Face-to-face tracheal intubation in adult patients: a comparison of the Airtraq™, Glidescope™ and Fastrach™ devices

被引:20
作者
Arslan, Zehra Ipek [1 ]
Alparslan, Volkan [1 ]
Ozdal, Pinar [1 ]
Toker, Kamil [1 ]
Solak, Mine [1 ]
机构
[1] Kocaeli Univ, Fac Med, Anaesthesiol & Reanimat, Umuttepe, Kocaeli, Turkey
关键词
Face-to-face; Intubation; Airtraq; Glidescope; Fastrach; LARYNGEAL MASK AIRWAY; MACINTOSH LARYNGOSCOPE; VIDEO LARYNGOSCOPES; VIDEO-AIRTRAQ(TM); MANAGEMENT; MOVEMENT; POSITION; TRIAL; SCOPE;
D O I
10.1007/s00540-015-2052-6
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Airway management in emergency settings can be difficult due to limited access to the patient. The use of video laryngoscopes along with the Fastrach (TM) device improves tracheal intubation; however, the use of such devices in a face-to-face intubation model has not been evaluated in adult patients. After obtaining official approval from the Local Research Ethics Committee and written informed consent from the patients, 120 patients were enrolled in this prospective randomized study. The patients were ASA I and ASA II according to the American Society of Anesthesiologists Physical Status Classification System. Rocuronium was administered for neuromuscular blockade following standard anesthesia monitoring and induction. The patients were divided into three groups (40 patients per group) and their tracheas were intubated via a face-to-face approach with the Airtraq (TM), Glidescope (TM) or Fastrach (TM) devices. The intubation success rates of the Airtraq (TM), Glidescope (TM) and Fastrach (TM) devices were similar (100, 98 and 90 %; p = 0.07). The insertion time for the Airtraq (TM) [8.5 (6-11) s] was the shortest followed by the Glidescope (TM) [11 (7-19) s] and the Fastrach (TM) [16.5 (14.3-21.8) s; p < 0.001]. The intubation time for the Airtraq (TM) [14 (10.3-18.8) s] was shorter than the Glidescope (TM) [25 (18-45) s], and Fastrach (TM) devices [46.5 (40-65) s; p < 0.001]. The Glidescope (TM) device required a greater number of optimization maneuvers (p = 0.009) and intubation attempts than the Airtraq (TM) (p = 0.004). Esophageal intubation (p = 0.001) and mucosal damage were more common in the Fastrach (TM) group (p = 0.03). The Airtraq (TM) device provided faster insertion and intubation times and enabled better Cormack-Lehane grades. Additionally, the Airtraq (TM) device required the minimum number of optimization maneuvers and was associated with fewer complications and fewer intubation attempts than the Glidescope (TM) and Fastrach (TM) devices during face-to-face tracheal intubation.
引用
收藏
页码:893 / 898
页数:6
相关论文
共 25 条
[1]   Simulating face-to-face tracheal intubation of a trapped patient: a randomized comparison of the LMA Fastrach™, the GlideScope™, and the Airtraq™ laryngoscope [J].
Amathieu, R. ;
Sudrial, J. ;
Abdi, W. ;
Luis, D. ;
Hahouache, H. ;
Combes, X. ;
Dhonneur, G. .
BRITISH JOURNAL OF ANAESTHESIA, 2012, 108 (01) :140-145
[2]  
[Anonymous], 2010, US MAN QUICK REF GUI
[3]   Tracheal intubation with restricted access: a randomised comparison of the Pentax-Airway Scope and Macintosh laryngoscope in a manikin [J].
Asai, T. .
ANAESTHESIA, 2009, 64 (10) :1114-1117
[4]   Advanced trauma life support (ATLS®): The ninth edition [J].
Brasel, Karen J. .
JOURNAL OF TRAUMA AND ACUTE CARE SURGERY, 2013, 74 (05) :1363-1366
[5]   DIFFICULT TRACHEAL INTUBATION IN OBSTETRICS [J].
CORMACK, RS ;
LEHANE, J .
ANAESTHESIA, 1984, 39 (11) :1105-1111
[6]   Urgent face-to-face tracheal re-intubation using Video-Airtraq™ in ICU patients placed in the sitting position [J].
Dhonneur, Gilles ;
Zraier, Sana ;
Sebbah, Jean Luc ;
Haouache, Hakim .
INTENSIVE CARE MEDICINE, 2014, 40 (04) :625-626
[7]   Comparison between intubation through ILMA and Airtraq, in different non-conventional patient positions: a manikin study [J].
Grosomanidis, V. ;
Amaniti, E. ;
Pourzitaki, Ch ;
Fyntanidou, V. ;
Mouratidis, K. ;
Vasilakos, D. .
EMERGENCY MEDICINE JOURNAL, 2012, 29 (01) :32-36
[8]  
Hilker T, 1999, Prehosp Emerg Care, V3, P74, DOI 10.1080/10903129908958911
[9]   A comparison of cervical spine movement during laryngoscopy using the Airtraq® or Macintosh laryngoscopes [J].
Hirabayashi, Y. ;
Fujita, A. ;
Seo, N. ;
Sugimoto, H. .
ANAESTHESIA, 2008, 63 (06) :635-640
[10]   Evaluation of ease of intubation with the GlideScope® or Macintosh laryngoscope by anaesthetists in simulated easy and difficult laryngoscopy [J].
Lim, TJ ;
Lim, Y ;
Liu, EHC .
ANAESTHESIA, 2005, 60 (02) :180-183