Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes

被引:129
作者
Yang, Yongfeng
Dokhale, Purushottam A.
Silverman, Robert W.
Shah, Kanai S.
McClish, Mickel A.
Farrell, Richard
Entine, Gerald
Cherry, Simon R.
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[2] Radiat Monitoring Devices Inc, Watertown, MA 02172 USA
[3] Univ Calif Los Angeles, Dept Mol & Med Pharmacol, Los Angeles, CA 90095 USA
关键词
D O I
10.1088/0031-9155/51/9/001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We explore dual-ended read out of LSO arrays with two position sensitive avalanche photodiodes (PSAPDs) as a high resolution, high efficiency depth-encoding detector for PET applications. Flood histograms, energy resolution and depth of interaction (DOI) resolution were measured for unpolished LSO arrays with individual crystal sizes of 1.0, 1.3 and 1.5 mm, and for a polished LSO array with 1.3 mm pixels. The thickness of the crystal arrays was 20 mm. Good flood histograms were obtained for all four arrays, and crystals in all four arrays can be clearly resolved. Although the amplitude of each PSAPD signal decreases as the interaction depth moves further from the PSAPD, the sum of the two PSAPD signals is essentially constant with irradiation depth for all four arrays. The energy resolutions were similar for all four arrays, ranging from 14.7% to 15.4%. A DOI resolution of 3-4 mm (including the width of the irradiation band which is similar to 2 mm) was obtained for all the unpolished arrays. The best DOI resolution was achieved with the unpolished 1mm array (average 3.5 mm). The DOI resolution for the 1.3 mm and 1.5 mm unpolished arrays was 3.7 and 4.0 mm respectively. For the polished array, the DOI resolution was only 16.5 mm. Summing the DOI profiles across all crystals for the 1 mm array only degraded the DOI resolution from 3.5 mm to 3.9 mm, indicating that it may not be necessary to calibrate the DOI response separately for each crystal within an array. The DOI response of individual crystals in the array confirms this finding. These results provide a detailed characterization of the DOI response of these PSAPD-based PET detectors which will be important in the design and calibration of a PET scanner making use of this detector approach.
引用
收藏
页码:2131 / 2142
页数:12
相关论文
共 32 条
[21]   Position sensitive APDs for small animal PET imaging [J].
Shah, KS ;
Grazioso, R ;
Farrell, R ;
Glodo, J ;
McClish, M ;
Entine, G ;
Dokhale, P ;
Cherry, SR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (01) :91-95
[22]   Position-sensitive avalanche photodiodes for gamma-ray imaging [J].
Shah, KS ;
Farrell, R ;
Grazioso, R ;
Harmon, ES ;
Karplus, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (04) :1687-1692
[23]   Dual APD array readout of LSO crystals: Optimization of crystal surface treatment [J].
Shao, Y ;
Meadors, K ;
Silverman, RW ;
Farrell, R ;
Cirignano, L ;
Grazioso, R ;
Shah, KS ;
Cherry, SR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (03) :649-654
[24]   Performance evaluation of the microPET P4: a PET system dedicated to animal imaging [J].
Tai, YC ;
Chatziioannou, A ;
Siegel, S ;
Young, J ;
Newport, D ;
Goble, RN ;
Nutt, RE ;
Cherry, SR .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (07) :1845-1862
[25]   MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging [J].
Tai, YC ;
Chatziioannou, AF ;
Yang, YF ;
Silverman, RW ;
Meadors, K ;
Siegel, S ;
Newport, DF ;
Stickel, JR ;
Cherry, SR .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (11) :1519-1537
[26]  
Tai YC, 2005, J NUCL MED, V46, P455
[27]   A four-layer depth of interaction detector block for small animal PET [J].
Tsuda, T ;
Murayama, H ;
Kitamura, K ;
Yamaya, T ;
Yoshida, E ;
Omura, T ;
Kawai, H ;
Inadama, N ;
Orita, N .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) :2537-2542
[28]   A high resolution animal PET scanner using compact PS-PMT detectors [J].
Watanabe, M ;
Okada, H ;
Shimizu, K ;
Omura, T ;
Yoshikawa, E ;
Kosugi, T ;
Mori, S ;
Yamashita, T .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1277-1282
[29]   Giant birefringent optics in multilayer polymer mirrors [J].
Weber, MF ;
Stover, CA ;
Gilbert, LR ;
Nevitt, TJ ;
Ouderkirk, AJ .
SCIENCE, 2000, 287 (5462) :2451-2456
[30]   Optimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging [J].
Yang, YF ;
Tai, YC ;
Siegel, S ;
Newport, DF ;
Bai, B ;
Li, QZ ;
Leahy, RM ;
Cherry, SR .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12) :2527-2545