Sensor-Based Physiologic Control Strategy for Biventricular Support with Rotary Blood Pumps

被引:16
作者
Wang, Yu [1 ]
Koenig, Steven C. [2 ,3 ]
Wu, Zhongjun [3 ]
Slaughter, Mark S. [3 ]
Giridharan, Guruprasad A. [2 ]
机构
[1] Dalian Univ Technol, Fac Elect Informat & Elect Engn, Dept Biomed Engn, Dalian, Liaoning, Peoples R China
[2] Univ Louisville, Cardiovasc Innovat Inst, Dept Bioengn, Room 407,302 East Muhammad Ali Blvd, Louisville, KY 40202 USA
[3] Univ Louisville, Cardiovasc Innovat Inst, Dept Cardiovasc & Thorac Surg, Louisville, KY 40292 USA
关键词
biventricular assist devices; flow balancing; physiologic control; suction prevention; sensor control algorithm; TOTAL ARTIFICIAL-HEART; VENTRICULAR ASSIST DEVICES; IN-VIVO EVALUATION; CONTROL-SYSTEMS; FLOW; PRESSURE; PRELOAD; SUCTION; SENSITIVITY; PARAMETERS;
D O I
10.1097/MAT.0000000000000671
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Rotary biventricular assist devices (BiVAD) are becoming a clinically accepted treatment option for end-stage biventricular failure. To improve BiVAD efficacy and safety, we propose a control algorithm to achieve the clinical objectives of maintaining left-right-sided balance, restoring physiologic flows, and preventing ventricular suction. The control algorithm consists of two proportional-integral (PI) controllers for left and right ventricular assist devices (LVAD and RVAD) to maintain differential pump pressure across LVAD (Delta P-L) and RVAD (Delta P-R) to provide left-right balance and physiologic flow. To prevent ventricular suction, LVAD and RVAD pump speed differentials (Delta RPML, Delta RPMR) were maintained above user-defined thresholds. Efficacy and robustness of the proposed algorithm were tested in silico for axial and centrifugal flow BiVAD using 1) normal and excessive Delta P-L and/or Delta P-R setpoints, 2) rapid threefold increase in pulmonary vascular or vena caval resistances, 3) transient responses from exercise to rest, and 4) ventricular fibrillation. The study successfully demonstrated that the proposed BiVAD algorithm achieved the clinical objectives but required pressure sensors to continuously measure Delta P-L and Delta P-R. The proposed control algorithm is device independent, should not require any modifications to the pump or inflow/outflow cannulae/grafts, and may be directly applied to current rotary blood pumps for biventricular support.
引用
收藏
页码:338 / 350
页数:13
相关论文
共 40 条
[1]   Physiological control of a total artificial heart: conductance- and arterial pressure-based control [J].
Abe, Y ;
Chinzei, T ;
Mabuchi, K ;
Snyder, AJ ;
Isoyama, T ;
Imanishi, K ;
Yonezawa, T ;
Matsuura, H ;
Kouno, A ;
Ono, T ;
Atsumi, K ;
Fujimasa, I ;
Imachi, K .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 84 (03) :868-876
[2]   Advance in animal experiments with the undulation pump total artificial heart: 50 and 54 day survival periods with 1/R control [J].
Abe, Y ;
Chinzei, T ;
Isoyama, T ;
Kobayashi, S ;
Ono, T ;
Saito, I ;
Iwasaki, K ;
Ishimaru, M ;
Baba, A ;
Kouno, A ;
Ozeki, T ;
Tohyama, T ;
Imachi, K .
ASAIO JOURNAL, 2003, 49 (03) :325-332
[3]   Implantation of the undulation pump total artificial heart in the goat [J].
Abe, Y ;
Chinzei, T ;
Ono, T ;
Mochizuki, S ;
Saito, I ;
Guba, P ;
Karita, T ;
Sun, YP ;
Kouno, A ;
Isoyama, T ;
Baba, K ;
Imachi, K .
ARTIFICIAL ORGANS, 1999, 23 (10) :932-938
[4]   Hemodynamic controller for left ventricular assist device based on pulsatility ratio [J].
Choi, Seongjin ;
Boston, J. Robert ;
Antaki, James F. .
ARTIFICIAL ORGANS, 2007, 31 (02) :114-125
[5]  
Choi SJ, 1997, P AMER CONTR CONF, P3714, DOI 10.1109/ACC.1997.609538
[6]   Control strategy for biventricular assistance with mixed-flow pumps [J].
Endo, G ;
Araki, K ;
Oshikawa, M ;
Kojima, K ;
Saitoh, T ;
Nakamura, K ;
Onitsuka, T .
ARTIFICIAL ORGANS, 2000, 24 (08) :594-599
[7]   Preload Sensitivity in Cardiac Assist Devices [J].
Fukamachi, Kiyotaka ;
Shiose, Akira ;
Massiello, Alex ;
Horvath, David J. ;
Golding, Leonard A. R. ;
Lee, Sangjin ;
Starling, Randall C. .
ANNALS OF THORACIC SURGERY, 2013, 95 (01) :373-380
[8]   An innovative, sensorless, pulsatile, continuous-flow total artificial heart: Device design and initial in vitro study [J].
Fukamachi, Kiyotaka ;
Horvath, David J. ;
Massietto, Alex L. ;
Fumoto, Hideyuki ;
Horai, Tetsuya ;
Rao, Santosh ;
Golding, Leonard A. R. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2010, 29 (01) :13-20
[9]   Comparison of Preload-Sensitive Pressure and Flow Controller Strategies for a Dual Device Biventricular Support System [J].
Gaddum, Nicholas Richard ;
Timms, Daniel Lee ;
Stevens, Michael ;
Mason, David ;
Lovell, Nigel ;
Fraser, John F. .
ARTIFICIAL ORGANS, 2012, 36 (03) :256-265
[10]   A Passively Controlled Biventricular Support Device [J].
Gaddum, Nicholas Richard ;
Timms, Daniel Lee ;
Pearcy, Mark John .