MEKC-MS/MS method using a volatile surfactant for the simultaneous determination of 12 synthetic cannabinoids

被引:25
作者
Akamatsu, Shigeki [1 ]
Mitsuhashi, Takao [1 ]
机构
[1] Hyogo Prefectural Inst Publ Hlth & Consumer Sci, Kobe, Hyogo 6520032, Japan
关键词
CE; Herbal incense blends; Micellar electrokinetic chromatography; MS/MS; Synthetic cannabinoids; MICELLAR ELECTROKINETIC CHROMATOGRAPHY; MASS-SPECTROMETRY; HAIR;
D O I
10.1002/jssc.201301132
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study describes a method for the simultaneous determination of 12 synthetic cannabinoids by MEKC-MS/MS using a volatile surfactant (ammonium perfluorooctanoate) as a constituent of the micellar pseudostationary phase. Although most synthetic cannabinoids comigrated by a CZE method, sufficient separation could be achieved by the proposed method. The best separation was made possible by 50 mM ammonium perfluorooctanoate in 20% v/v acetonitrile/water (apparent pH(star) 9.0) as the BGE, followed by MS detection using a sheath liquid composed of 5 mM ammonium formate in 50% v/v methanol/water mixed hydro-organic solvent. The standard calibration curve for all analytes showed good linearity (r > 0.99). Satisfactory recoveries, ranging from 89.5 to 101.7%, were obtained. The LODs were 6.5-76.5 mu g/g for the target analytes. This method appears to be a useful tool for the identification of synthetic cannabinoids in illegal herbal incense blends.
引用
收藏
页码:304 / 307
页数:4
相关论文
共 23 条
[21]   ELECTROKINETIC SEPARATIONS WITH MICELLAR SOLUTIONS AND OPEN-TUBULAR CAPILLARIES [J].
TERABE, S ;
OTSUKA, K ;
ICHIKAWA, K ;
TSUCHIYA, A ;
ANDO, T .
ANALYTICAL CHEMISTRY, 1984, 56 (01) :111-113
[22]   Ammonium perfluorooctanoate as a volatile surfactant for the analysis of N-methylcarbamates by MEKC-ESI-MS [J].
Van Biesen, Geert ;
Bottaro, Christina S. .
ELECTROPHORESIS, 2006, 27 (22) :4456-4468
[23]   Qualitative Confirmation of 9 Synthetic Cannabinoids and 20 Metabolites in Human Urine Using LC-MS/MS and Library Search [J].
Wohlfarth, Ariane ;
Scheidweiler, Karl B. ;
Chen, Xiaohong ;
Liu, Hua-fen ;
Huestis, Marilyn A. .
ANALYTICAL CHEMISTRY, 2013, 85 (07) :3730-3738