A comparison of modified Howland circuits as current generators with current mirror type circuits

被引:72
作者
Bertemes, P [1 ]
Brown, BH [1 ]
Wilson, AJ [1 ]
机构
[1] Univ Sheffield, Royal Hallamshire Hosp, Dept Med Phys & Clin Engn, Sheffield S10 2JF, S Yorkshire, England
关键词
current sources; impedance imaging; evaluation;
D O I
10.1088/0967-3334/21/1/301
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Multi-frequency electrical impedance tomography (EIT)) systems require stable voltage controlled current generators that will work over a wide frequency range and with a large variation in load impedance. In this paper we compare the performance of two commonly used designs: the first is a modified Howland circuit whilst the second is based on a current mirror. The output current and the output impedance of both circuits were determined through PSPICE simulation and through measurement. Both circuits were stable over the frequency ranges 1 kHz to 1 MHz. The maximum variation of output current with frequency for the modified Howland circuit was 2.0% and for the circuit based on a current mirror 1.6%. The output impedance for both circuits was greater than 100 k Omega for frequencies up to 100 kHz. However, neither circuit achieved this output impedance at 1 MHz. Comparing the results from the two circuits suggests that there is little to choose between them in terms of a practical implementation.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 17 条
[1]   A CURRENT INJECTING DEVICE FOR ELECTRICAL-IMPEDANCE TOMOGRAPHY [J].
BLAD, B ;
LINDSTROM, K ;
BERTENSTAM, L ;
PERSSON, BRR ;
HOLMER, NG .
PHYSIOLOGICAL MEASUREMENT, 1994, 15 :A69-A77
[2]  
BOONE KG, 1996, PHYSL MEAS S1A, V17, pA229
[3]   A WIDE-BAND AC-COUPLED CURRENT SOURCE FOR ELECTRICAL-IMPEDANCE TOMOGRAPHY [J].
BRAGOS, R ;
ROSELL, J ;
RIU, P .
PHYSIOLOGICAL MEASUREMENT, 1994, 15 :A91-A99
[4]   MULTIFREQUENCY IMAGING AND MODELING OF RESPIRATORY-RELATED ELECTRICAL-IMPEDANCE CHANGES [J].
BROWN, BH ;
BARBER, DC ;
WANG, W ;
LU, LQ ;
LEATHARD, AD ;
SMALLWOOD, RH ;
HAMPSHIRE, AR ;
MACKAY, R ;
HATZIGALANIS, K .
PHYSIOLOGICAL MEASUREMENT, 1994, 15 :A1-A12
[5]   ASSESSMENT OF 4 ENZYME-IMMUNOASSAY KITS FOR SERUM BETA-HCG SCREENING [J].
BROWN, D ;
ALLEN, LC .
JOURNAL OF IMMUNOASSAY, 1987, 8 (01) :1-9
[6]   ACT3 - A HIGH-SPEED, HIGH-PRECISION ELECTRICAL-IMPEDANCE TOMOGRAPH [J].
COOK, RD ;
SAULNIER, GJ ;
GISSER, DG ;
GOBLE, JC ;
NEWELL, JC ;
ISAACSON, D .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (08) :713-722
[7]  
CUSICK G, 1994, INNOV TECH BIOL S1A, V15, pA33
[8]   A HIGH-OUTPUT IMPEDANCE CURRENT SOURCE [J].
DENYER, CW ;
LIDGEY, FJ ;
ZHU, QS ;
MCLEOD, CN .
PHYSIOLOGICAL MEASUREMENT, 1994, 15 :A79-A82
[9]   BIOELECTRICAL SPECTROSCOPY FROM MULTIFREQUENCY EIT [J].
GRIFFITHS, H ;
JOSSINET, J .
PHYSIOLOGICAL MEASUREMENT, 1994, 15 :A59-A63
[10]   ACTIVE CURRENT ELECTRODES FOR IN-VIVO ELECTRICAL-IMPEDANCE TOMOGRAPHY [J].
JOSSINET, J ;
TOURTEL, C ;
JARRY, R .
PHYSIOLOGICAL MEASUREMENT, 1994, 15 :A83-A90