Suppression of dark current radiation in step-and-shoot intensity modulated radiation therapy by the initial pulse-forming network

被引:13
作者
Cheng, CW
Das, IJ
Ndlovu, AM
机构
[1] Morristown Mem Hosp, Morristown, NJ 07962 USA
[2] Hosp Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
关键词
IMRT; dark current radiation; whole body dose; machine commissioning;
D O I
10.1118/1.1500403
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effect of the initial pulse forming network (IPFN) on the suppression of dark current is investigated for a Siemens Primus accelerator. The dark current produces a spurious radiation, which is referred to as dark current radiation (DCR) in this study. In the step-and-shoot delivery of an intensity modulated radiation therapy (IMRT), the DCR could be of some concern for whole body dose along with leakage radiation through collimator jaws or multileaf collimator. By adjusting the IPFN-to-PFN ratio to >0.8, the DCR can be measured with an ion chamber during the "PAUSE" state of the accelerator in the IMRT mode. For 15 MV x rays, the magnitude of the DCR is approximately equal to 0.7% of the dose at d(max) for a 10 x 10 cm 2 field. The DCR has a similar central axis depth dose as a 15 MV beam as determined from a water phantom scan. When the IPFN-to-PFN ratio is lowered to <0.8, no DCR is detected. For low energy x rays (6 MV), no DCR is detected regardless of the IPFN-to-PFN ratio. Although the DCR is studied only for the Siemens Primus model accelerator, the same precaution applies to other models of modern accelerators from other vendors. Due to the large number of field segments used in a step-and-shoot IMRT, it is imperative therefore, that dark current evaluation be part of machine commissioning and annual calibration for high-energy photon beams. Should DCR be detected, the medical physicist should work with a service engineer to rectify the problem. In view of DCR and whole body dose, low-energy photon beams are advisable for IMRT. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:1974 / 1979
页数:6
相关论文
共 24 条
[1]   TELETHERAPY BEAM CHARACTERISTICS - THE 1ST 2ND [J].
BARISH, RJ ;
FLEISCHMAN, RC ;
PIPMAN, YM .
MEDICAL PHYSICS, 1987, 14 (04) :657-661
[2]   Improved delivery efficiency for step and shoot intensity modulated radiotherapy using a fast-tuning magnetron [J].
Budgell, GJ ;
Martens, C ;
Claus, F .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (11) :N253-N261
[3]   Comparison of beam characteristics in intensity modulated radiation therapy (IMRT) and those under normal treatment condition [J].
Cheng, CW ;
Das, IJ .
MEDICAL PHYSICS, 2002, 29 (02) :226-230
[4]   Beam characteristics of a retrofitted double-focused multileaf collimator [J].
Das, IJ ;
Desobry, GE ;
McNeeley, SW ;
Cheng, EC ;
Schultheiss, TE .
MEDICAL PHYSICS, 1998, 25 (09) :1676-1684
[5]   HIGHER ENERGY - IS IT NECESSARY, IS IT WORTH THE COST FOR RADIATION ONCOLOGY [J].
DAS, IJ ;
KASE, KR .
MEDICAL PHYSICS, 1992, 19 (04) :917-925
[6]   Dose estimation to critical organs from vertex field treatment of brain tumors [J].
Das, IJ ;
Cheng, CW ;
Fein, DA ;
Coia, LR ;
Curran, WJ ;
Fowble, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (05) :1023-1029
[7]   DOSIMETRIC PROBLEMS AT LOW MONITOR UNIT SETTINGS FOR SCANNED AND SCATTERING FOIL ELECTRON-BEAMS [J].
DAS, IJ ;
HARRINGTON, JC ;
AKBER, SF ;
TOMER, AF ;
MURRAY, JC ;
CHENG, CW .
MEDICAL PHYSICS, 1994, 21 (06) :821-826
[8]   DOSIMETRIC ACCURACY AT LOW MONITOR UNIT SETTINGS [J].
DAS, IJ ;
KASE, KR ;
TELLO, VM .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (765) :808-811
[9]   Validation of a new virtual wedge model [J].
Desobry, GE ;
Waldron, TJ ;
Das, IJ .
MEDICAL PHYSICS, 1998, 25 (01) :71-72
[10]   Estimates of whole-body dose equivalent produced by beam intensity modulated conformal therapy [J].
Followill, D ;
Geis, P ;
Boyer, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03) :667-672