Comparative Study of Underwater Single Pulse and Orthogonal Double Pulse Laser-Induced Breakdown Spectroscopy of Barium Element in Solution

被引:9
作者
Li Wenping [1 ,2 ]
Zhou Weidong [1 ,2 ]
机构
[1] Zhejiang Normal Univ, Coll Phys & Elect Informat Engn, Jinhua 321004, Zhejiang, Peoples R China
[2] Key Lab Opt Informat Detecting & Display Technol, Jinhua 321004, Zhejiang, Peoples R China
来源
CHINESE JOURNAL OF LASERS-ZHONGGUO JIGUANG | 2019年 / 46卷 / 09期
关键词
spectroscopy; single pulse laser-induced breakdown spectroscopy; double pulse laser-induced breakdown spectroscopy; underwater; laser energy; detection sensitivity; METAL ELEMENTS; WATER; PLASMA; NANOSECOND; LIBS; EMISSION; LIQUIDS;
D O I
10.3788/CJL201946.0911003
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Two Q -switched Nd YAG lasers operating at 1061 nm arc used and combined in orthogonal beam geometry to perform underwater single and double pulse laser-induced breakdown spectroscopy (SP-LIBS and DPLIES) experiments in barium chloride (BaCl2) solution. The influences of key experimental parameters on the spectral characteristics and signal enhancement of underwater SP-LIBS and DP-LIBS arc investigated systematically. There is a time window from approximately 500 ns to 900 ns in underwater DP-LIBS, where signal-to-noise ratios arc very high. The time window favors the selection of delay time for spectral acquisition and analysis. By properly setting the relative position of two laser foci and using optimized laser energy, a 20 -time enhancement of the Ba II 155.1 nm line intensity of DP-LIBS over the maximal intensity of SP-LIBS is obtained. The spectral line of DP-LIBS is narrower and last longer than that of SP-LIBS. In addition, dependence of spectral line intensity of DP-LIBS on the second laser energy is observed, which demonstrates an exponential increase. This is quite different to the linear growth observed in collinear DP-LIBS experiment on the liquid surface, and the difference in growth trend may be related to the gaseous environment in which the second laser induced plasma is produced and confined. Finally, the detection capability of underwater SP-LIBS and DP-LIBS is compared. The results show that the detection sensitivity of DP-LIBS is improved by a factor of 37, and the limit of detection of Ba in bulk water is reduced from 31.35X 10(-6) in SP-LIBS to 1.78X 10(-6) in DP-LIBS.
引用
收藏
页数:10
相关论文
共 32 条
[1]   Material ejection in nanosecond Er:YAG laser ablation of water, liver, and skin [J].
Apitz, I ;
Vogel, A .
APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 2005, 81 (02) :329-338
[2]   Sensitive detection of metals in water using laser-induced breakdown spectroscopy on wood sample substrates [J].
Chen, Zhijiang ;
Godwal, Yogesh ;
Tsui, Ying Yin ;
Fedosejevs, Robert .
APPLIED OPTICS, 2010, 49 (13) :C87-C94
[3]   SPECTROCHEMICAL ANALYSIS OF LIQUIDS USING THE LASER SPARK [J].
CREMERS, DA ;
RADZIEMSKI, LJ ;
LOREE, TR .
APPLIED SPECTROSCOPY, 1984, 38 (05) :721-729
[4]   Double pulse laser produced plasma on metallic target in seawater: basic aspects and analytical approach [J].
De Giacomo, A ;
Dell'Aglio, M ;
Colao, F ;
Fantoni, R .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2004, 59 (09) :1431-1438
[5]   From single pulse to double pulse ns-Laser Induced Breakdown Spectroscopy under water: Elemental analysis of aqueous solutions and submerged solid samples [J].
De Giacomo, A. ;
Dell'Aglio, M. ;
De Pascale, O. ;
Capitelli, M. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2007, 62 (08) :721-738
[6]   Single Pulse-Laser Induced Breakdown Spectroscopy in aqueous solution [J].
De Giacomo, A ;
Dell'Aglio, M ;
De Pascale, O .
APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 2004, 79 (4-6) :1035-1038
[7]   STUDY OF THE TEMPORAL AND SPATIAL DYNAMICS OF PLASMAS INDUCED IN LIQUIDS BY NANOSECOND ND-YAG LASER-PULSES .1. ANALYSIS OF THE PLASMA STARTING TIMES [J].
DOCCHIO, F ;
REGONDI, P ;
CAPON, MRC ;
MELLERIO, J .
APPLIED OPTICS, 1988, 27 (17) :3661-3668
[8]  
Hou HM, 2014, J ANAL ATOM SPECTROM, V29, P169, DOI [10.1039/c3ja50244a, 10.1039/C3JA50244A]
[9]   Continuous Online Detection Method of Heavy Metals in Water Based on LIBS Technology [J].
Jia Yao ;
Zhao Nanjing ;
Liu Wenqing ;
Fang Li ;
Ma Mingjun ;
Meng Deshuo .
CHINESE JOURNAL OF LASERS-ZHONGGUO JIGUANG, 2018, 45 (06)
[10]   Laser induced breakdown spectroscopy (LIBS) as an analytical tool for the detection of metal ions in aqueous solutions [J].
Knopp, R ;
Scherbaum, FJ ;
Kim, JI .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1996, 355 (01) :16-20