Sensitivity evaluation of two commercial quality assurance systems to organ-dose variations of patient-specific VMAT plans

被引:4
作者
Oderinde, Oluwaseyi M. [1 ]
Du Plessis, Freek [1 ]
机构
[1] Univ Free State, Dept Med Phys, POB 339, ZA-9300 Bloemfontein, South Africa
基金
英国医学研究理事会;
关键词
Sensitivity assessment; dose variation; IQM; MatriXX; radiotherapy verification; IONIZATION-CHAMBER; TRANSMISSION DETECTOR; MONITORING-SYSTEM; DIODE-ARRAY; IMRT; VERIFICATION; MLC; MATRIXX; QA;
D O I
10.1080/16878507.2019.1618080
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The purpose of this study was to assess the dose variation sensitivity of two quality assurance (QA) devices (Integral quality monitoring (IQM (R)) and MatriXXEvolution systems) used for radiotherapy verification. Six volumetric-modulated arc therapy (VMAT) radiation plans were calculated, namely, three head-and-neck and three prostate cases. For sensitivity evaluation, the planning target volume (PTV) dose for each patient's plan was modified by 0.5%, 1.0%, 2.0%, and 3.0% of its original dose. The IQM and MatriXX detectors are sensitive to the dose errors considered. At 0.5% PTV dose modification, the average local percentage differences for the IQM are 0.27 +/- 0.29, 0.24 +/- 0.35, 0.42 +/- 0.39, 0.74 +/- 0.28, 0.41 +/- 0.24, and 0.26 +/- 0.32, while the average local percentage differences for the MatriXX device are 1.37 +/- 0.25, 1.30 +/- 0.75, 2.82 +/- 1.46, 1.34 +/- 1.29, 1.58 +/- 0.97, and 1.13 +/- 0.97. The sensitivity of the detectors is more pronounced in VMAT plan errors containing larger segments. This shows that the sensitivities of the detectors are plan and fraction specific. Both detectors are sensitive to dose variation in the clinical plans to a minimal dose deviation of 0.5%. The IQM detector shows the capability to be used for QA procedures and for real-time beam output monitoring.
引用
收藏
页码:132 / 139
页数:8
相关论文
共 34 条
[1]   Sensitivity of a helical diode array dosimeter to Volumetric Modulated Arc Therapy delivery errors [J].
Arumugam, Sankar ;
Xing, Aitang ;
Young, Tony ;
Holloway, Lois .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (08) :1043-1054
[2]  
Bawazeer O., 2014, J PHYS C SERIES, V489, P1
[3]   Influence of the Integral Quality Monitor transmission detector on high energy photon beams: A multi-centre study [J].
Casar, Bozidar ;
Pasler, Marlies ;
Wegener, Sonja ;
Hoffman, David ;
Talamonti, Cinzia ;
Qian, Jianguo ;
Mendez, Ignasi ;
Brojan, Denis ;
Perrin, Bruce ;
Kusters, Martijn ;
Canters, Richard ;
Pallotta, Stefania ;
Peterlini, Primoz .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (03) :232-242
[4]   Patient QA systems for rotational radiation therapy: A comparative experimental study with intentional errors [J].
Fredh, Anna ;
Scherman, Jonas Bengtsson ;
Fog, Lotte S. ;
Af Rosenschold, Per Munck .
MEDICAL PHYSICS, 2013, 40 (03)
[5]   Investigation of a real-time EPID-based patient dose monitoring safety system using site-specific control limits [J].
Fuangrod, Todsaporn ;
Greer, Peter B. ;
Woodruff, Henry C. ;
Simpson, John ;
Bhatia, Shashank ;
Zwan, Benjamin ;
vanBeek, Timothy A. ;
McCurdy, Boyd M. C. ;
Middleton, Richard H. .
RADIATION ONCOLOGY, 2016, 11
[6]   Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external beam radiation therapy [J].
Hoffman, David ;
Chung, Eunah ;
Hess, Clayton ;
Stern, Robin ;
Benedict, Stanley .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (01) :40-48
[7]   A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems [J].
Hussein, Mohammad ;
Rowshanfarzad, Pejman ;
Ebert, Martin A. ;
Nisbet, Andrew ;
Clark, Catharine H. .
RADIOTHERAPY AND ONCOLOGY, 2013, 109 (03) :370-376
[8]   Study on the Sensitivity of Gamma-Index Evaluation Methods to Positioning Errors of High-Definition MLC of True Beam STx in VMAT Quality Assurance for SBRT [J].
Hwang, S. ;
Ye, S. ;
Park, J. ;
Kim, J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 :S858-S859
[9]   An integral quality monitoring system for real-time verification of intensity modulated radiation therapy [J].
Islam, Mohammad K. ;
Norrlinger, Bernhard D. ;
Smale, Jason R. ;
Heaton, Robert K. ;
Galbraith, Duncan ;
Fan, Cary ;
Jaffray, David A. .
MEDICAL PHYSICS, 2009, 36 (12) :5420-5428
[10]  
Kosaka K., 2016, INT J MED PHYS CLIN, V05, P317, DOI DOI 10.4236/IJMPCERO.2016.54031