Integrated PET/MRI scanner with oxygen-15 labeled gases for quantification of cerebral blood flow, cerebral blood volume, cerebral oxygen extraction fraction and cerebral metabolic rate of oxygen

被引:4
作者
Ito, Hiroshi [1 ,2 ]
Kubo, Hitoshi [2 ]
Takahashi, Kazuhiro [2 ]
Nishijima, Ken-Ichi [2 ]
Ukon, Naoyuki [2 ]
Nemoto, Ayaka [2 ]
Sugawara, Shigeyasu [1 ,2 ]
Yamakuni, Ryo [1 ]
Ibaraki, Masanobu [3 ]
Ishii, Shiro [1 ]
机构
[1] Fukushima Med Univ, Dept Radiol & Nucl Med, 1 Hikariga Oka, Fukushima 9601295, Japan
[2] Fukushima Med Univ, Adv Clin Res Ctr, Fukushima, Japan
[3] Akita Res Inst Brain & Blood Vessels, Dept Radiol & Nucl Med, Akita, Japan
基金
日本学术振兴会;
关键词
PET; MRI; CBF; CMRO2; Oxygen-15; Brain;
D O I
10.1007/s12149-021-01578-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives Measurement of cerebral blood flow (CBF), cerebral blood volume (CBV), cerebral oxygen extraction fraction (OEF) and cerebral metabolic rate of oxygen (CMRO2) by PET with oxygen-15 labeled gases is useful for diagnosis and treatment planning in cases of chronic occlusive cerebrovascular disease. In the present study, CBF, CBV, OEF and CMRO2 were measured using the integrated design of PET/MRI scanner system. This is a first attempt to measure cerebral perfusion and oxygen metabolism using PET/MRI with oxygen-15 labeled gases. Methods PET/MRI measurements with the steady-state method of oxygen-15 labeled gases, carbon monoxide ((CO)-O-15), oxygen (O-15(2)), and carbon dioxide ((CO2)-O-15) were performed on nine healthy men. Two kinds of attenuation correction for PET were performed using MRI with Dixon sequence (DIXON) and Dixon sequence with model-based bone segmentation (DIXONbone). A real-time motion correction of PET images was also performed using simultaneously measured MR images to detect head motion. Results Mean and SD values of CBF, CBV, OEF, and CMRO2 in the cerebral cortices with attenuation correction by DIXON were 31 +/- 4 mL/100 mL/min, 2.7 +/- 0.2 mL/mL, 0.40 +/- 0.07, and 2.5 +/- 0.3 mL/100 mL/min without real-time motion correction, and 33 +/- 4 mL/100 mL/min, 2.7 +/- 0.2 mL/mL, 0.40 +/- 0.07, and 2.6 +/- 0.3 mL/100 mL/min with real-time motion correction, respectively. Values with of CBF, CBV, OEF, and CMRO2 with attenuation correction by DIXONbone were 35 +/- 5 mL/100 mL/min, 2.8 +/- 0.2 mL/mL, 0.40 +/- 0.07, and 2.8 +/- 0.3 mL/100 mL/min without real-time motion correction, and 38 +/- 5 mL/100 mL/min, 2.8 +/- 0.2 mL/mL, 0.40 +/- 0.07, and 3.0 +/- 0.4 mL/100 mL/min with real-time motion correction, respectively. Conclusions Using PET/MRI with oxygen-15 labeled gases, CBF, CBV, OEF, and CMRO2 could be measured. Values of CBF, CBV, and CMRO2 measured with attenuation correction by DIXON were significantly lower than those measured with correction by DIXONbone. One of the reasons for this is that attenuation correction of DIXON does not take into consideration of the photon absorption by bone. OEF values, corresponding to ratios of CMRO2 to CBF, were not affected by attenuation correction methods. Values of CBF and CMRO2 with a real-time motion correction were significantly higher than those without correction. Using PET/MRI with adequate corrections, similar values of CBF, CBV, OEF, and CMRO2 as PET alone scanner system reported previously were obtained. Trail registration The UMIN clinical trial number: UMIN000033382.
引用
收藏
页码:421 / 428
页数:8
相关论文
共 27 条
  • [1] PET/MR in Dementia and Other Neurodegenerative Diseases
    Barthel, Henryk
    Schroeter, Matthias L.
    Hoffmann, Karl-Titus
    Sabri, Osama
    [J]. SEMINARS IN NUCLEAR MEDICINE, 2015, 45 (03) : 224 - 233
  • [2] MRI-Assisted PET Motion Correction for Neurologic Studies in an Integrated MR-PET Scanner
    Catana, Ciprian
    Benner, Thomas
    van der Kouwe, Andre
    Byars, Larry
    Hamm, Michael
    Chonde, Daniel B.
    Michel, Christian J.
    El Fakhri, Georges
    Schmand, Matthias
    Sorensen, Gregory
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (01) : 154 - 161
  • [3] Delso G, 2011, J NUCL MED, V52, P1914, DOI 10.2967/jnumed.111.092726
  • [4] PET/MRI system design
    Delso, Gaspar
    Ziegler, Sibylle
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 : 86 - 92
  • [5] Cerebral hemodynamic impairment - Methods of measurement and association with stroke risk
    Derdeyn, CP
    Grubb, RL
    Powers, WJ
    [J]. NEUROLOGY, 1999, 53 (02) : 251 - 259
  • [6] A comparison of CT- and MR-based attenuation correction in neurological PET
    Dickson, John C.
    O'Meara, Celia
    Barnes, Anna
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 (06) : 1176 - 1189
  • [7] QUANTITATIVE MEASUREMENT OF REGIONAL CEREBRAL BLOOD-FLOW AND OXYGEN-METABOLISM IN MAN USING O-15 AND POSITRON EMISSION TOMOGRAPHY - THEORY, PROCEDURE, AND NORMAL VALUES
    FRACKOWIAK, RSJ
    LENZI, GL
    JONES, T
    HEATHER, JD
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1980, 4 (06) : 727 - 736
  • [8] GIBBS JM, 1984, LANCET, V1, P310
  • [9] A systematic performance evaluation of head motion correction techniques for3 commercial PET scanners using a reproducible experimental acquisition protocol
    Inomata, Takato
    Watanuki, Shoichi
    Odagiri, Hayato
    Nambu, Takeyuki
    Karakatsanis, Nicolas A.
    Ito, Hiroshi
    Watabe, Hiroshi
    Tashiro, Manabu
    Shidahara, Miho
    [J]. ANNALS OF NUCLEAR MEDICINE, 2019, 33 (07) : 459 - 470
  • [10] Error analysis of measured cerebral vascular response to acetazolamide stress by I-123-IMP autoradiographic method with single photon emission computed tomography: Errors due to distribution volume of I-123-IMP
    Ito, H
    Inoue, K
    Goto, R
    Kinomura, S
    Sato, T
    Kaneta, T
    Sato, K
    Fukuda, H
    [J]. ANNALS OF NUCLEAR MEDICINE, 2004, 18 (03) : 221 - 226