Temperature rises in the round window caused by various light sources during insertion of rigid endoscopes: an experimental animal study

被引:11
作者
Dundar, R. [1 ]
Bulut, H. [2 ]
Yukkaldiran, A. [1 ]
Guler, O. K. [1 ]
Demirtas, Y. [2 ]
Iynen, I. [1 ]
Bozkus, F. [1 ]
Soy, F. K. [1 ]
机构
[1] Harran Univ, Dept Otorhinolaryngol, TR-414200 Sanliurfa, Turkey
[2] Harran Univ, Fac Engn, Dept Mech Engn, TR-414200 Sanliurfa, Turkey
关键词
D O I
10.1111/coa.12468
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
ObjectivesThe instruments used in endoscopic surgery include rigid endoscopes of different diameters and angles, sources of light and monitors. In this study, we explored temperature rises in the round window caused by insertion of rigid endoscopes of different diameters into the middle ear; the endoscopes were fitted with different light sources. DesignAn experimental animal study. SettingWe subjected seven guinea pigs to simulate otological surgery at room temperature. We monitored body temperatures, along with temperature rises caused by the use of 0 degrees rigid endoscopes of diameters 3 or 4mm, fitted with light sources including a halogen light, a light-emitting diode (LED) and a xenon light. Data were collected every second from a sensor placed in the round window. ParticipantsAn experimental study on 7 guinea pigs. Main Outcome MeasuresRise of the temperature on round window. ResultsRigid endoscopes caused the temperature of the tympanum to rise when xenon and halogen light sources were used, regardless of endoscope diameter. However, the temperature rise was less when a LED light source was employed. ConclusionThe endoscopic instruments used in middle ear surgery caused the temperature of the round window to rise. The rise varied with endoscope diameter and the type of light source used.
引用
收藏
页码:44 / 50
页数:7
相关论文
共 16 条
[1]  
[Anonymous], 2000, JAMA, V284, P3043
[2]  
Aoki K, 2001, Diagn Ther Endosc, V7, P99, DOI 10.1155/DTE.7.99
[3]   Instrumentation and Technologies in Endoscopic Ear Surgery [J].
Badr-El-Dine, Mohamed ;
James, Adrian L. ;
Panetti, Giuseppe ;
Marchioni, Daniele ;
Presutti, Livio ;
Nogueira, Joao Flavio .
OTOLARYNGOLOGIC CLINICS OF NORTH AMERICA, 2013, 46 (02) :211-+
[4]   In vitro and in vivo determination of the thermal effect of middle ear endoscopy [J].
Bottrill, I ;
Perrault, DF ;
Poe, D .
LARYNGOSCOPE, 1996, 106 (02) :213-216
[5]   NOISE-INDUCED REDUCTION OF INNER-EAR MICROPHONIC RESPONSE - DEPENDENCE ON BODY-TEMPERATURE [J].
DRESCHER, DG .
SCIENCE, 1974, 185 (4147) :273-274
[6]  
EATOCK RA, 1981, J COMP PHYSIOL, V142, P219, DOI 10.1007/BF00605740
[7]   Influence of hyperthermia on cochlear micromechanical properties in humans [J].
FerberViart, C ;
Savourey, G ;
Garcia, C ;
Duclaux, R ;
Bittel, J ;
Collet, L .
HEARING RESEARCH, 1995, 91 (1-2) :202-207
[8]  
Glasscock ME, 2003, SURG EAR, P350
[9]   A comparative study of endoscope assisted myringoplasty and micrsoscope assisted myringoplasty [J].
Harugop, A. S. ;
Mudhol, R. S. ;
Godhi, R. A. .
INDIAN JOURNAL OF OTOLARYNGOLOGY AND HEAD & NECK SURGERY, 2008, 60 (04) :298-302
[10]   Total radiated power, infrared output, and heat generation by cold light sources at the distal end of endoscopes and fiber optic bundle of light cables [J].
Hensman, C ;
Hanna, GB ;
Drew, T ;
Moseley, H ;
Cuschieri, A .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1998, 12 (04) :335-337