Positioning error of custom 3D-printed surgical guides for the radius: influence of fitting location and guide design

被引:32
作者
Caiti, G. [1 ]
Dobbe, J. G. G. [1 ]
Strijkers, G. J. [1 ]
Strackee, S. D. [2 ]
Streekstra, G. J. [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Biomed Engn & Phys, Amsterdam, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Plast Reconstruct & Hand Surg, Amsterdam, Netherlands
关键词
Radius osteotomy; Patient-specific surgical guides; 3D-printing; Computer-assisted orthopedic surgery; DISTAL RADIUS; CORRECTIVE OSTEOTOMY; COMPUTER-SIMULATION; PLATE; INSTRUMENT; NAVIGATION; ACCURACY; FIXATION;
D O I
10.1007/s11548-017-1682-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Utilization of 3D-printed patient-specific surgical guides is a promising navigation approach for orthopedic surgery. However, navigation errors can arise if the guide is not correctly positioned at the planned bone location, compromising the surgical outcome. Quantitative measurements of guide positioning errors are rarely reported and have never been related to guide design and underlying bone anatomy. In this study, the positioning accuracy of a standard and an extended guide design with lateral extension is evaluated at different fitting locations (distal, mid-shaft and proximal) on the volar side of the radius. Four operators placed the surgical guides on 3D-printed radius models obtained from the CT scans of six patients. For each radius model, every operator positioned two guide designs on the three fitting locations. The residual positioning error was quantified with a CT-based image analysis method in terms of the mean target registration error (mTRE), total translation error () and total rotation error () by comparing the actual guide position with the preoperatively planned position. Three generalized linear regression models were constructed to evaluate if the fitting location and the guide design affected mTRE, and . mTRE, and were significantly higher for mid-shaft guides () compared to distal guides. The guide extension significantly improved the target registration and translational accuracy in all the volar radius locations (). However, in the mid-shaft region, the guide extension yielded an increased total rotational error (). Our study demonstrates that positioning accuracy depends on the fitting location and on the guide design. In distal and proximal radial regions, the accuracy of guides with lateral extension is higher than standard guides and is therefore recommended for future use.
引用
收藏
页码:507 / 518
页数:12
相关论文
共 29 条
[1]   Automatic Segmentation of Wrist Bones in CT Using a Statistical Wrist Shape plus Pose Model [J].
Abu Anas, Emran Mohammad ;
Rasoulian, Abtin ;
Seitel, Alexander ;
Darras, Kathryn ;
Wilson, David ;
St John, Paul ;
Pichora, David ;
Mousavi, Parvin ;
Rohling, Robert ;
Abolmaesumi, Purang .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2016, 35 (08) :1789-1801
[2]   Adding Value in Additive Manufacturing [J].
Banks, Jim .
IEEE PULSE, 2013, 4 (06) :22-26
[3]   Nonoperative Treatment of Both-bone Forearm Shaft Fractures in Children: Predictors of Early Radiographic Failure [J].
Bowman, Eric N. ;
Mehlman, Charles T. ;
Lindsell, Christopher J. ;
Tamai, Junichi .
JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2011, 31 (01) :23-32
[4]   Emerging applications of bedside 3D printing in plastic surgery [J].
Chae, Michael P. ;
Rozeni, Warren M. ;
McMenamin, Paul G. ;
Findlay, Michael W. ;
Spychal, Robert T. ;
Hunter-Smith, David J. .
FRONTIERS IN SURGERY, 2015, 2
[5]   Patient-specific distal radius locking plate for fixation and accurate 3D positioning in corrective osteotomy [J].
Dobbe, J. G. G. ;
Vroemen, J. C. ;
Strackee, S. D. ;
Streekstra, G. J. .
STRATEGIES IN TRAUMA AND LIMB RECONSTRUCTION, 2014, 9 (03) :179-183
[6]   Patient-tailored plate for bone fixation and accurate 3D positioning in corrective osteotomy [J].
Dobbe, J. G. G. ;
Vroemen, J. C. ;
Strackee, S. D. ;
Streekstra, G. J. .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2013, 51 (1-2) :19-27
[7]   Computer-Assisted Planning and Navigation for Corrective Distal Radius Osteotomy, Based on Pre- and Intraoperative Imaging [J].
Dobbe, J. G. G. ;
Strackee, S. D. ;
Schreurs, A. W. ;
Jonges, R. ;
Carelsen, B. ;
Vroemen, J. C. ;
Grimbergen, C. A. ;
Streekstra, G. J. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (01) :182-190
[8]  
Hao-Yuan Kuo, 2014, 2014 IEEE International Conference on Automation Science and Engineering (CASE), P1264, DOI 10.1109/CoASE.2014.6899489
[9]   Corrective limb osteotomy using patient specific 3D-printed guides: A technical note [J].
Hoekstra, H. ;
Rosseels, W. ;
Sermon, A. ;
Nijs, S. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (10) :2375-2380
[10]   A Simple 3-Dimensional Printed Aid for a Corrective Palmar Opening Wedge Osteotomy of the Distal Radius [J].
Honigmann, Philipp ;
Thieringer, Florian ;
Steiger, Regula ;
Haefeli, Mathias ;
Schumacher, Ralf ;
Henning, Julia .