Accuracy of weighted CTDI in estimating average dose delivered to CTDI phantoms: An experimental study

被引:7
作者
Treb, Kevin [1 ]
Li, Ke [1 ,2 ]
机构
[1] Univ Wisconsin, Dept Med Phys, 1111 Highland Ave, Madison, WI 53705 USA
[2] Univ Wisconsin, Dept Radiol, 600 Highland Ave, Madison, WI 53792 USA
关键词
bowtie filter; CTDI; CT dosimetry; radiation dose; radiation output; weighted CTDI; INDEX CTDI; DOSIMETRY;
D O I
10.1002/mp.14528
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The concept of the weighted computed tomography dose index (CTDIw) was proposed in 1995 to represent the average CTDI across an axial section of a cylindrical phantom. The purpose of this work was to experimentally re-examine the validity of the underlying assumptions behind CTDIw for modern MDCT systems. Methods To enable experimental mapping of CTDI100 in the axial plane, in-house 16 and 32 cm cylindrical phantoms were fabricated to allow the pencil chamber to reach any arbitrary axial location within the phantoms. The phantoms were scanned on a clinical MDCT with five beam collimation widths, three bowtie filters, and four kV levels. To evaluate the linearity and rotational invariance assumptions implicitly made when the weighting factors of 1/3 and 2/3 in the CTDIw formula were originally derived, CTDI100 was measured at different radial and angular locations within the phantom for different collimation, bowtie, and kV combinations. The average CTDI (CTDIavg) across the axial plane was calculated from the experimental two-dimensional (2D) dose distribution and was compared with the traditional CTDIw. Results For both phantoms under all scan conditions, the axial dose distributions were found to have significant angular dependence, potentially due to the x-ray attenuation by the patient couch or the head holder. The radial dose profiles were also found to significantly deviate from linearity in many cases due to the presence of the bowtie filter. When only the 12 o'clock peripheral CTDI100 and the traditional weighting factors were used to calculate CTDIw, the average dose was overestimated in the 16 cm phantom by up to 8.4% at isocenter and up to 35.3% when the phantom was off-centered by 6 cm; in the 32 cm phantom at isocenter, the average dose was overestimated by up to 12.8%. Using an average of the four peripheral CTDI100 measurements at the 12, 3, 6, and 9 o'clock locations reduced the error of CTDIw to within 1.2% in the 16 cm phantom. For the 32 cm phantom, even by using the average of the peripheral measurements, the traditional CTDIw underestimated the average dose by up to 4.3% due to aggressive drop-off of the CTDI100 at the phantom periphery. Conclusions The linearity and rotational-invariance assumptions behind the traditional CTDIw formalism may not be valid for modern CT systems and thus CTDIw may not accurately represent the average dose or radiation output within a CTDI phantom. Utilizing data from all four peripheral locations always improves accuracy of CTDIw in representing the true average dose. For the large (32 cm) phantom, nonlinear models and more measurement points are needed if a more precise estimation of the average axial dose is required.
引用
收藏
页码:6484 / 6499
页数:16
相关论文
共 30 条
[1]  
AAPM Task Group 111, 2010, 111 AAPM
[2]  
AAPM Task Group 204, 2011, 2044 AAPM
[3]  
AAPM Task Group 246 and EFOMP, 2019, 246 AAPM
[4]  
AAPM Task Group 293, 2019, 293 AAPM
[5]   An assessment of the efficiency of methods for measurement of the computed tomography dose index (CTDI) for cone beam (CBCT) dosimetry by Monte Carlo simulation [J].
Abuhaimed, Abdullah ;
Martin, Colin J. ;
Sankaralingam, Marimuthu ;
Gentle, David J. ;
McJury, Mark .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (21) :6307-6326
[6]  
American Association of Physicists in Medicine, 2020, 111 AAPM
[7]  
[Anonymous], 2019, AM COLL RAD ACCR PRO
[8]   The trouble with CTDI100 [J].
Boone, John M. .
MEDICAL PHYSICS, 2007, 34 (04) :1364-1371
[9]  
Cederquist B, 2008, THESIS
[10]   A Numerical Investigation for the Optimal Positions and Weighting Coefficients of Point Dose Measurements in the Weighted CTDI [J].
Choi, Jang-Hwan ;
Constantin, Dragos ;
Fahrig, Rebecca .
MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412