This paper reports on different sources of errors that occur in the calibration process of a superconducting gravimeter (SG), determined by comparison with a ballistic absolute gravimeter (AG); some of them have never been discussed in the literature. We then provide methods to mitigate the impact of those errors, to achieve a robust calibration estimate at the level. We demonstrate that a standard deviation at the level of can be reached within 48 h by measuring at spring tides and by increasing the AG sampling rate. This is much shorter than what is classically reported in previous empirical studies. Measuring more than 5 days around a tidal extreme does not improve the precision in the calibration factor significantly, as the variation in the error as a function of does not apply, considering the decrease in signal amplitude due to the tidal modulation. However, we investigate the precision improvement up to 120 days, which can be useful if an AG is run continuously: at mid-latitude it would require 21 days to ensure a calibration factor at the level with a 99.7 % confidence interval. We also show that restricting the AG measurement periods to tidal extrema can reduce instrument demand, while this does not affect the precision on the calibration factor significantly. Then, we quantify the effect of high microseismic noise causing aliasing in the AG time series. We eventually discuss the attenuation bias that might be induced by noisy time series of the SG. When experiments are performed at the level, 7 are needed to ensure that the error in the calibration estimate will be at the 1 per mille level with a 99 % confidence.
引用
收藏
页码:91 / 99
页数:9
相关论文
共 27 条
[1]   A CALIBRATION SYSTEM FOR SUPERCONDUCTING GRAVIMETERS [J].
ACHILLI, V ;
BALDI, P ;
CASULA, G ;
ERRANI, M ;
FOCARDI, S ;
GUERZONI, M ;
PALMONARI, F ;
RAGUNI, G .
BULLETIN GEODESIQUE, 1995, 69 (02) :73-80
[2]  
[Anonymous], 1963, NATL BUREAU STANDARD
[3]   Validating Earth and ocean tide models using tidal gravity measurements [J].
Baker, TF ;
Bos, MS .
GEOPHYSICAL JOURNAL INTERNATIONAL, 2003, 152 (02) :468-485
[4]   Noise levels of superconducting gravimeters at seismic frequencies [J].
Banka, D ;
Crossley, D .
GEOPHYSICAL JOURNAL INTERNATIONAL, 1999, 139 (01) :87-97
[5]   A review of the GGP network and scientific challenges [J].
Crossley, David ;
Hinderer, Jacques .
JOURNAL OF GEODYNAMICS, 2009, 48 (3-5) :299-304
[6]   Precision gravimetry with atomic sensors [J].
de Angelis, M. ;
Bertoldi, A. ;
Cacciapuoti, L. ;
Giorgini, A. ;
Lamporesi, G. ;
Prevedelli, M. ;
Saccorotti, G. ;
Sorrentino, F. ;
Tino, G. M. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2009, 20 (02)
[7]  
Faller J.E., 1983, BUR GRAV INT B INF, V53, P87
[8]   Calibration of a superconducting gravimeter by comparison with an absolute gravimeter FG5 in Boulder [J].
Francis, O ;
Niebauer, TM ;
Sasagawa, G ;
Klopping, F ;
Gschwind, J .
GEOPHYSICAL RESEARCH LETTERS, 1998, 25 (07) :1075-1078
[9]  
Francis O., 2001, J GEODETIC SOC JAPAN, V47, P16, DOI DOI 10.11366/SOKUCHI1954.47.16
[10]  
Francis O., 1997, INT ASS GEODESY S, P212