Automated on-line liquid chromatography-photodiode array-mass spectrometr method with dilution line for the determination of bisphenol A and 4-octylphenol in serum

被引:67
作者
Liu, Min
Hashi, Yuki [1 ]
Pan, Fengyun
Yao, Jianguo
Song, Guanqun
Lin, Jin-Ming
机构
[1] Shimadzu Hong Kong Ltd, Beijing Off, Analyt Applicat Ctr, Beijing 100020, Peoples R China
[2] Chinese Acad Sci, State Key Lab Environm Chem & Ecotoxicol, Res Ctr Ecoenvironm Sci, Beijing 100085, Peoples R China
基金
中国国家自然科学基金;
关键词
on-line LC-DAD-MS; dilution line; bisphenol A; 4-octyphenol;
D O I
10.1016/j.chroma.2006.08.009
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel on-line liquid chromatography-photodiode array detection-mass spectrometry (LC-DAD-MS) system was established with restricted-access media (RAM) pre-column and dilution line combined with a column-switching valve. The serum samples were injected directly onto pre-column under diluted condition by dilution line. After elution of proteins in the serum, the analytes were backflushed onto an ODS analytical column using a six-port column-switching device. The influence of the composition of the mobile phase, for instance, organic modifer, ionic strength, pH, dilution times and the rotation time of the switching valve have been investigated using bisphenol A (BPA) and 4-octyphenol (4-OP) as analytes. The evaluations for peak responses and sensitivity were conducted by MS, and proteins were removed by RAM-column with DAD monitoring at 280 nm. The peak shape was improved by adding a dilution line, especially in the case of large volume injection (LVI), which increased the sensitivity of the analysis. The selective and sensitive quantification of BPA and 4-OP in serum sample could be finished within 25 min. The method had linearity in the range 0.1-500 ng/mL with a limit of quantification for BPA and 4-OP of 0.1 and 0.5 ng/mL, respectively. The recoveries were in the range of 80-101% with less than 9.0% RSDs. This on-line LC-MS method demonstrates potential application to evaluating the exposure and risk of BPA and 4-OP in human. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:142 / 148
页数:7
相关论文
共 35 条
  • [1] Determination of SCH 211803 by nanoelectrospray infusion mass spectrometry: evaluation of matrix effect and comparison with liquid chromatography-tandem mass spectrometry
    Chen, JW
    Yang, LY
    Kapron, JT
    Ma, L
    Pace, E
    Van Pelt, CK
    Rudewicz, PJ
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 809 (02): : 205 - 210
  • [2] Use of a novel cation-exchange restricted-access material for automated sample clean-up prior to the determination of basic drugs in plasma by liquid chromatography
    Chiap, P
    Rbeida, O
    Christiaens, B
    Hubert, P
    Lubda, D
    Boos, KS
    Crommen, J
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2002, 975 (01) : 145 - 155
  • [3] A new and rapid method for monitoring the new oxazolidinone antibiotic linezolid in serum and urine by high performance liquid chromatography-integrated sample preparation
    Ehrlich, M
    Trittler, R
    Daschner, FD
    Kümmerer, K
    [J]. JOURNAL OF CHROMATOGRAPHY B, 2001, 755 (1-2): : 373 - 377
  • [4] Determination of retinol and retinyl esters in human plasma by high-performance liquid chromatography with automated column switching and ultraviolet detection
    Hartmann, S
    Froescheis, O
    Ringenbach, F
    Wyss, R
    Bucheli, F
    Bischof, S
    Bausch, J
    Wiegand, UW
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2001, 751 (02): : 265 - 275
  • [5] Determination of pesticides in vegetables using large-volume injection column liquid chromatography-electrospray tandem mass spectrometry
    Hogenboom, AC
    Hofman, MP
    Kok, SJ
    Niessen, WMA
    Brinkman, UAT
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2000, 892 (1-2) : 379 - 390
  • [6] HOLGER MK, 2003, J CHROMATOGR B, V784, P169
  • [7] Size-exclusion flow extraction of bisphenol A in human urine for liquid chromatography-mass spectrometry
    Inoue, K
    Kawaguchi, M
    Funakoshi, Y
    Nakazawa, H
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 798 (01): : 17 - 23
  • [8] Application of liquid chromatography-mass spectrometry to the quantification of bisphenol A in human semen
    Inoue, K
    Wada, M
    Higuchi, T
    Oshio, S
    Umeda, T
    Yoshimura, Y
    Nakazawa, H
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 773 (02): : 97 - 102
  • [9] Quantitative detection of bisphenol A and bisphenol A diglycidyl ether metabolites in human plasma by liquid chromatography-electrospray mass spectrometry
    Inoue, K
    Yamaguchi, A
    Wada, M
    Yoshimura, Y
    Makino, T
    Nakazawa, H
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2001, 765 (02): : 121 - 126
  • [10] Determination of bisphenol A in river water and body fluid samples by stir bar sorptive extraction with in situ derivatization and thermal desorption-gas chromatography-mass spectrometry
    Kawaguchi, M
    Inoue, K
    Yoshimura, M
    Ito, R
    Sakui, N
    Okanouchi, N
    Nakazawa, H