Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device

被引:96
作者
Pasma, KL [1 ]
Dirkx, MLP [1 ]
Kroonwijk, M [1 ]
Visser, AG [1 ]
Heijmen, BJM [1 ]
机构
[1] Univ Rotterdam Hosp, Dr Daniel Den Hoed Canc Ctr, Dept Radiotherapy, Div Clin Phys, NL-3075 EA Rotterdam, Netherlands
关键词
intensity modulation; dynamic multileaf collimation; quality assurance; portal imaging; EPID;
D O I
10.1118/1.598752
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dose distributions can often be significantly improved by modulating the two-dimensional intensity profile of the individual x-ray beams. One technique for delivering intensity modulated beams is dynamic multileaf collimation (DMLC). However, DMLC is complex and requires extensive quality assurance. In this paper a new method is presented for a pretreatment dosimetric verification of these intensity modulated beams utilizing a charge-coupled device camera based fluoroscopic electronic portal imaging device (EPID). In the absence of the patient, EPID images are acquired for all beams produced with DMLC. These images are then converted into two-dimensional dose distributions and compared with the calculated dose distributions. The calculations are performed with a pencil beam algorithm as implemented in a commercially available treatment planning system using the same absolute beam fluence profiles as used for calculation of the patient dose distribution. The method allows an overall verification of (i) the leaf trajectory calculation (including the models to incorporate collimator scatter and leaf transmission), (ii) the correct transfer of the leaf sequencing file to the treatment machine, and (iii) the mechanical and dosimetrical performance of the treatment unit. The method was tested for intensity modulated 10 and 25 MV photon beams; both model cases and real clinical cases were studied. Dose profiles measured with the EPID were also compared with ionization chamber measurements. In all cases both predictions and EPID measurements and EPID and ionization chamber measurements agreed within 2% (1 sigma). The study has demonstrated that the proposed method allows fast and accurate pretreatment verification of DMLC. (C) 1999 American Association of Physicists in Medicine. [S0093-2405(99)01411-X].
引用
收藏
页码:2373 / 2378
页数:6
相关论文
共 38 条
[11]   Leaf trajectory calculation for dynamic multileaf collimation to realize optimized fluence profiles [J].
Dirkx, MLP ;
Heijmen, BJM ;
van Santvoort, JPC .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (05) :1171-1184
[12]  
DIRKX MLP, 1998, RADIOTHER ONCOL, V48, pS186
[13]   PORTAL DOSE MEASUREMENT IN RADIOTHERAPY USING AN ELECTRONIC PORTAL IMAGING DEVICE (EPID) [J].
HEIJMEN, BJM ;
PASMA, KL ;
KROONWIJK, M ;
ALTHOF, VGM ;
DEBOER, JCJ ;
VISSER, AG ;
HUIZENGA, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) :1943-1955
[14]   SHAPING OF ARBITRARY DOSE DISTRIBUTIONS BY DYNAMIC MULTILEAF COLLIMATION [J].
KALLMAN, P ;
LIND, B ;
EKLOF, A ;
BRAHME, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (11) :1291-1300
[15]   AN ALGORITHM FOR MAXIMIZING THE PROBABILITY OF COMPLICATION-FREE TUMOR-CONTROL IN RADIATION-THERAPY [J].
KALLMAN, P ;
LIND, BK ;
BRAHME, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (04) :871-890
[16]   In vivo dosimetry for prostate cancer patients using an electronic portal imaging device (EPID); demonstration of internal organ motion [J].
Kroonwijk, M ;
Pasma, KL ;
Quint, S ;
Koper, PCM ;
Visser, AG ;
Heijmen, BJM .
RADIOTHERAPY AND ONCOLOGY, 1998, 49 (02) :125-132
[17]  
Kroonwijk M, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P276
[18]   Conformal radiation treatment of prostate cancer using inversely-planned intensity-modulated photon beams produced with dynamic multileaf collimation [J].
Ling, CC ;
Burman, C ;
Chui, CS ;
Kutcher, GJ ;
Leibel, SA ;
LoSasso, T ;
Mohan, R ;
Bortfeld, T ;
Reinstein, L ;
Spirou, S ;
Wang, XH ;
Wu, QW ;
Zelefsky, M ;
Fuks, Z .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (04) :721-730
[19]   Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 1998, 25 (10) :1919-1927
[20]   Quality assurance for dynamic multileaf collimator modulated fields using a fast beam imaging system [J].
Ma, LJ ;
Geis, PB ;
Boyer, AL .
MEDICAL PHYSICS, 1997, 24 (08) :1213-1220