The biomechanical effects of cervical multilevel oblique corpectomy

被引:25
作者
Çagli, S
Chamberlain, RH
Sonntag, VKH
Crawford, NR
机构
[1] Barrow Neurol Inst, Phoenix, AZ 85013 USA
[2] Ege Univ, Sch Med, Dept Neurosurg, Izmir, Turkey
关键词
oblique corpectomy; multilevel corpectomy; cervical spondylotic myelopathy; anterior plate; biomechanics; kinematics;
D O I
10.1097/01.BRS.0000129896.80044.B6
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. A repeated-measures flexibility test was performed in vitro using human cadaveric spines. Objectives. To compare changes in cervical biomechanics associated with multilevel oblique corpectomy and standard grafted corpectomy with or without plating. Summary of Background Data. Standard multilevel plated and unplated corpectomies are susceptible to instability in vitro. The authors are unaware of any previous research on the biomechanics of multilevel oblique corpectomy. Methods. Six human cadaveric cervical spine specimens (C3 - T1) were tested: 1) normal; 2) after 2-level multilevel oblique corpectomy; 3) after expanding multilevel oblique corpectomy to represent standard grafted and plated corpectomy; and 4) after removing the anterior plate. Pure moments were applied to induce flexion, extension, lateral bending, and axial rotation while recording motion stereophotogrammetrically. Results. Compared to normal, the range of motion after multilevel oblique corpectomy increased 15% during flexion, 18% during extension, 11% during lateral bending, and 18% during axial rotation. These increases were about one-third of the increases observed after standard corpectomy without plating. Multilevel oblique corpectomy caused few alterations in locations of axes of rotation and coupling patterns, whereas standard corpectomy with or without plating significantly altered these parameters in several instances. Conclusions. Multilevel oblique corpectomy ( without graft) induced significantly less instability and altered kinematics less than standard unplated corpectomy with graft. Multilevel oblique corpectomy allowed significantly more motion than standard plated corpectomy with graft. However, the goal of standard corpectomy is fusion. Our results indicate that plating significantly limits spinal mobility after 2-level corpectomy, improving the environment for fusion.
引用
收藏
页码:1420 / 1427
页数:8
相关论文
共 39 条
  • [1] CERVICAL STABILIZATION BY PLATE AND BONE FUSION
    BROWN, JA
    HAVEL, P
    EBRAHEIM, N
    GREENBLATT, SH
    JACKSON, WT
    [J]. SPINE, 1988, 13 (03) : 236 - 240
  • [2] ANTERIOR PLATE STABILIZATION FOR BURSTING TEARDROP FRACTURES OF THE CERVICAL-SPINE
    CABANELA, ME
    EBERSOLD, MJ
    [J]. SPINE, 1988, 13 (08) : 888 - 891
  • [3] Biomechanical effects of progressive anterior cervical decompression
    Chen, TY
    Crawford, NR
    Sonntag, VKH
    Dickman, CA
    [J]. SPINE, 2001, 26 (01) : 6 - 13
  • [4] Biomechanical evaluation of Caspar and Cervical Spine Locking Plate systems in a cadaveric model
    Clausen, JD
    Ryken, TC
    Traynelis, VC
    Sawin, PD
    Dexter, F
    Goel, VK
    [J]. JOURNAL OF NEUROSURGERY, 1996, 84 (06) : 1039 - 1045
  • [5] Uncinate processes and Luschka joints influence the biomechanics of the cervical spine: Quantification using a finite element model of the C5-C6 segment
    Clausen, JD
    Goel, VK
    Traynelis, VC
    Scifert, J
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 1997, 15 (03) : 342 - 347
  • [6] A new technique for determining 3-D joint angles: the tilt/twist method
    Crawford, NR
    Yamaguchi, GT
    Dickman, CA
    [J]. CLINICAL BIOMECHANICS, 1999, 14 (03) : 153 - 165
  • [7] AN APPARATUS FOR APPLYING PURE NONCONSTRAINING MOMENTS TO SPINE SEGMENTS IN-VITRO
    CRAWFORD, NR
    BRANTLEY, AGU
    DICKMAN, CA
    KOENEMAN, EJ
    [J]. SPINE, 1995, 20 (19) : 2097 - 2100
  • [8] Construction of local vertebral coordinate systems using a digitizing probe - Technical note
    Crawford, NR
    Dickman, CA
    [J]. SPINE, 1997, 22 (05) : 559 - 563
  • [9] Crawford NR, 1998, J SPINAL DISORD, V11, P416
  • [10] Anterior cervical plating reverses load transfer through multilevel strut-grafts
    DiAngelo, DJ
    Foley, KT
    Vossel, KA
    Rampersaud, YR
    Jansen, TH
    [J]. SPINE, 2000, 25 (07) : 783 - 795