Long-Term Clinical Experience With Amplatzer Ductal Occluder II for Closure of the Persistent Arterial Duct in Children

被引:25
作者
Kang, Sok-Leng
Morgan, Gareth
Forsey, Jonathan
Tometzki, Andrew
Martin, Robin
机构
[1] Univ Hosp Bristol, NHS Fdn Trust, Bristol Royal Hosp Children, Bristol Congenital Heart Ctr, Bristol, Avon, England
[2] Univ Hosp Bristol, NHS Fdn Trust, Bristol Royal Infirm, Bristol, Avon, England
关键词
transcatheter; device; patent ductus arteriosus; TRANSCATHETER CLOSURE; OCCLUSION; COILS; MULTICENTER; INFANTS;
D O I
10.1002/ccd.25393
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ObjectivesTo describe the long-term clinical experience and follow-up with the Amplatzer Ductal Occluder II (ADO II) in children. MethodsAll patients undergoing attempted transcatheter closure of patent arterial duct (PDA) with the ADO II were included. Data collected included demographic, clinical, and echocardiographic parameters. ResultsFrom March 2008 until March 2013, 62 patients with a median age of 1.2 years (range 0.43-11.1 years) and median weight of 9 kg (range 4.7-31.4 kg) underwent the procedure. The median measurement for minimal ductal diameter was 2.7 mm (range 1.3-5 mm). An ADO II was implanted in 60 patients (96.8%). Two patients had significant residual shunting following deployment of the ADO II and underwent closure with the Amplatzer ductal occluder I (ADO I) during the same procedure. In six patients, the initial ADO II was unsatisfactory, and after recapture a different size ADO II was deployed. Device embolization of the ADO II to the pulmonary artery occurred in 6.7% of patients. Of these, one underwent surgical closure and three were closed with an ADO I. Complete occlusion on echocardiography was noted prior to discharge in 87.5% of the deployed occluders and 100% at first follow-up. Five year follow-up (n=25) revealed a 100% occlusion rate. There were three cases of persistent mild left pulmonary artery stenosis at long-term follow-up. ConclusionsThe ADO II is effective for occlusion of PDA with variable anatomy from either arterial or venous approaches with a low profile delivery system. Stable occluder position is highly dependent on accurate device sizing, good quality imaging to visualize device configuration after deployment and operator experience. (c) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:1102 / 1108
页数:7
相关论文
共 17 条
[1]  
[Anonymous], GER MED MON
[2]   Clinical Evaluation of the New Amplatzer Duct Occluder II for Patent Arterial Duct Occlusion [J].
Bhole, V. ;
Miller, P. ;
Mehta, C. ;
Stumper, O. ;
Reinhardt, Z. ;
De Giovanni, J. V. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 74 (05) :762-769
[3]   The Amplatzer duct occluder: Experience in 209 patients [J].
Bilkis, AA ;
Alwi, M ;
Hasri, S ;
Haifa, AL ;
Geetha, K ;
Rehman, MA ;
Hasanah, I .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (01) :258-261
[4]   Early Clinical Experience With the New Amplatzer Ductal Occluder II for Closure of the Persistent Arterial Duct [J].
Forsey, Jonathan ;
Kenny, Damien ;
Morgan, Gareth ;
Hayes, Alison ;
Turner, Mark ;
Tometzki, Andrew ;
Martin, Robin .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 74 (04) :615-623
[5]   RESULTS OF ANTEROGRADE TRANSCATHETER CLOSURE OF PATENT DUCTUS-ARTERIOSUS USING SINGLE OR MULTIPLE GIANTURCO COILS [J].
HIJAZI, ZM ;
GEGGEL, RL .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 74 (09) :925-929
[6]   Early Clinical Experience With a Modified Amplatzer Ductal Occluder for Transcatheter Arterial Duct Occlusion in Infants and Small Children [J].
Kenny, Damien ;
Morgan, Gareth J. ;
Bentham, James R. ;
Wilson, Neil ;
Martin, Robin ;
Tometzki, Andrew ;
Oslizlok, Paul ;
Walsh, Kevin P. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 82 (04) :534-540
[7]   ANGIOGRAPHIC CLASSIFICATION OF THE ISOLATED, PERSISTENTLY PATENT DUCTUS-ARTERIOSUS AND IMPLICATIONS FOR PERCUTANEOUS CATHETER OCCLUSION [J].
KRICHENKO, A ;
BENSON, LN ;
BURROWS, P ;
MOES, CAF ;
MCLAUGHLIN, P ;
FREEDOM, RM .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 63 (12) :877-880
[8]   Comparison of the Results of Transcatheter Closure of Patent Ductus Arteriosus with Newer Amplatzer Devices [J].
Liddy, Stephen ;
Oslizlok, Paul ;
Walsh, Kevin P. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 82 (02) :253-259
[9]   PERCUTANEOUS CLOSURE OF THE SMALL PATENT DUCTUS-ARTERIOSUS USING OCCLUDING SPRING COILS [J].
MOORE, JW ;
GEORGE, L ;
KIRKPATRICK, SE ;
MATHEWSON, JW ;
SPICER, RL ;
UZARK, K ;
ROTHMAN, A ;
CAMBIER, PA ;
SLACK, MC ;
KIRBY, WC .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (03) :759-765
[10]   Transcatheter Closure of Long Tubular Patent Arterial Ducts: The Amplatzer Duct Occluder II - A New and Valuable Tool [J].
Morgan, Gareth ;
Tornetzki, Andrew J. ;
Martin, Robin P. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 73 (04) :576-580