Improving automatic analysis of the electrocardiogram acquired during magnetic resonance imaging using magnetic field gradient artefact suppression

被引:9
作者
Abaecherli, Roger [1 ]
Hornaff, Sven
Leber, Remo
Schmid, Hans-Jakob
Felblinger, Jacques
机构
[1] SCHILLER AG, CH-6341 Baar, Switzerland
[2] Univ Henri Poincare, INSERM, ERI 13, F-54511 Nancy, France
关键词
electrocardiogram; artefacts; magnetic field gradients; patient monitoring; MRI synchronization; rhythm analysis;
D O I
10.1016/j.jelectrocard.2006.05.028
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The electrocardiogram (ECG) used for patient monitoring during magnetic resonance imaging (MRI) unfortunately suffers from severe artefacts. These artefacts are due to the special environment of the MRI. Modeling helped in finding solutions for the suppression of these artefacts superimposed on the ECG signal. After we validated the linear and time invariant model for the magnetic field gradient artefact generation, we applied offline and online filters for their suppression. Wiener filtering (offline) helped in generating reference annotations of the ECG beats. In online filtering, the least-mean-square filter suppressed the magnetic field gradient artefacts before the acquired ECG signal was input to the arrhythmia algorithm. Comparing the results of two runs (one run using online filtering and one run without) to our reference annotations, we found an eminent improvement in the arrhythmia module's performance, enabling reliable patient monitoring and MRI synchronization based on the ECG signal. (C) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:S134 / S139
页数:6
相关论文
共 42 条
[1]   Suppression of MR gradient artefacts on electrophysiological signals based on an adaptive real-time filter with LMS coefficient updates [J].
Abächerli, R ;
Pasquier, C ;
Odille, F ;
Kraemer, M ;
Schmid, JJ ;
Felblinger, J .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2005, 18 (01) :41-50
[2]  
ABACHERLI R, 2005, RESTORATION ANAL ELE
[3]   RF-induced temperature elevation along metallic wires in clinical magnetic resonance imaging: Influence of diameter and length [J].
Armenean, C ;
Perrin, E ;
Armenean, M ;
Beuf, O ;
Pilleul, F ;
Saint-Jalmes, H .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (05) :1200-1206
[4]  
Association for the Advancement of Medical Instrumentation, 1998, ANSI/AAMI EC57
[5]   Magnetic resonance pharmacological stress for detecting coronary disease - Comparison with echocardiography [J].
Baer, FM ;
Crnac, J ;
Schmidt, M ;
Jochims, M ;
Theissen, P ;
Schneider, C ;
Schicha, H ;
Erdmann, E .
HERZ, 2000, 25 (04) :400-408
[6]   Static magnetic field effects on human subjects related to magnetic resonance imaging systems [J].
Chakeres, DW ;
de Vocht, F .
PROGRESS IN BIOPHYSICS & MOLECULAR BIOLOGY, 2005, 87 (2-3) :255-265
[7]  
Chia JM, 2000, J MAGN RESON IMAGING, V12, P678, DOI 10.1002/1522-2586(200011)12:5<678::AID-JMRI4>3.0.CO
[8]  
2-5
[9]   NEW APPROACH TO THE DIGITAL ELIMINATION OF 50 HZ INTERFERENCE FROM THE ELECTROCARDIOGRAM [J].
CHRISTOV, II ;
DOTSINSKY, IA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (04) :431-434
[10]   Dynamic powerline interference subtraction from biosignals [J].
Christov, II .