Determination of trace amounts of mercury, lead and cadmium in steels by ICP-MS after ion exchange chromatographic separation

被引:9
作者
Fujimoto, K
Chino, A
机构
[1] JFE Steel Corp, Steel Res Lab, Chuo Ku, Chiba 2600835, Japan
[2] JFE Steel Corp, Steel Res Lab, Kawasaki Ku, Kawasaki, Kanagawa 2100855, Japan
关键词
cation exchange chromatographic separation; trace mercury; lead and cadmium in steel; simultaneous analysis by ICP-MS;
D O I
10.2116/bunsekikagaku.55.245
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Trace amounts of mercury, lead, and cadmium at a level of ng g(-1) in steel were determined by ICP-MS after cation-exchange chromatographic separation. The sample was decomposed with nitric acid, and the analytes were separated from the iron matrix in 1 M hydrofluoric acid solution, Using the cation-exchange resin packed in a mini-column. The elements adsorbed on the resin were eluted by 8 M hydrochloric acid solution and determined by ICP-MS simultaneously. he lower limits of detection based on 3 sigma of blank values were 2 ng g(-1) for mercury, 9 ng g(-1) lead, and 0.5 ng g(-1) for cadmium.
引用
收藏
页码:245 / 249
页数:5
相关论文
共 6 条
[1]  
FARIS JP, 1960, ANAL CHEM, V32, P521
[2]   Direct determination of methylmercury and inorganic mercury in biological materials by solid sampling-electrothermal vaporization-inductively coupled plasma-isotope dilution-mass spectrometry [J].
Gelaude, I ;
Dams, R ;
Resano, M ;
Vanhaecke, F ;
Moens, L .
ANALYTICAL CHEMISTRY, 2002, 74 (15) :3833-3842
[3]  
KRAUS KA, 1955, P INT C PEAC US AT E, V7, P118
[4]  
MINCZEWSKI J, 1982, SEPARATION PRECONCEN, P420
[5]  
NELSON F, 1964, J CHROMATOGR, V13, P532
[6]   On-line preconcentration and determination of lead in iron and steel by flow injection-flame atomic absorption spectrometry [J].
Seki, T ;
Takigawa, H ;
Hirano, Y ;
Ishibashi, Y ;
Oguma, K .
ANALYTICAL SCIENCES, 2000, 16 (05) :513-516