On the optimization of bone SPECT/CT in terms of image quality and radiation dose

被引:6
作者
Tulik, Monika [1 ]
Tulik, Piotr [2 ]
Kowalska, Teresa [1 ]
机构
[1] Maria Sklodowska Curie Natl Res Inst Oncol, Krakow Branch, Krakow, Poland
[2] Warsaw Univ Technol, Inst Metrol & Biomed Engn, Fac Mechatron, Warsaw, Poland
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2020年 / 21卷 / 11期
关键词
bone SPECT; CT; image quality; optimization; radiation dose; ATTENUATION CORRECTION; ITERATIVE RECONSTRUCTION; CT; PET;
D O I
10.1002/acm2.13069
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction The purpose of this study was to present the optimization process of CT parameters to reduce patient exposure during bone SPECT/CT without affecting the quality of SPECT images with attenuation correction (AC). Material and methods A fillable phantom reflecting realistic bone scintigraphy conditions was developed and acquired on an AnyScan SC. SPECT/CT scans were carried out with different x-ray tube current values (10, 20, 30, 40, 50, 60, 70, 90, 110, 130, 150, and 200 mA) at three different high-voltage values (80, 100, and 120 kV). The contrast (C) and coefficients of variation (CV) in the SPECT images as well as the signal-to-noise ratio (SNR) and noise (SDCT) in the CT images with CTDIvol were measured. An optimal acquisition protocol that obtained SPECT/CT images with no artifacts on both CT and SPECT images, acceptable C, SNR, CV, and SDCT values, and the largest reduction in patient exposure compared to the reference acquisition procedure was sought. Results The optimal set of parameters for bone SPECT/CT was determined based on a phantom study. It has been implemented in clinical practice. Two groups of patients were examined according to the baseline and optimized protocols, respectively. The new SPECT/CT protocol substantially reduced patients' radiation exposure compared to the old protocol while maintaining the required diagnostic quality of SPECT and CT images. Conclusions In the study, we present a methodology that finds a compromise between diagnostic information and patient exposure during bone SPECT/CT procedures.
引用
收藏
页码:237 / 246
页数:10
相关论文
共 39 条
[1]   Accuracy of CT-based attenuation correction in PET/CT bone imaging [J].
Abella, Monica ;
Alessio, Adam M. ;
Mankoff, David A. ;
MacDonald, Lawrence R. ;
Jose Vaquero, Juan ;
Desco, Manuel ;
Kinahan, Paul E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (09) :2477-2490
[2]  
[Anonymous], 2009, Computed tomography: principles, design, artifacts, and recent advances
[3]  
Blocklet D, 1999, J NUCL MED, V40, P1978
[4]  
Boone J.M., 2011, SIZE SPECIFIC DOSE E
[5]   Radiation risk and protection of patients in clinical SPECT/CT [J].
Brix, Gunnar ;
Nekolla, Elke A. ;
Borowski, Markus ;
Nosske, Dietmar .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 :S125-S136
[6]   SPECT/CT [J].
Buck, Andreas K. ;
Nekolla, Stephan ;
Ziegler, Sibylle ;
Beer, Ambros ;
Krause, Bernd J. ;
Herrmann, Ken ;
Scheidhauer, Klemens ;
Wester, Hans-Juergen ;
Rummeny, Ernst J. ;
Schwaiger, Markus ;
Drzezga, Alexander .
JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (08) :1305-1319
[7]  
Case JA, 1999, J NUCL MED, V40, P604
[8]   Functional and Hybrid Imaging of Bone Metastases [J].
Cook, Gary J. R. ;
Goh, Vicky .
JOURNAL OF BONE AND MINERAL RESEARCH, 2018, 33 (06) :961-972
[9]   Bone equivalent liquid solution to assess accuracy of transmission measurements in SPECT and PET [J].
deDreuille, O ;
Strijckmans, V ;
Almeida, P ;
Loch, C ;
Bendriem, B .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1186-1190
[10]   Optimization of the CT component of SPECT-CT and establishment of local CT diagnostic reference levels for clinical practice [J].
Dennis, Jennifer L. ;
Gemmell, Alastair J. ;
Nicol, Alice J. .
NUCLEAR MEDICINE COMMUNICATIONS, 2018, 39 (06) :493-499