Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: a cadaveric study

被引:41
作者
Lamartina, Claudio [1 ]
Cecchinato, Riccardo [1 ]
Fekete, Zsolt [2 ]
Lipari, Alberto [3 ]
Fiechter, Meinrad [3 ]
Berjano, P. [1 ]
机构
[1] IRCCS Ist Ortoped Galeazzi, Milan, Italy
[2] Neuro & Wirbelsaulenzentrum Zent Schweiz, Luzern, Switzerland
[3] Medacta Int SA, CH-6874 Castel San Pietro, Switzerland
关键词
Spine; Surgery; Technique; in vitro; Pedicle screws; Patient-specific instrumentation; FREE-HAND TECHNIQUE; OF-THE-LITERATURE; POSTERIOR INSTRUMENTATION; THORACOLUMBAR SPINE; SCOLIOSIS SURGERY; NAVIGATION; FIXATION; METAANALYSIS; GUIDANCE; SYSTEMS;
D O I
10.1007/s00586-015-4261-y
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose Pedicle screw placement is an increasingly common procedure for the correction of spine degenerative disease, deformity and trauma. However, screw placement is demanding, with complications resulting from inaccurate screw placement. While several different techniques have been developed to improve accuracy, they all have their limitations. Methods We examined the MySpine (Medacta International SA, Castel San Pietro, CH) patient-matched pedicle targeting guide in three cadaveric spine specimens operated on by three surgeons. A three-dimensional (3D) preoperative plan was constructed from spinal computed tomography scans, from which individualised guides were developed for the placement of Medacta Unconstrained Screw Technology pedicle screws. Following screw placement, the 3D positioning of the screws was compared to the preoperative plan against a series of pre-defined criteria. Results Of 46 inserted screws eligible for assessment, 91.3 % were fully inside the pedicle. There were no cases of Grade B (2-4 mm) or C (>4 mm) pedicle perforation. The mean deviation between the planned and actual screw position at the midpoint of the pedicle was 0.70 mm, the mean horizontal deviation was 0.60 mm and the mean vertical deviation was 0.77 mm. The mean angular deviation in the sagittal plane was 1.74 degrees, versus 1.32 degrees in the transverse plane. The mean deviation in screw depth was 1.55 mm. On all measures, the accuracy of screw placement was within the predefined criteria. Conclusions Our cadaver study indicates that pedicle screw placement with the system is accurate and should be investigated in larger in vitro and in vivo studies.
引用
收藏
页码:S937 / S941
页数:5
相关论文
共 32 条
[11]  
Kim YJJ, 2005, NEUROL INDIA, V53, P512
[12]   Pedicle screw placement accuracy - A meta-analysis [J].
Kosmopoulos, Victor ;
Schizas, Constantin .
SPINE, 2007, 32 (03) :E111-E120
[13]   Adolescent idiopathic scoliosis surgery with patient-specific screw placement-guides [J].
Lamartina, C. ;
Capuzzo, A. ;
Cecchinato, R. ;
Zerbi, A. ;
Berjano, P. .
EUROPEAN SPINE JOURNAL, 2014, 23 (12) :2765-2766
[14]  
Landi A, 2015, ORTHOP MUSCULAR SYST, V4, pe116, DOI [10.4127/2161-0533.1000e116, DOI 10.4127/2161-0533.1000E116]
[15]   Complications associated with pedicle screws [J].
Lonstein, JE ;
Denis, F ;
Perra, JH ;
Pinto, MR ;
Smith, MD ;
Winter, RB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (11) :1519-1528
[16]   Efficacy and accuracy of a novel rapid prototyping drill template for cervical pedicle screw placement [J].
Lu, Sheng ;
Xu, Yong Q. ;
Chen, Guo P. ;
Zhang, Yuan Z. ;
Lu, Di ;
Chen, Yu B. ;
Shi, Ji H. ;
Xu, Xing M. .
COMPUTER AIDED SURGERY, 2011, 16 (05) :240-248
[17]   The accuracy of pedicle screw placement using intraoperative image guidance systems [J].
Mason, Alexander ;
Paulsen, Renee ;
Babuska, Jason M. ;
Rajpal, Sharad ;
Burneikiene, Sigita ;
Nelson, E. Lee ;
Villavicencio, Alan T. .
JOURNAL OF NEUROSURGERY-SPINE, 2014, 20 (02) :196-203
[18]   Fluoroscopy-based navigation system in spine surgery [J].
Merloz, P. ;
Troccaz, J. ;
Vottaillat, H. ;
Vasile, C. ;
Tonetti, I. ;
Eid, A. ;
Plaweski, S. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H7) :813-820
[19]   Computer-assisted surgical planning and image-guided surgical navigation in refractory adult scoliosis surgery - Case report and review of the literature [J].
Metz, Lionel N. ;
Burch, Shane .
SPINE, 2008, 33 (09) :E287-E292
[20]   Accuracy of thoracic pedicle screw placement in scoliosis using the ideal pedicle entry point during the freehand technique [J].
Modi, Hitesh ;
Suh, Seung Woo ;
Song, Hae-Ryong ;
Yang, Jae-Hyuk .
INTERNATIONAL ORTHOPAEDICS, 2009, 33 (02) :469-475