A Measurement System to Estimate the Pleural Pressure From the CVP for Respiratory System Monitoring

被引:6
作者
Addabbo, Tommaso [1 ]
Cataldo, Gennaro [2 ]
Cevenini, Gabriele [2 ]
Fort, Ada [1 ]
Franchi, Federico [2 ]
Moretti, Riccardo [1 ]
Mugnaini, Marco [1 ]
Scolletta, Sabino [3 ]
Vignoli, Valerio [1 ]
机构
[1] Univ Siena, Dept Informat Engn & Math, I-53100 Siena, Italy
[2] Univ Siena, Dept Med Biotechnol, I-53100 Siena, Italy
[3] Univ Siena, Dept Med Surg & Neurosci, I-53100 Siena, Italy
关键词
Biomedical electronics; biomedical measurement; biomedical monitoring; biomedical signal processing; ESOPHAGEAL PRESSURE; TRANSPULMONARY PRESSURE; MECHANICAL VENTILATION; DISTRESS-SYNDROME; PARAMETERS; CARE;
D O I
10.1109/TIM.2019.2909995
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We present a technique to estimate the pleural pressure combining the information provided by the central venous pressure (CVP) and electrocardiographic signals (ECG) for respiratory system monitoring. We discuss the characterization of the proposed measurement method with theoretical modeling, numerical analysis, and experiments based on signals acquired from intensive care unit patients. The results confirm the validity of the proposal suitable to define a novel method for the continuous monitoring of the respiratory mechanics. Since mechanically ventilated patients are supposed to have a central venous catheter in position, the CVP filtering technique can results in an useful clinical approach to estimate the pleural pressure, avoiding additional invasive devices (i.e., balloon-tipped esophageal catheters).
引用
收藏
页码:2469 / 2478
页数:10
相关论文
共 33 条
[1]  
Addabbo T, 2018, 2018 IEEE SENSORS APPLICATIONS SYMPOSIUM (SAS), P6, DOI 10.1109/SAS.2018.8336709
[2]   The Application of Esophageal Pressure Measurement in Patients with Respiratory Failure [J].
Akoumianaki, Evangelia ;
Maggiore, Salvatore M. ;
Valenza, Franco ;
Bellani, Giacomo ;
Jubran, Amal ;
Loring, Stephen H. ;
Pelosi, Paolo ;
Talmor, Daniel ;
Grasso, Salvatore ;
Chiumello, Davide ;
Guerin, Claude ;
Patroniti, Nicolo ;
Ranieri, V. Marco ;
Gattinoni, Luciano ;
Nava, Stefano ;
Terragni, Pietro-Paolo ;
Pesenti, Antonio ;
Tobin, Martin ;
Mancebo, Jordi ;
Brochard, Laurent .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2014, 189 (05) :520-531
[3]  
Aliverti A., 2017, ACUTE RESP DISTRESS, P225
[4]   REAL-TIME TRACKING OF PARAMETERS OF LUNG-MECHANICS - EMPHASIS ON ALGORITHM TUNING [J].
AVANZOLINI, G ;
BARBINI, P ;
CAPPELLO, A ;
CEVENINI, G .
JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (06) :489-495
[5]   Central Venous Pressure as Popular Resuscitation Surrogate Not Totally Unjustified Response [J].
Baram, Michael ;
Marik, Paul E. .
CHEST, 2008, 134 (06) :1352-1353
[6]   REAL-TIME TRACKING OF BREATHING PARAMETERS IN MECHANICALLY VENTILATED DOGS [J].
BARBINI, P ;
CAPPELLO, A ;
CEVENINI, G .
PROCEEDINGS OF THE ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, PTS 1-4, 1988, :700-702
[7]   ESTIMATING RESPIRATORY MECHANICAL PARAMETERS OF VENTILATED PATIENTS - A CRITICAL-STUDY IN THE ROUTINE INTENSIVE-CARE UNIT [J].
BARBINI, P ;
CEVENINI, G ;
LUTCHEN, KR ;
URSINO, M .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1994, 32 (02) :153-160
[8]  
BAYDUR A, 1982, AM REV RESPIR DIS, V126, P788
[9]   Epidemiology, Patterns of Care, and Mortality for Patients With Acute Respiratory Distress Syndrome in Intensive Care Units in 50 Countries [J].
Bellani, Giacomo ;
Laffey, John G. ;
Pham, Tai ;
Fan, Eddy ;
Brochard, Laurent ;
Esteban, Andres ;
Gattinoni, Luciano ;
van Haren, Frank ;
Larsson, Anders ;
McAuley, Daniel F. ;
Ranieri, Marco ;
Rubenfeld, Gordon ;
Thompson, B. Taylor ;
Wrigge, Hermann ;
Slutsky, Arthur S. ;
Pesenti, Antonio .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2016, 315 (08) :788-800
[10]   Extraction of Breathing Signal and Suppression of Its Effects in Oscillometric Blood Pressure Measurement [J].
Chen, Silu ;
Bolic, Miodrag ;
Groza, Voicu Z. ;
Dajani, Hilmi R. ;
Batkin, Izmail ;
Rajan, Sreeraman .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (05) :1741-1750