Development of 15 MHz 2-2 piezo-composite ultrasound linear array transducers for ophthalmic imaging

被引:35
作者
Cha, Jung Hyui [1 ]
Chang, Jin Ho [1 ,2 ,3 ]
机构
[1] Sogang Univ, Interdisciplinary Program Integrated Biotechnol, Seoul 121742, South Korea
[2] Sogang Univ, Sogang Inst Adv Technol, Seoul 121742, South Korea
[3] Sogang Univ, Dept Elect Engn, Seoul 121742, South Korea
关键词
High frequency ultrasound; Ophthalmic imaging; Linear array transducer; 2-2; piezo-composite; Anterior segment; Posterior segment; DESIGN;
D O I
10.1016/j.sna.2014.06.024
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
High frequency (>15 MHz) ultrasound imaging is capable of producing high spatial resolution B-mode and color Doppler images for the efficient diagnosis of ocular diseases. However, commercial high frequency ultrasound imaging systems cannot currently provide its full capability for ophthalmic applications because either single element or annular array transducers are used in conjunction with mechanical scanning. To overcome the current limitations, 15 MHz 128-element linear array transducers were developed, which are suitable to cover the whole eye from the eyelid in anterior segment to orbital fundus in posterior segment. To achieve the highest electromechanical coupling coefficient, 2-2 piezo-composite was designed and fabricated as an active layer. For the design, the simple model of the thickness-mode oscillation was derived and used in the 1-D model simulator such as PiezoCAD. Also, the usability of the simple model was verified by comparing its simulation results with those by finite element model-based PZFlex. In addition, easy fabrication procedure for the high frequency 2-2 piezo-composite arrays was proposed. In performance evaluation, the pulse-echo response of a representative composite element showed the center frequency of 14.4 MHz and the -6 dB fractional bandwidth of 69.4%. Measured insertion loss was achieved higher than -33.9 dB and crosstalk level between elements less than -30 dB. From the evaluation results, it is expected that the linear array transducer developed using the proposed design and fabrication methods can provide a potential alternative for high frequency ophthalmic imaging with the desired capability. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:39 / 48
页数:10
相关论文
共 28 条
[1]  
Abboud N., 1998, SPIE INT S MED IM
[2]  
Abramoff Michael D, 2010, IEEE Rev Biomed Eng, V3, P169, DOI 10.1109/RBME.2010.2084567
[3]   Development of a 35-MHz piezo-composite ultrasound array for medical imaging [J].
Cannata, JM ;
Williams, JA ;
Zhou, QF ;
Ritter, TA ;
Shung, KK .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (01) :224-236
[4]  
Cannata JM, 2003, ULTRASON, P1658
[5]   A High-Frequency Linear Ultrasonic Array Utilizing an Interdigitally Bonded 2-2 Piezo-Composite [J].
Cannata, Jonathan M. ;
Williams, Jay A. ;
Zhang, Lequan ;
Hu, Chang-Hong ;
Shung, K. Kirk .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (10) :2202-2212
[6]  
Cattaneo P, 2007, TEX HEART I J, V34, P105
[7]   A High-Frequency Annular-Array Transducer Using an Interdigital Bonded 1-3 Composite [J].
Chabok, Hamid Reza ;
Cannata, Jonathan M. ;
Kim, Hyung Ham ;
Williams, Jay A. ;
Park, Jinhyoung ;
Shung, K. Kirk .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (01) :206-214
[8]  
Coleman D., 2006, ULTRASONOGRAPHY EYE
[9]   DESIGN OF EFFICIENT BROAD-BAND PIEZOELECTRIC TRANSDUCERS [J].
DESILETS, CS ;
FRASER, JD ;
KINO, GS .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1978, 25 (03) :115-125
[10]   Ultrasound in ophthalmology [J].
Fledelius, HC .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (03) :365-375