Numerical investigations on the strain-adaptive bone remodelling in the periprosthetic femur: Influence of the boundary conditions

被引:52
作者
Behrens, Bernd-Arno [1 ]
Nolte, Ingo [2 ]
Wefstaedt, Patrick [2 ]
Stukenborg-Colsman, Christina [3 ]
Bouguecha, Anas [1 ]
机构
[1] Leibniz Univ Hannover, Inst Met Forming & Met Forming Machines, Hannover, Germany
[2] Univ Vet Med Hannover, Small Anim Clin, Hannover, Germany
[3] Hannover Med Sch, Dept Orthopaed, D-3000 Hannover, Germany
关键词
HIP PROSTHESES; DENSITY; FORCES; STEMS;
D O I
10.1186/1475-925X-8-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: There are several numerical investigations on bone remodelling after total hip arthroplasty (THA) on the basis of the finite element analysis (FEA). For such computations certain boundary conditions have to be defined. The authors chose a maximum of three static load situations, usually taken from the gait cycle because this is the most frequent dynamic activity of a patient after THA. Materials and methods: The numerical study presented here investigates whether it is useful to consider only one static load situation of the gait cycle in the FE calculation of the bone remodelling. For this purpose, 5 different loading cases were examined in order to determine their influence on the change in the physiological load distribution within the femur and on the resulting strain-adaptive bone remodelling. First, four different static loading cases at 25%, 45%, 65% and 85% of the gait cycle, respectively, and then the whole gait cycle in a loading regime were examined in order to regard all the different loadings of the cycle in the simulation. Results: The computed evolution of the apparent bone density (ABD) and the calculated mass losses in the periprosthetic femur show that the simulation results are highly dependent on the chosen boundary conditions. Conclusion: These numerical investigations prove that a static load situation is insufficient for representing the whole gait cycle. This causes severe deviations in the FE calculation of the bone remodelling. However, accompanying clinical examinations are necessary to calibrate the bone adaptation law and thus to validate the FE calculations.
引用
收藏
页数:9
相关论文
共 33 条
[1]   Periprosthetic bone remodelling after cementless total hip replacement - A prospective comparison of two different implant designs [J].
Ang, KC ;
De, SD ;
Goh, JCH ;
Low, SL ;
Bose, K .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (04) :675-679
[2]   Numerical investigations of stress shielding in total hip prostheses [J].
Behrens, B-A ;
Wirth, C. J. ;
Windhagen, H. ;
Nolte, I. ;
Meyer-Lindenberg, A. ;
Bouguecha, A. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H5) :593-600
[3]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[4]   The effect of muscle loading on the simulation of bone remodelling in the proximal femur [J].
Bitsakos, C ;
Kerner, J ;
Fisher, I ;
Amis, AA .
JOURNAL OF BIOMECHANICS, 2005, 38 (01) :133-139
[5]  
BOUGUECHA A, 2007, JAHRESTAGUNG DTSCH G
[6]   COMPRESSIVE BEHAVIOR OF BONE AS A 2-PHASE POROUS STRUCTURE [J].
CARTER, DR ;
HAYES, WC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (07) :954-962
[7]   TRABECULAR BONE-DENSITY AND LOADING HISTORY - REGULATION OF CONNECTIVE-TISSUE BIOLOGY BY MECHANICAL ENERGY [J].
CARTER, DR ;
FYHRIE, DP ;
WHALEN, RT .
JOURNAL OF BIOMECHANICS, 1987, 20 (08) :785-+
[8]   Influence of muscle forces on femoral strain distribution [J].
Duda, GN ;
Heller, M ;
Albinger, J ;
Schulz, O ;
Schneider, E ;
Claes, L .
JOURNAL OF BIOMECHANICS, 1998, 31 (09) :841-846
[9]   Internal forces and moments in the femur during walking [J].
Duda, GN ;
Schneider, E ;
Chao, EYS .
JOURNAL OF BIOMECHANICS, 1997, 30 (09) :933-941
[10]  
Ebbecke B, 2005, J STRUCTURAL MECH, V38, P177