COMPARISON OF GLUCOSE POINT-OF-CARE VALUES WITH LABORATORY VALUES IN CRITICALLY ILL PATIENTS

被引:25
|
作者
Shearer, Anna [1 ]
Boehmer, Marilyn [1 ]
Closs, Melanie [1 ]
Dela Rosa, Rosalina [1 ]
Hamilton, Jean [1 ]
Horton, Karen [1 ]
McGrath, Rose [1 ]
Schulman, Christine
机构
[1] Providence St Vincents Med Ctr, Intens Care Unit, Portland, OR 97225 USA
关键词
INTENSIVE INSULIN THERAPY; ARTERIAL-BLOOD; METERS; HEMATOCRIT; CAPILLARY; ACCURACY; REFLECTANCE; AGREEMENT; MONITORS; PH;
D O I
10.4037/ajcc2009448
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background Blood from central venous or arterial catheters as well as from fingersticks is often used for point-of-care glucose testing. Objectives To compare glucose values obtained with a point-of-care device for catheter and fingerstick blood samples with values obtained by clinical laboratory analysis of a catheter blood sample. Methods A method-comparison design was used. Point-of-care values for fingerstick and catheter blood samples were compared with laboratory values for blood from catheters in a convenience sample of 63 critically ill patients. Device bias and precision were calculated and graphed according to the Bland-Altman method, and a t test was used to determine differences in glucose values for the 2 methods. Results Laboratory glucose values for blood from a catheter sample differed significantly from point-of-care values for blood from a fingerstick (t(1,61) = 5.01; P < .001) and from a catheter (t(1,61) = 3.91; P < .001). Bias and precision for the point-of-care device was large (fingerstick, 8.7 and 13.7; catheter, 7.0 and 14.0); 20% of the values differed from the laboratory glucose values by more than 20 mg/dL. Point-of-care glucose values for fingerstick and catheter samples did not differ (P = .41). Conclusions Glucose values obtained with a point-of-care device differ significantly from those obtained by laboratory analysis. The magnitude of these differences calls into question the widespread practice of using point-of-care glucose testing to guide insulin titration for tight glucose control. Errors in dosing could easily be made because of the large bias and precision associated with a point-of-care device. (American Journal of Critical Care. 2009; 18: 224-230)
引用
收藏
页码:224 / 230
页数:7
相关论文
共 50 条
  • [31] Comparison of a Point-of-Care Glucometer and a Laboratory Autoanalyzer for Measurement of Blood Glucose Concentrations in Domestic Pigeons (Columba livia domestica)
    Mohsenzadeh, Mahdieh Sadat
    Zaeemi, Mahdieh
    Razmyar, Jamshid
    Azizzadeh, Mohammad
    JOURNAL OF AVIAN MEDICINE AND SURGERY, 2015, 29 (03) : 181 - 186
  • [32] Reliability of Central Venous Blood Gas Values Compared With Arterial Blood Gas Values in Critically Ill Patients
    Hyun, Dong-Gon
    Ji, Wonjun
    Ahn, Jee Hwan
    Oh, Dong Kyu
    Huh, Jin Won
    Hong, Sang-Bum
    Koh, Younsuck
    Lim, Chae-Man
    RESPIRATORY CARE, 2022, 67 (07) : 863 - 870
  • [33] Comparison of Point-of-Care Testing Glucose Results from Intensive Care Patients Measured with Network-Ready Devices
    Hopf, Susanne
    Graf, Bernhard
    Gruber, Michael
    DIABETES TECHNOLOGY & THERAPEUTICS, 2011, 13 (10) : 1047 - 1056
  • [34] Evaluation of a point-of-care blood glucose monitor in healthy goats
    Quandt, Jane E.
    Barletta, Michele
    Cornell, Karen K.
    Giguere, Steeve
    Hofmeister, Erik H.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2018, 28 (01) : 45 - 53
  • [35] The utility of point-of-care urinary lipoarabinomannan testing for the diagnosis of tuberculosis in critically ill patients: a prospective observational study
    de Vasconcellos, Kim
    Ramjathan, Praksha
    Singh, Dhivendra
    BMC INFECTIOUS DISEASES, 2021, 21 (01)
  • [36] Agreement between point-of-care glucometry, blood gas and laboratory-based measurement of glucose in an equine neonatal intensive care unit
    Russell, C.
    Palmer, J. E.
    Boston, R. C.
    Wilkins, P. A.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2007, 17 (03) : 236 - 242
  • [37] Comparison of point-of-care versus central laboratory measurement of hematocrit, hemoglobin, and electrolyte concentrations
    Gavala, Alexandra
    Myrianthefs, Pavios
    HEART & LUNG, 2017, 46 (04): : 246 - 250
  • [38] The incidence, characteristics, outcomes and associations of small short-term point-of-care creatinine increases in critically ill patients
    Toh, Lisa
    Bitker, Laurent
    Eastwood, Glenn M.
    Bellomo, Rinaldo
    JOURNAL OF CRITICAL CARE, 2019, 52 : 227 - 232
  • [39] Comparison of three point-of-care blood glucose meters for use in adult and juvenile alpacas
    Tennent-Brown, Brett S.
    Koenig, Amie
    Williamson, Lisa H.
    Boston, Raymond C.
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2011, 239 (03): : 380 - 386
  • [40] Inaccurate point-of-care blood glucose measurement in a dog with secondary erythrocytosis
    McQuinn, Erin R.
    Viall, Austin K.
    Hirschfield, Margaret A.
    Ward, Jessica L.
    Jeffery, Unity
    LeVine, Dana N.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2020, 30 (01) : 81 - 85