Image-based view-angle independent cardiorespiratory motion gating and coronary sinus catheter tracking for x-ray-guided cardiac electrophysiology procedures

被引:5
作者
Panayiotou, Maria [1 ]
Rhode, Kawal S. [1 ]
King, Andrew P. [1 ]
Ma, Yingliang [1 ]
Cooklin, Michael [2 ]
O'Neill, Mark [2 ]
Gill, Jaswinder [2 ]
Rinaldi, C. A. [2 ]
Housden, R. James [1 ]
机构
[1] Kings Coll London, Div Imaging Sci & Biomed Engn, London SE1 7EH, England
[2] Guys & St Thomas Hosp NHS Fdn Trust, Dept Cardiol, London, England
基金
英国工程与自然科学研究理事会;
关键词
cardiorespiratory motion gating; principal component analysis; cardiac electrophysiology; 3D rotational angiography; x-ray fluoroscopy; RESPIRATORY MOTION; INTERVENTIONAL NAVIGATION; ORGAN MOTION; ABLATION; COMPENSATION; REGISTRATION; SYSTEM; MODEL; MR;
D O I
10.1088/0031-9155/60/20/8087
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Determination of the cardiorespiratory phase of the heart has numerous applications during cardiac imaging. In this article we propose a novel view-angle independent near-real time cardiorespiratory motion gating and coronary sinus (CS) catheter tracking technique for x-ray fluoroscopy images that are used to guide cardiac electrophysiology procedures. The method is based on learning CS catheter motion using principal component analysis and then applying the derived motion model to unseen images taken at arbitrary projections, using the epipolar constraint. This method is also able to track the CS catheter throughout the x-ray images in any arbitrary subsequent view. We also demonstrate the clinical application of our model on rotational angiography sequences. We validated our technique in normal and very low dose phantom and clinical datasets. For the normal dose clinical images we established average systole, end-expiration and end-inspiration gating success rates of 100%, 85.7%, and 92.3%, respectively. For very low dose applications, the technique was able to track the CS catheter with median errors not exceeding 1 mm for all tracked electrodes. Average gating success rates of 80.3%, 71.4%, and 69.2% were established for the application of the technique on clinical datasets, even with a dose reduction of more than 10 times. In rotational sequences at normal dose, CS tracking median errors were within 1.2 mm for all electrodes, and the gating success rate was 100%, for view angles from RAO 90 degrees to LAO 90 degrees. This view-angle independent technique can extract clinically useful cardiorespiratory motion information using x-ray doses significantly lower than those currently used in clinical practice.
引用
收藏
页码:8087 / 8108
页数:22
相关论文
共 45 条
  • [31] A statistical method for retrospective cardiac and respiratory motion gating of interventional cardiac x-ray images
    Panayiotou, Maria
    King, Andrew P.
    Housden, R. James
    Ma, YingLiang
    Cooklin, Michael
    O'Neill, Mark
    Gill, Jaswinder
    Rinaldi, C. Aldo
    Rhode, Kawal S.
    [J]. MEDICAL PHYSICS, 2014, 41 (07)
  • [32] Penney G P, 2001, THESIS
  • [33] Modeling and Registration for Electrophysiology Procedures Based on Three-Dimensional Imaging
    Rhode K.
    Sermesant M.
    [J]. Current Cardiovascular Imaging Reports, 2011, 4 (2) : 116 - 126
  • [34] A system for real-time XMR guided cardiovascular intervention
    Rhode, KS
    Sermesant, M
    Brogan, D
    Hegde, S
    Hipwell, J
    Lambiase, P
    Rosenthal, E
    Bucknall, C
    Qureshi, SA
    Gill, JS
    Razavi, R
    Hill, DLG
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (11) : 1428 - 1440
  • [35] GEOMETRICAL CALIBRATION OF X-RAY-IMAGING CHAINS FOR 3-DIMENSIONAL RECONSTRUCTION
    ROUGEE, A
    PICARD, C
    PONCHUT, C
    TROUSSET, Y
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1993, 17 (4-5) : 295 - 300
  • [36] Model-Based Respiratory Motion Compensation for Image-Guided Cardiac Interventions
    Schneider, Matthias
    Sundar, Hari
    Liao, Rui
    Hornegger, Joachim
    Xu, Chenyang
    [J]. 2010 IEEE CONFERENCE ON COMPUTER VISION AND PATTERN RECOGNITION (CVPR), 2010, : 2948 - 2954
  • [37] Prospective motion correction of X-ray images for coronary interventions
    Shechter, G
    Shechter, B
    Resar, JR
    Beyar, R
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (04) : 441 - 450
  • [38] Respiratory motion of the heart from free breathing coronary angiograms
    Shechter, G
    Ozturk, C
    Resar, JR
    McVeigh, ER
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2004, 23 (08) : 1046 - 1056
  • [39] Registration of 3D computed tomographic images with interventional systems: Implications for catheter ablation of atrial fibrillation
    Sra, Jasbir
    Narayan, Girish
    Krum, David
    Akhtar, Masood
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2006, 16 (03) : 141 - 148
  • [40] Strat T.M., 1987, Readings In Computer Vision, P93, DOI DOI 10.1016/B978-0-08-051581-6.50017-9