A Lexicon for Wear of Metal-on-Metal Hip Prostheses

被引:16
|
作者
McKellop, Harry A. [1 ,2 ]
Hart, Alister [3 ,4 ]
Park, Sang-Hyun [1 ,2 ]
Hothi, Harry [3 ,4 ]
Campbell, Pat [1 ,2 ]
Skinner, John A. [3 ,4 ]
机构
[1] Univ Calif Los Angeles, Dept Orthopaed Surg, Los Angeles, CA 90095 USA
[2] Orthopaed Inst Children, Los Angeles, CA USA
[3] Univ Coll London, Stanmore, Middx, England
[4] Royal Natl Orthopaed Hosp, Inst Orthopaed & Musculoskeletal Sci, Stanmore HA7 4LP, Middx, England
关键词
wear; cobalt-chrome alloy; retrieval; corrosion; hip arthroplasty; IN-VIVO WEAR; RETRIEVAL ANALYSIS; COMPONENTS; BEARINGS;
D O I
10.1002/jor.22651
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Research on metal-on-metal (MoM) hip bearings has generated an extensive vocabulary to describe the wear processes and resultant surface damage. However, a lack of consistency and some redundancy exist in the current terminology. To facilitate the understanding of MoM tribology and to enhance communication of results among researchers and clinicians, we propose four categories of wear terminology: wear modes refer to the in vivo conditions under which the wear occurred; wear mechanisms refer to fundamental wear processes (adhesion, abrasion, fatigue, and tribochemical reactions); wear damage refers to the resultant changes in the morphology and/or composition of the surfaces; and wear features refer to the specific wear phenomena that are described in terms of the relevant modes, mechanisms, and damage. Clarifying examples are presented, but it is expected that terms will be added to the lexicon as new mechanisms and types of damage are identified. Corrosion refers to electrochemical processes that can remove or add material and thus also generate damage. Corrosion can act alone or may interact with mechanical wear. Examples of corrosion damage are also presented. However, an in-depth discussion of the many types of corrosion and their effects is beyond the scope of the present wear lexicon. (C) 2014 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.
引用
收藏
页码:1221 / 1233
页数:13
相关论文
共 50 条
  • [21] Biotribological study of multilayer coated metal-on-metal hip prostheses in a hip joint simulator
    Ortega-Saenz, J. A.
    Alvarez-Vera, M.
    Hernandez-Rodriguez, M. A. L.
    WEAR, 2013, 301 (1-2) : 234 - 242
  • [22] Which Factors Determine the Wear Rate of Large-Diameter Metal-on-Metal Hip Replacements?
    Hart, A. J.
    Muirhead-Allwood, S.
    Porter, M.
    Matthies, A.
    Ilo, K.
    Maggiore, P.
    Underwood, R.
    Cann, P.
    Cobb, J.
    Skinner, J. A.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (08) : 678 - 685
  • [23] Wear Simulation of Metal-on-Metal Hip Replacements With Frictional Contact
    Mattei, Lorenza
    Di Puccio, Francesca
    JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2013, 135 (02):
  • [24] What Is a "Normal" Wear Pattern for Metal-on-Metal Hip Bearings?
    Bowsher, J. G.
    Clarke, I. C.
    Williams, P. A.
    Donaldson, T. K.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2009, 91B (01) : 297 - 308
  • [25] Predictive wear modeling of the articulating metal-on-metal hip replacements
    Gao, Leiming
    Dowson, Duncan
    Hewson, Robert W.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2017, 105 (03) : 497 - 506
  • [26] Metal-on-metal hip replacements: Wear performance and cellular response to wear particles
    Campbell, P
    McKellop, H
    Alim, R
    Mirra, J
    Nutt, S
    Dorr, L
    Amstutz, HC
    COBALT-BASE ALLOYS FOR BIOMEDICAL APPLICATIONS, 1999, 1365 : 193 - 209
  • [27] Metal-on-metal hip prostheses: Correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream
    De Pasquale, Dalila
    Stea, Susanna
    Squarzoni, Stefano
    Bordini, Barbara
    Amabile, Marilina
    Catalani, Simona
    Apostoli, Pietro
    Toni, Aldo
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (03) : 469 - 475
  • [28] Factors Associated With Trunnionosis in the Metal-on-Metal Pinnacle Hip
    Hothi, Harry S.
    Eskelinen, Antti P.
    Berber, Reshid
    Lainiala, Olli S.
    Moilanen, Teemu P. S.
    Skinner, John A.
    Hart, Alister J.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (01) : 286 - 290
  • [29] Characterization of protein degradation in serum-based lubricants during simulation wear testing of metal-on-metal hip prostheses
    Maskiewicz, Victoria K.
    Williams, Paul A.
    Prates, Sarah J.
    Bowsher, John G.
    Clarke, Ian C.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2010, 94B (02) : 429 - 440
  • [30] Bioengineering reasons for the failure of metal-on-metal hip prostheses AN ENGINEER'S PERSPECTIVE
    Fisher, J.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (08): : 1001 - 1004