Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device

被引:96
作者
Pasma, KL [1 ]
Dirkx, MLP [1 ]
Kroonwijk, M [1 ]
Visser, AG [1 ]
Heijmen, BJM [1 ]
机构
[1] Univ Rotterdam Hosp, Dr Daniel Den Hoed Canc Ctr, Dept Radiotherapy, Div Clin Phys, NL-3075 EA Rotterdam, Netherlands
关键词
intensity modulation; dynamic multileaf collimation; quality assurance; portal imaging; EPID;
D O I
10.1118/1.598752
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dose distributions can often be significantly improved by modulating the two-dimensional intensity profile of the individual x-ray beams. One technique for delivering intensity modulated beams is dynamic multileaf collimation (DMLC). However, DMLC is complex and requires extensive quality assurance. In this paper a new method is presented for a pretreatment dosimetric verification of these intensity modulated beams utilizing a charge-coupled device camera based fluoroscopic electronic portal imaging device (EPID). In the absence of the patient, EPID images are acquired for all beams produced with DMLC. These images are then converted into two-dimensional dose distributions and compared with the calculated dose distributions. The calculations are performed with a pencil beam algorithm as implemented in a commercially available treatment planning system using the same absolute beam fluence profiles as used for calculation of the patient dose distribution. The method allows an overall verification of (i) the leaf trajectory calculation (including the models to incorporate collimator scatter and leaf transmission), (ii) the correct transfer of the leaf sequencing file to the treatment machine, and (iii) the mechanical and dosimetrical performance of the treatment unit. The method was tested for intensity modulated 10 and 25 MV photon beams; both model cases and real clinical cases were studied. Dose profiles measured with the EPID were also compared with ionization chamber measurements. In all cases both predictions and EPID measurements and EPID and ionization chamber measurements agreed within 2% (1 sigma). The study has demonstrated that the proposed method allows fast and accurate pretreatment verification of DMLC. (C) 1999 American Association of Physicists in Medicine. [S0093-2405(99)01411-X].
引用
收藏
页码:2373 / 2378
页数:6
相关论文
共 38 条
[1]   Physical characteristics of a commercial electronic portal imaging device [J].
Althof, VGM ;
deBoer, JCJ ;
Huizenga, H ;
Stroom, JC ;
Visser, AG ;
Swanenburg, BN .
MEDICAL PHYSICS, 1996, 23 (11) :1845-1855
[2]  
BALTER JM, 1997, MED PHYS, V24, P1002
[3]   METHODS OF IMAGE-RECONSTRUCTION FROM PROJECTIONS APPLIED TO CONFORMATION RADIOTHERAPY [J].
BORTFELD, T ;
BURKELBACH, J ;
BOESECKE, R ;
SCHLEGEL, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) :1423-1434
[4]   X-RAY FIELD COMPENSATION WITH MULTILEAF COLLIMATORS [J].
BORTFELD, TR ;
KAHLER, DL ;
WALDRON, TJ ;
BOYER, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03) :723-730
[5]   OPTIMIZATION OF STATIONARY AND MOVING BEAM RADIATION-THERAPY TECHNIQUES [J].
BRAHME, A .
RADIOTHERAPY AND ONCOLOGY, 1988, 12 (02) :129-140
[6]   SOLUTION OF AN INTEGRAL-EQUATION ENCOUNTERED IN ROTATION THERAPY [J].
BRAHME, A ;
ROOS, JE ;
LAX, I .
PHYSICS IN MEDICINE AND BIOLOGY, 1982, 27 (10) :1221-1229
[7]   Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate [J].
Burman, C ;
Chui, CS ;
Kutcher, G ;
Leibel, S ;
Zelefsky, M ;
LoSasso, T ;
Spirou, S ;
Wu, QW ;
Yang, J ;
Stein, J ;
Mohan, R ;
Fuks, Z ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (04) :863-873
[8]   THE GENERATION OF INTENSITY-MODULATED FIELDS FOR CONFORMAL RADIOTHERAPY BY DYNAMIC COLLIMATION [J].
CONVERY, DJ ;
ROSENBLOOM, ME .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (06) :1359-1374
[9]   Verification of segmented beam delivery using a commercial electronic portal imaging device [J].
Curtin-Savard, AJ ;
Podgorsak, EB .
MEDICAL PHYSICS, 1999, 26 (05) :737-742
[10]   DAILY DOSIMETRIC QUALITY-CONTROL OF THE MM50 RACETRACK MICROTRON USING AN ELECTRONIC PORTAL IMAGING DEVICE [J].
DIRKX, MLP ;
KROONWIJK, M ;
DEBOER, JCJ ;
HEIJMEN, BJM .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (01) :55-60