extraction for tissue segmentation in Monte Carlo dose calculations [J].
Bazalova, Magdalena ;
Carrier, Jean-Francois ;
Beaulieu, Luc ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (09) :2439-2456
[8]   Tissue segmentation in Monte Carlo treatment planning: A simulation study using dual-energy CT images [J].
Bazatova, Magdatena ;
Carrier, Jean-Francois ;
Beautieu, Luc ;
Verhaegen, Frank .
RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) :93-98
[9]   Quantitative therapy response assessment by volumetric iodine-uptake measurement: Initial experience in patients with advanced hepatocellular carcinoma treated with sorafenib [J].
Dai, Xu ;
Schlemmer, Heinz-Peter ;
Schmidt, Bernhard ;
Hoeh, Karolin ;
Xu, Ke ;
Ganten, Tom M. ;
Ganten, Maria-Katharina .
EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (02) :327-334
[10]   Performance of today's dual energy CT and future multi energy CT in virtual non-contrast imaging and in iodine quantification: A simulation study [J].
Faby, Sebastian ;
Kuchenbecker, Stefan ;
Sawall, Stefan ;
Simons, David ;
Schlemmer, Heinz-Peter ;
Lell, Michael ;
Kachelriess, Marc .
MEDICAL PHYSICS, 2015, 42 (07) :4349-4366