Reduced Patient Radiation Exposure during Neurodiagnostic and Interventional X-Ray Angiography with a New Imaging Platform

被引:9
作者
van der Marel, K. [1 ]
Vedantham, S. [1 ]
van der Bom, I. M. J. [2 ]
Howk, M. [1 ]
Narain, T. [1 ]
Ty, K. [1 ]
Karellas, A. [1 ]
Gounis, M. J. [1 ]
Puri, A. S. [1 ]
Wakhloo, A. K. [1 ]
机构
[1] Univ Massachusetts, Sch Med, Dept Radiol, New England Ctr Stroke Res, Worcester, MA USA
[2] Philips Healthcare, Best, Netherlands
关键词
NOISE-REDUCTION TECHNOLOGY; SKIN; RADIOLOGY; NEURORADIOLOGY; QUALITY;
D O I
10.3174/ajnr.A5049
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: Advancements in medical device and imaging technology as well as accruing clinical evidence have accelerated the growth of the endovascular treatment of cerebrovascular diseases. However, the augmented role of these procedures raises concerns about the radiation dose to patients and operators. We evaluated patient doses from an x-ray imaging platform with radiation dose-reduction technology, which combined image noise reduction, motion correction, and contrast-dependent temporal averaging with optimized x-ray exposure settings. MATERIALS AND METHODS: In this single-center, retrospective study, cumulative dose-area product inclusive of fluoroscopy, angiography, and 3D acquisitions for all neurovascular procedures performed during a 2-year period on the dose-reduction platform were compared with a reference platform. Key study features were the following: The neurointerventional radiologist could select the targeted dose reduction for each patient with the dose-reduction platform, and the statistical analyses included patient characteristics and the neurointerventional radiologist as covariates. The analyzed outcome measures were cumulative dose (kerma)-area product, fluoroscopy duration, and administered contrast volume. RESULTS: A total of 1238 neurointerventional cases were included, of which 914 and 324 were performed on the reference and dose-reduction platforms, respectively. Over all diagnostic and neurointerventional procedures, the cumulative dose-area product was significantly reduced by 53.2% (mean reduction, 160.3 Gy x cm(2); P < .0001), fluoroscopy duration was marginally significantly increased (mean increase, 5.2 minutes; P = .0491), and contrast volume was nonsignificantly increased (mean increase, 15.3 mL; P = .1616) with the dose-reduction platform. CONCLUSIONS: A significant reduction in patient radiation dose is achievable during neurovascular procedures by using dose-reduction technology with a minimal impact on workflow.
引用
收藏
页码:442 / 449
页数:8
相关论文
共 30 条
  • [11] HURVICH CM, 1989, BIOMETRIKA, V76, P297, DOI 10.2307/2336663
  • [12] Validation and Initial Clinical Use of Automatic Peak Skin Dose Localization with Fluoroscopic and Interventional Procedures
    Khodadadegan, Yasaman
    Zhang, Muhong
    Pavlicek, William
    Paden, Robert G.
    Chong, Brian
    Huettl, Eric A.
    Schueler, Beth A.
    Fetterly, Kenneth A.
    Langer, Steve G.
    Wu, Teresa
    [J]. RADIOLOGY, 2013, 266 (01) : 246 - 255
  • [13] Patient dose during interventional neuroradiology procedures: Results from a multi-center study
    Kien, N.
    Rehel, J. -L.
    Etard, C.
    Aubert, B.
    [J]. JOURNAL DE RADIOLOGIE, 2011, 92 (12): : 1101 - 1112
  • [14] Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190
    Lin, Pei-Jan P.
    Schueler, Beth A.
    Balter, Stephen
    Strauss, Keith J.
    Wunderle, Kevin A.
    LaFrance, M. Terry
    Kim, Don-Soo
    Behrman, Richard H.
    Shepard, S. Jeff
    Bercha, Ishtiaq H.
    [J]. MEDICAL PHYSICS, 2015, 42 (12) : 6815 - 6829
  • [15] Staff Doses in Interventional Radiology: A National Survey
    Mariano Sanchez, Roberto
    Vano, Eliseo
    Fernandez, Jose M.
    Rosales, Francisco
    Sotil, Jesus
    Carrera, Francisco
    Garcia, Maria A.
    Soler, Maria M.
    Hernandez-Armas, Jose
    Martinez, Luis C.
    Verdu, Jose F.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 23 (11) : 1496 - 1501
  • [16] Radiation doses in interventional radiology procedures: The RAD-IR Study - Part I: Overall measures of dose
    Miller, DL
    Balter, S
    Cole, PE
    Lu, HT
    Schueler, BA
    Geisinger, M
    Berenstein, A
    Albert, R
    Georgia, JD
    Noonan, PT
    Cardella, JF
    George, JS
    Russell, EJ
    Malisch, TW
    Vogelzang, RL
    Miller, GL
    Anderson, J
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (06) : 711 - 727
  • [17] International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion
    Molyneux, AJ
    Kerr, RSC
    Yu, LM
    Clarke, M
    Sneade, M
    Yarnold, JA
    Sandercock, P
    [J]. LANCET, 2005, 366 (9488) : 809 - 817
  • [18] Patient radiation dose reduction using an X-ray imaging noise reduction technology for cardiac angiography and intervention
    Nakamura, Shigeru
    Kobayashi, Tomoko
    Funatsu, Atsushi
    Okada, Tadahisa
    Mauti, Maria
    Waizumi, Yuki
    Yamada, Shinichi
    [J]. HEART AND VESSELS, 2016, 31 (05) : 655 - 663
  • [19] Norbash AM, 1996, AM J NEURORADIOL, V17, P41
  • [20] Radiation Exposure in Vascular Angiographic Procedures
    Pitton, Michael Bernhard
    Kloeckner, Roman
    Schneider, Jens
    Ruckes, Christian
    Bersch, Anton
    Dueber, Christoph
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 23 (11) : 1487 - 1495