Evaluating five dedicated automatic devices for haemoglobinopathy diagnostics in multi-ethnic populations

被引:104
作者
Van Delft, P.
Lenters, E. [2 ]
Bakker-Verweij, M.
de Korte, M.
Baylan, U.
Harteveld, C. L.
Giordano, P. C. [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Human & Clin Genet, Haemoglobinopathies Lab,Ctr Human & Clin Genet, NL-2300 RC Leiden, Netherlands
[2] Isala Clin, Dept Clin Chem, Zwolle, Netherlands
关键词
HPLC; capillary electrophoresis; thalassaemia; sickle cell disease; diagnosis; A(2); VARIANTS; RETENTION; MUTATION; SAMPLES; SYSTEM;
D O I
10.1111/j.1751-553X.2009.01158.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
P>We have tested five haemoglobin (Hb) separation apparatuses, dedicated to haemoglobinopathy diagnostics. These are the four high performance liquid chromatography devices: VARIANT II (TM), HA 8160, G7, Ultra(2) and the Capillary Electrophoresis apparatus from Sebia. In the first place, we focussed on the capacity of all apparatuses to detect the most common structural variants relevant for public health, these being HbS, HbC, HbE, HbD-Punjab and HbO-Arab. We then compared how the high HbA(2) beta-thalassaemia carriers were identified. All apparatuses were able to identify carriers of these traits with the expected sensitivity and specificity. With the primary goal of a high degree of conformity in basic diagnostics of haemoglobinopathies, we present the interpretation and the significance of the results on all apparatuses, and we comment on the unavoidable problems and solutions.
引用
收藏
页码:484 / 495
页数:12
相关论文
共 20 条
[1]  
Aulesa Carlos, 1995, Sangre (Saragossa), V40, P369
[2]  
Bouva MJ, 2006, HAEMATOLOGICA, V91, P129
[3]  
DALAND GA, 1948, J LAB CLIN MED, V33, P1082
[4]  
Giordano PC, 2000, ANN CLIN LAB SCI, V30, P430
[5]  
GIORDANO PC, 2006, NED TIJDSCHR KLIN CH, V31, P301
[6]  
Giordano PC, 2006, HAEMATOLOGICA, V91, P153
[7]   Inaccuracy of high-performance liquid chromatography estimation of haemoglobin F in the presence of increased haemoglobin A1c [J].
Grey, V. ;
Wilkinson, M. ;
Phelan, L. ;
Hughes, C. ;
Bain, B. J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2007, 29 (01) :42-44
[8]   α-thalassemia phenotype induced by the new IVS-II-2 (T→A) splice donor site mutation on the α2-globin gene [J].
Harteveld, CL ;
Jebbink, MCW ;
van der Lely, N ;
van Delft, P ;
Akkermans, N ;
Arkesteyn, S ;
Giordano, PC .
HEMOGLOBIN, 2006, 30 (01) :3-7
[9]   Hb St. Jozef, a val→leu N-terminal mutation leading to retention of the methionine, and partial acetylation found in the globin gene in cis with a -α3.7 thalassemia deletion [J].
Harteveld, Cornelis L. ;
Versteegh, Florens G. A. ;
van Leer, Eduard H. G. ;
Starreveld, Jaap S. ;
Kok, Peter J. M. J. ;
van Rooijen-Nijdam, Irene ;
van Delft, Peter ;
Zanella-Cleon, Isabelle ;
Becchi, Michel ;
Wajcman, Henri ;
Giordano, Piero C. .
HEMOGLOBIN, 2007, 31 (03) :313-323
[10]   Identification of hemoglobin variants by HPLC [J].
Hoyer, JD ;
Scheidt, RM .
CLINICAL CHEMISTRY, 2005, 51 (07) :1303-1304