Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: a cadaveric study

被引:41
作者
Lamartina, Claudio [1 ]
Cecchinato, Riccardo [1 ]
Fekete, Zsolt [2 ]
Lipari, Alberto [3 ]
Fiechter, Meinrad [3 ]
Berjano, P. [1 ]
机构
[1] IRCCS Ist Ortoped Galeazzi, Milan, Italy
[2] Neuro & Wirbelsaulenzentrum Zent Schweiz, Luzern, Switzerland
[3] Medacta Int SA, CH-6874 Castel San Pietro, Switzerland
关键词
Spine; Surgery; Technique; in vitro; Pedicle screws; Patient-specific instrumentation; FREE-HAND TECHNIQUE; OF-THE-LITERATURE; POSTERIOR INSTRUMENTATION; THORACOLUMBAR SPINE; SCOLIOSIS SURGERY; NAVIGATION; FIXATION; METAANALYSIS; GUIDANCE; SYSTEMS;
D O I
10.1007/s00586-015-4261-y
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose Pedicle screw placement is an increasingly common procedure for the correction of spine degenerative disease, deformity and trauma. However, screw placement is demanding, with complications resulting from inaccurate screw placement. While several different techniques have been developed to improve accuracy, they all have their limitations. Methods We examined the MySpine (Medacta International SA, Castel San Pietro, CH) patient-matched pedicle targeting guide in three cadaveric spine specimens operated on by three surgeons. A three-dimensional (3D) preoperative plan was constructed from spinal computed tomography scans, from which individualised guides were developed for the placement of Medacta Unconstrained Screw Technology pedicle screws. Following screw placement, the 3D positioning of the screws was compared to the preoperative plan against a series of pre-defined criteria. Results Of 46 inserted screws eligible for assessment, 91.3 % were fully inside the pedicle. There were no cases of Grade B (2-4 mm) or C (>4 mm) pedicle perforation. The mean deviation between the planned and actual screw position at the midpoint of the pedicle was 0.70 mm, the mean horizontal deviation was 0.60 mm and the mean vertical deviation was 0.77 mm. The mean angular deviation in the sagittal plane was 1.74 degrees, versus 1.32 degrees in the transverse plane. The mean deviation in screw depth was 1.55 mm. On all measures, the accuracy of screw placement was within the predefined criteria. Conclusions Our cadaver study indicates that pedicle screw placement with the system is accurate and should be investigated in larger in vitro and in vivo studies.
引用
收藏
页码:S937 / S941
页数:5
相关论文
共 32 条
[1]   Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review [J].
Aoude, Ahmed A. ;
Fortin, Maryse ;
Figueiredo, Rainer ;
Jarzem, Peter ;
Ouellet, Jean ;
Weber, Michael H. .
EUROPEAN SPINE JOURNAL, 2015, 24 (05) :990-1004
[2]   The evolution of image-guided lumbosacral spine surgery [J].
Bourgeois, Austin C. ;
Faulkner, Austin R. ;
Pasciak, Alexander S. ;
Bradley, Yong C. .
ANNALS OF TRANSLATIONAL MEDICINE, 2015, 3 (05)
[3]   Quantitative analysis of misplaced pedicle screws in the thoracic spine: how much pullout strength is lost? Presented at the 2009 Joint Spine Section Meeting Laboratory investigation [J].
Brasiliense, Leonardo B. C. ;
Theodore, Nicholas ;
Lazaro, Bruno C. R. ;
Sayed, Zafar A. ;
Deniz, Fatih Ersay ;
Sonntag, Volker K. H. ;
Crawford, Neil R. .
JOURNAL OF NEUROSURGERY-SPINE, 2010, 12 (05) :503-508
[4]   The risks of aorta impingement from pedicle screw may increase due to aorta movement during posterior instrumentation in Lenke 5C curve: a computed tomography study [J].
Chen, Ling ;
Xu, Leilei ;
Qiu, Yong ;
Qiao, Jun ;
Wang, Fei ;
Liu, Zhen ;
Shi, Benglong ;
Qian, Bang-ping ;
Zhu, Zezhang .
EUROPEAN SPINE JOURNAL, 2015, 24 (07) :1481-1489
[5]   The use of pedicle-screw internal fixation for the operative treatment of spinal disorders [J].
Gaines, RW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (10) :1458-1476
[6]   Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques [J].
Gelalis, Ioannis D. ;
Paschos, Nikolaos K. ;
Pakos, Emilios E. ;
Politis, Angelos N. ;
Arnaoutoglou, Christina M. ;
Karageorgos, Athanasios C. ;
Ploumis, Avraam ;
Xenakis, Theodoros A. .
EUROPEAN SPINE JOURNAL, 2012, 21 (02) :247-255
[7]   ACCURACY OF PEDICULAR SCREW PLACEMENT INVIVO [J].
GERTZBEIN, SD ;
ROBBINS, SE .
SPINE, 1990, 15 (01) :11-14
[8]   Worldwide Survey on the Use of Navigation in Spine Surgery [J].
Haertl, Roger ;
Lam, Khai Sing ;
Wang, Jeffrey ;
Korge, Andreas ;
Kandziora, Frank ;
Audige, Laurent .
WORLD NEUROSURGERY, 2013, 79 (01) :162-172
[9]   Intraoperative spinal navigation [J].
Holly, LT ;
Foley, KT .
SPINE, 2003, 28 (15) :S54-S61
[10]   Pedicle screw placement with a free hand technique in thoracolumbar spine: Is it safe? [J].
Karapinar, Levent ;
Erel, Nuri ;
Ozturk, Hasan ;
Altay, Taskin ;
Kaya, Ahmet .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2008, 21 (01) :63-67