The Impact of Aortic/Subclavian Outflow Cannulation for Cardiopulmonary Bypass and Cardiac Support: A Computational Fluid Dynamics Study

被引:32
作者
Kaufmann, Tim A. S. [1 ]
Hormes, Marcus [1 ]
Laumen, Marco [1 ]
Timms, Daniel L. [1 ]
Linde, Torsten [1 ]
Schmitz-Rode, Thomas [1 ]
Moritz, Anton [2 ]
Dzemali, Omer [2 ]
Steinseifer, Ulrich [1 ]
机构
[1] RWTHAachen Univ, Helmholtz Inst, Aachen, Germany
[2] JW Goethe Univ Hosp, Dept Thorac & Cardiovasc Surg, Frankfurt, Germany
关键词
Computational fluid dynamics; Flow distribution; Cardiopulmonary bypass; Oxygen deficiency; Cardiac support; Ventricular assist devices; Outflow cannula; Outflow cannula positioning; AORTIC-ARCH; NONPULSATILE PERFUSION; BLOOD-FLOW; PULSATILE; STROKE;
D O I
10.1111/j.1525-1594.2009.00848.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Approximately 100 000 cases of oxygen deficiency in the brain occur during cardiopulmonary bypass (CPB) procedures each year. In particular, perfusion of the carotid and vertebral arteries is affected. The position of the outflow cannula influences the blood flow to the cardiovascular system and thus end organ perfusion. Traditionally, the cannula returns blood into the ascending aorta. But some surgeons prefer cannulation to the right subclavian artery. A computational fluid dynamics study was initially undertaken for both approaches. The vessel model was created from real computed tomography/magnetic resonance imaging data of young healthy patients. The simulations were run with usual CPB conditions. The flow distribution for different cannula positions in the aorta was studied, as well as the impact of the cannula tip distance to vertebral artery for the subclavian position. The study presents a fast method of analyzing the flow distribution in the cardiovascular system, and can be adapted for other applications such as ventricular assist device support. It revealed that two effects cause the loss of perfusion seen clinically: a vortex under the brachiocephalic trunk and low pressure regions near the cannula jet. The results suggest that cannulation to the subclavian artery is preferred if the cannula tip is sufficiently far away from the branch of the vertebral artery. For the aortic positions, however, the cannula should be injected from the left body side.
引用
收藏
页码:727 / 732
页数:6
相关论文
共 22 条
[1]   Is there any impact of the shape of aortic end-hole cannula on stroke occurrence? Clinical evaluation of straight and bent-tip aortic cannulae [J].
Albert, AA ;
Beller, CJ ;
Arnrich, B ;
Walter, JA ;
Rosendahl, UP ;
Hetzel, A ;
Priss, H ;
Ennker, J .
PERFUSION-UK, 2002, 17 (06) :451-456
[2]   Aortic atheroma burden and cognitive dysfunction after coronary artery bypass graft surgery [J].
Bar-Yosef, S ;
Anders, M ;
Mackensen, B ;
Ti, LK ;
Mathew, JP ;
Phillips-Bute, B ;
Messier, RH ;
Grocott, HP .
ANNALS OF THORACIC SURGERY, 2004, 78 (05) :1556-1563
[3]  
BLAUTH CI, 1988, J THORAC CARDIOV SUR, V95, P668
[4]   The effects of pulsatile and nonpulsatile systemic perfusion on renal sympathetic nerve activity in anesthetized dogs [J].
Fukae, K ;
Tominaga, R ;
Tokunaga, S ;
Kawachi, Y ;
Imaizumi, T ;
Yasui, H .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1996, 111 (02) :478-484
[5]   Hydrodynamics of aortic arch vessels during perfusion through the right subclavian artery [J].
Gerdes, A ;
Joubert-Hübner, E ;
Esders, K ;
Sievers, HH .
ANNALS OF THORACIC SURGERY, 2000, 69 (05) :1425-1430
[6]   EFFECT OF CANNULA LENGTH ON AORTIC-ARCH FLOW - PROTECTION OF THE ATHEROMATOUS AORTIC-ARCH [J].
GROSSI, EA ;
KANCHUGER, MS ;
SCHWARTZ, DS ;
MCLOUGHLIN, DE ;
LEBOUTILLIER, M ;
RIBAKOVE, GH ;
MARSCHALL, KE ;
GALLOWAY, AC ;
COLVIN, SB .
ANNALS OF THORACIC SURGERY, 1995, 59 (03) :710-712
[7]   Pulsatile and nonpulsatile perfusion: The continuing controversy [J].
Hornick, P ;
Taylor, K .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1997, 11 (03) :310-315
[8]   An evaluation of the benefits of pulsatile versus nonpulsatile perfusion during cardiopulmonary bypass procedures in pediatric and adult cardiac patients [J].
Ji, Bingyang ;
Uendar, Akif .
ASAIO JOURNAL, 2006, 52 (04) :357-361
[9]   Non-Newtonian blood flow in human right coronary arteries: steady state simulations [J].
Johnston, BM ;
Johnston, PR ;
Corney, S ;
Kilpatrick, D .
JOURNAL OF BIOMECHANICS, 2004, 37 (05) :709-720
[10]   An in vitro evaluation of a new cannula tip design compared with two clinically established cannula-tip designs regarding aortic arch vessel perfusion characteristics [J].
Joubert-Huebner, E ;
Gerdes, A ;
Sievers, HH .
PERFUSION-UK, 2000, 15 (01) :69-76