COMPARISON OF GLUCOSE POINT-OF-CARE VALUES WITH LABORATORY VALUES IN CRITICALLY ILL PATIENTS

被引:25
|
作者
Shearer, Anna [1 ]
Boehmer, Marilyn [1 ]
Closs, Melanie [1 ]
Dela Rosa, Rosalina [1 ]
Hamilton, Jean [1 ]
Horton, Karen [1 ]
McGrath, Rose [1 ]
Schulman, Christine
机构
[1] Providence St Vincents Med Ctr, Intens Care Unit, Portland, OR 97225 USA
关键词
INTENSIVE INSULIN THERAPY; ARTERIAL-BLOOD; METERS; HEMATOCRIT; CAPILLARY; ACCURACY; REFLECTANCE; AGREEMENT; MONITORS; PH;
D O I
10.4037/ajcc2009448
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background Blood from central venous or arterial catheters as well as from fingersticks is often used for point-of-care glucose testing. Objectives To compare glucose values obtained with a point-of-care device for catheter and fingerstick blood samples with values obtained by clinical laboratory analysis of a catheter blood sample. Methods A method-comparison design was used. Point-of-care values for fingerstick and catheter blood samples were compared with laboratory values for blood from catheters in a convenience sample of 63 critically ill patients. Device bias and precision were calculated and graphed according to the Bland-Altman method, and a t test was used to determine differences in glucose values for the 2 methods. Results Laboratory glucose values for blood from a catheter sample differed significantly from point-of-care values for blood from a fingerstick (t(1,61) = 5.01; P < .001) and from a catheter (t(1,61) = 3.91; P < .001). Bias and precision for the point-of-care device was large (fingerstick, 8.7 and 13.7; catheter, 7.0 and 14.0); 20% of the values differed from the laboratory glucose values by more than 20 mg/dL. Point-of-care glucose values for fingerstick and catheter samples did not differ (P = .41). Conclusions Glucose values obtained with a point-of-care device differ significantly from those obtained by laboratory analysis. The magnitude of these differences calls into question the widespread practice of using point-of-care glucose testing to guide insulin titration for tight glucose control. Errors in dosing could easily be made because of the large bias and precision associated with a point-of-care device. (American Journal of Critical Care. 2009; 18: 224-230)
引用
收藏
页码:224 / 230
页数:7
相关论文
共 50 条
  • [1] DIFFERENCES IN GLUCOSE VALUES OBTAINED FROM POINT-OF-CARE GLUCOSE METERS AND LABORATORY ANALYSIS IN CRITICALLY ILL PATIENTS
    Cook, Anjannette
    Laughlin, Delyla
    Moore, Margery
    North, Doreen
    Wilkins, Kathleen
    Wong, Gay
    Wallace-Scroggs, Allyson
    Halvorsen, Lisa
    AMERICAN JOURNAL OF CRITICAL CARE, 2009, 18 (01) : 65 - 71
  • [2] Comparison of point-of-care and laboratory glucose analysis in critically ill patients
    Lacara, Teresita
    Domagtoy, Caroline
    Lickliter, Donna
    Quattrocchi, Kathy
    Snipes, Lydia
    Kuszaj, Joanne
    Prasnikar, MaryClare
    AMERICAN JOURNAL OF CRITICAL CARE, 2007, 16 (04) : 336 - 346
  • [3] The effects of precision, haematocrit, pH and oxygen tension on point-of-care glucose measurement in critically ill patients: a prospective study
    Watkinson, Peter James
    Barber, Vicki Suzanne
    Amira, Esther
    James, Timothy
    Taylor, Richard
    Young, John Duncan
    ANNALS OF CLINICAL BIOCHEMISTRY, 2012, 49 : 144 - 151
  • [4] Inaccuracy of Venous Point-of-Care Glucose Measurements in Critically Ill Patients: A Cross-Sectional Study
    Pereira, Adriano Jose
    Correa, Thiago Domingos
    de Almeida, Francisca Pereira
    Deliberato, Rodrigo Octavio
    Lobato, Michelle dos Santos
    Akamine, Nelson
    Silva, Eliezer
    Cavalcanti, Alexandre Biasi
    PLOS ONE, 2015, 10 (06):
  • [5] Point-of-care blood glucose measurement errors overestimate hypoglycaemia rates in critically ill patients
    Nya-Ngatchou, Jean-Jacques
    Corl, Dawn
    Onstad, Susan
    Yin, Tom
    Tylee, Tracy
    Suhr, Louise
    Thompson, Rachel E.
    Wisse, Brent E.
    DIABETES-METABOLISM RESEARCH AND REVIEWS, 2015, 31 (02) : 147 - 154
  • [6] Comparing the accuracy of point-of-care with laboratory (capillary, venous, and arterial) blood glucose levels in critically ill patients with and without shock
    Alshaer, Abdulaziz
    Badgheish, Basma A.
    Alsadah, Zahra Hashim
    Sewify, Khalid
    Alghazal, Sarah
    Alzahrani, Sarah
    Qadi, Abeer
    Alqahtani, Reham
    Farsani, Ghadeer Abdullah
    Shilash, Amal
    BMC RESEARCH NOTES, 2022, 15 (01)
  • [7] Precision and Accuracy Comparison of Point-of-Care and Laboratory Glucose Concentrations in Cardiothoracic Surgery Patients
    Denfeld, Quin E.
    Goodell, Teresa T.
    Stafford, Kelly N.
    Kazmierczak, Steve
    JOURNAL OF CARDIOVASCULAR NURSING, 2011, 26 (06) : 512 - 518
  • [8] Point-of-care creatinine vs. central laboratory creatinine in the critically ill
    White, Kyle C.
    McCullough, James
    Shekar, Kiran
    Senthuran, Siva
    Laupland, Kevin B.
    Dimeski, Goce
    Serpa-Neto, Ary
    Bellomo, Rinaldo
    CRITICAL CARE AND RESUSCITATION, 2024, 26 (03) : 198 - 203
  • [9] Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients
    Allardet-Servent, Jerome
    Lebsir, Melissa
    Dubroca, Christian
    Fabrigoule, Martine
    Jordana, Sylvie
    Signouret, Thomas
    Castanier, Matthias
    Thomas, Guillemette
    Soundaravelou, Rettinavelou
    Lepidi, Anne
    Delapierre, Laurence
    Penaranda, Guillaume
    Halfon, Philippe
    Seghboyan, Jean-Marie
    PLOS ONE, 2017, 12 (01):
  • [10] Point-of-care assessment of microvascular blood flow in critically ill patients
    Arnold, Ryan C.
    Parrillo, Joseph E.
    Dellinger, R. Phillip
    Chansky, Michael E.
    Shapiro, Nathan I.
    Lundy, David J.
    Trzeciak, Stephen
    Hollenberg, Steven M.
    INTENSIVE CARE MEDICINE, 2009, 35 (10) : 1761 - 1766