摘要
Background: Tissue temperature monitoring during cutaneous laser therapy can lead to safer and more effective treatments. In this study, we investigate the use of speckle variance optical coherence tomography (svOCT) to monitor real-time temperature changes in the excised human skin tissue sample during laser irradiation. Methods: To accomplish this, we combined the pulse laser system with a reference-based svOCT system. To calibrate the svOCT, the ex-vivo human skin samples from three individuals with tissues collected from the arm, face, and back were heated with 1-degree increments. Additionally, linear regression was used to extract and evaluate the linear relationship between the temperature and normalized speckle variance value. Experiments were conducted on excised human skin sample to monitor the temperature change during laser therapy with a svOCT system. Thermal modeling of ex-vivo human skin was used to numerically simulate the laser-tissue interaction and estimate the thermal diffusion and peak temperature of the tissue during the laser treatment. Results and Conclusion: These results showed that normalized speckle variance had a linear relationship with the tissue temperature before the onset of tissue coagulation (52 degrees) and we were able to measure the rapid increase of the tissue temperature during laser therapy. The result of the experiment is also in good agreement with the numerical simulation result that estimated the laser-induced peak temperature and thermal relaxation time.
引用
收藏
页数:8
相关论文
共 33 条
[1]   SELECTIVE PHOTOTHERMOLYSIS - PRECISE MICROSURGERY BY SELECTIVE ABSORPTION OF PULSED RADIATION [J].
ANDERSON, RR ;
PARRISH, JA .
SCIENCE, 1983, 220 (4596) :524-527
[2]   TEMPERATURE RESPONSE OF BIOLOGICAL-MATERIALS TO PULSED NON-ABLATIVE CO2-LASER IRRADIATION [J].
BRUGMANS, MJP ;
KEMPER, J ;
GIJSBERS, GHM ;
VANDERMEULEN, FW ;
VANGEMERT, MJC .
LASERS IN SURGERY AND MEDICINE, 1991, 11 (06) :587-594
[3]   Analysis of thermal relaxation during laser irradiation of tissue [J].
Choi, B ;
Welch, AJ .
LASERS IN SURGERY AND MEDICINE, 2001, 29 (04) :351-359
[4]  
COOPER TE, 1971, AEROSPACE MED, V42, P24
[5]  
Fok D. C. C., 2017, PLAST AESTHETIC RES, V4, P1
[6]   Optical coherence tomography in dermatology: technical and clinical aspects [J].
Gambichler, Thilo ;
Jaedicke, Volker ;
Terras, Sarah .
ARCHIVES OF DERMATOLOGICAL RESEARCH, 2011, 303 (07) :457-473
[7]  
Goldberg David J, 2012, J Clin Aesthet Dermatol, V5, P45
[8]   Accurate prediction of collapse temperature using optical coherence tomography-based freeze-drying microscopy [J].
Greco, Kristyn ;
Mujat, Mircea ;
Galbally-Kinney, Kristin L. ;
Hammer, Daniel X. ;
Ferguson, R. Daniel ;
Iftimia, Nicusor ;
Mulhall, Phillip ;
Sharma, Puneet ;
Kessler, William J. ;
Pikal, Michael J. .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2013, 102 (06) :1773-1785
[9]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[10]   Correlation of temperature rise and optical coherence tomography characteristics in patient retinal photocoagulation [J].
Koinzer, Stefan ;
Schlott, Kerstin ;
Portz, Lea ;
Ptaszynski, Lars ;
Baade, Alexander ;
Bever, Marco ;
Saeger, Mark ;
Caliebe, Amke ;
Denner, Rene ;
Birngruber, Reginald ;
Brinkmann, Ralf ;
Roider, Johann .
JOURNAL OF BIOPHOTONICS, 2012, 5 (11-12) :889-902