Development and validation of a novel leaky surface acoustic wave immunosensor array for label-free and high-sensitive detection of cyclosporin A in whole-blood samples

被引:20
作者
Chang, Kai [1 ]
Wang, Feng [1 ]
Ding, Yi [2 ]
Pan, Feng [1 ]
Li, Fake [1 ]
Jia, Shuangrong [1 ]
Lu, Weiping [1 ]
Deng, Shaoli [1 ]
Shi, Jianfeng [2 ]
Chen, Ming [1 ]
机构
[1] Third Mil Med Univ, Daping Hosp, Inst Surg Res, Dept Clin Lab Med, Chongqing 400042, Peoples R China
[2] Chinese Elect Sci & Tech Grp Co, Res Inst 26, Chongqing 400060, Peoples R China
关键词
Leaky surface acoustic wave; Immunosensor; Array; Cyclosporin A; PIEZOELECTRIC IMMUNOSENSOR; RECENT TRENDS; QUANTIFICATION; IMMUNOASSAY; BIOSENSORS;
D O I
10.1016/j.bios.2013.10.066
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This manuscript described a novel 2 x 3 model of leaky surface acoustic wave (LSAW) immunosensor array for label-free and high-sensitive detection of Cyclosporin A (CsA) in whole-blood samples. In this technique, every resonator crystal unit of the LSAW immunosensor array had an individual oscillator circuit to work without mutual interference. The LSAW immunosensor was first immobilized with protein A from Staphylococcus aureus and monoclonal anti-CsA antibody on the gold electrode surface of 100 MHz LiTaO3 piezoelectric crystals, which then captured the CsA. The CsA increased the mass loading of LSAW immunosensor and leaded to phase shifts of LSAW. Consequently, under optimal conditions, the designed LSAW immunosensor exhibited a detection limit of 0.89 ng/mL, quantification limit of 2.96 ng/mL, and wide dynamic linear range from 1 ng/mL to 1000 ng/mL for CsA detection. Application of the LSAW immunosensor array to clinical sample revealed that consistency and comparability between LSAW immunosensor and the enzyme multiplied immunoassay method were good. Moreover, the immunosensor could be regenerated for ten times without appreciable loss of activity. Therefore, the self-designed LSAW immunosensor array provided a rapid, accurate, label-free, easy handling, and dynamic real-time method for the detection of immunosuppressive drugs in clinical laboratory. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:151 / 157
页数:7
相关论文
共 31 条
[1]   Evaluation of antibody immobilization methods for piezoelectric biosensor application [J].
Babacan, S ;
Pivarnik, P ;
Letcher, S ;
Rand, AG .
BIOSENSORS & BIOELECTRONICS, 2000, 15 (11-12) :615-621
[2]   Acousto-optic interaction with leaky surface acoustic waves in Y-cut LiTaO3 crystals [J].
Belovickis, Jaroslavas ;
Rimeika, Romualdas ;
Ciplys, Daumantas .
ULTRASONICS, 2012, 52 (05) :593-597
[3]   Construction of a novel peptide nucleic acid piezoelectric gene sensor microarray detection system [J].
Chen, M ;
Liu, MH ;
Yu, LL ;
Cai, GR ;
Chen, QH ;
Wu, R ;
Wang, F ;
Zhang, B ;
Jiang, TL ;
Fu, WL .
JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2005, 5 (08) :1266-1272
[4]   Surface engineering: Optimization of antigen presentation in self-assembled monolayers [J].
Duschl, C ;
SevinLandais, AF ;
Vogel, H .
BIOPHYSICAL JOURNAL, 1996, 70 (04) :1985-1995
[5]   Surface acoustic wave sensors in the bioanalytical field: Recent trends and challenges [J].
Gronewold, Thomas M. A. .
ANALYTICA CHIMICA ACTA, 2007, 603 (02) :119-128
[6]   The possibility of using the specific RIA method for the area-under-time-concentration curve sparse sampling calculation of cyclosporin A despite a large post-dose overestimation [J].
Grundmann, M. ;
Koristkova, B. ;
Brozmanova, H. ;
Perinova, I. ;
Safarcik, K. .
INTERNATIONAL JOURNAL OF CLINICAL PHARMACOLOGY AND THERAPEUTICS, 2011, 49 (01) :30-37
[7]  
Hamwi A, 1999, AM J CLIN PATHOL, V112, P358
[8]   Recent trends in antibody based sensors [J].
Holford, Timothy R. J. ;
Davis, Frank ;
Higson, Seamus P. J. .
BIOSENSORS & BIOELECTRONICS, 2012, 34 (01) :12-24
[9]   Analysis and suppression of side radiation in leaky SAW resonators [J].
Inoue, Shogo ;
Tsutsumi, Jun ;
Matsuda, Takashi ;
Ueda, Masanori ;
Ikata, Osamu ;
Satoh, Yoshio .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (08) :1692-1699
[10]   Evaluation of different micro/nanobeads used as amplifiers in QCM immunosensor for more sensitive detection of E. coli O157:H7 [J].
Jiang, Xuesong ;
Wang, Ronghui ;
Wang, Yun ;
Su, Xiaoli ;
Ying, Yibin ;
Wang, Jianping ;
Li, Yanbin .
BIOSENSORS & BIOELECTRONICS, 2011, 29 (01) :23-28