Assessing the Placement of the Cochlear Slim Perimodiolar Electrode Array by Trans Impedance Matrix Analysis: A Temporal Bone Study

被引:3
作者
Ramos de Miguel, Angel [1 ,2 ]
Riol Sancho, Diego [3 ]
Carlos Falcon-Gonzalez, Juan [1 ,2 ]
Pavone, Joana [1 ,2 ]
Rodriguez Herrera, Leandro [1 ,2 ]
Borkoski Barreiro, Silvia [1 ,2 ]
Falcon Benitez, Nadia [4 ]
Ramos Macias, Angel [1 ,2 ,4 ]
机构
[1] Complejo Hosp Univ Insular Materno Infantil Gran, Dept Otolaryngol Head & Neck Surg, Las Palmas Gran Canaria 35016, Spain
[2] Las Palmas Univ ULPGC, Univ Inst Intelligent Syst & Numer Applicat Engn, Dept Otolaryngol Psychoacoust & Equilibrium Lab, Las Palmas Gran Canaria 35016, Spain
[3] Complejo Hosp Univ Insular Materno Infantil Gran, Dept Radiol Head Neck Surg, Las Palmas Gran Canaria 35016, Spain
[4] Univ Las Palmas de Gran Canaria ULPGC, Dept Clin Sci, Las Palmas Gran Canaria 35001, Spain
关键词
cochlear implant; trans impedance; temporal bone; perimodiolar electrode array; hearing loss; fold-over of the electrode; MODIOLAR ELECTRODE; INSERTION DEPTH; FOLD-OVER; LOCALIZATION; STIMULATION;
D O I
10.3390/jcm11143930
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
New cochlear implant (CI) electrode arrays provide softer insertion dynamics; however, due to their high flexibility, the possibilities of fold-overs or intraoperative displacements must be taken into account. The position of each individual electrode can only be determined by using high-resolution computed tomography or cone-beam CT. The trans-impedance matrix test (TIM) is an electrophysiological method based on electric field imaging that can provide images of electrode position and electrode folding. Objective: In this experimental research, we evaluated the result of TIM as a method of monitoring cochlear insertion for a precurved slim modiolar electrode array in fresh human temporal bones by analyzing the transimpedance matrix patterns and their correlation with electrode position using high-resolution computed tomography. Material and Methods: Sixteen slim modiolar electrode arrays were inserted into eight fresh Human Temporal Bones. Eight electrodes were inserted according to the correct methodology of insertion, and eight were intentionally folded over. After all insertions, a trans-impedance matrix analysis and a Cone Beam CT (CBCT) were performed in each temporal bone. Results: If we correlated the TIM patterns with the radiological electrode position, we observed that better electrode intracochlear positions indicated more "homogeneous" TIM patterns (intracochlear voltage dropped monotonically as the distance between stimulation and recording contact increased, both toward the apex and toward the base). A correlation where fold-over was detected in the TIM results was found in all eight temporal bone radiological findings. Conclusions: Trans-Impedance Matrix patterns were correlated with the radiological CI electrode position. When a tip fold-over appeared, a matrix with a secondary ridge in addition to the primary ridge was observed in all cases. TIM can be an effective method in the control of electrode positioning.
引用
收藏
页数:12
相关论文
共 32 条
  • [1] Preservation of basal inner ear structures in cochlear implantation
    Adunka, O
    Gstoettner, W
    Hambek, M
    Unkelbach, MH
    Radeloff, A
    Kiefer, J
    [J]. ORL-JOURNAL FOR OTO-RHINO-LARYNGOLOGY AND ITS RELATED SPECIALTIES, 2004, 66 (06): : 306 - 312
  • [2] Intraoperative Impedance-Based Estimation of Cochlear Implant Electrode Array Insertion Depth
    Aebischer, Philipp
    Meyer, Stefan
    Caversaccio, Marco
    Wimmer, Wilhelm
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2021, 68 (02) : 545 - 555
  • [3] In Vivo Real-time Remote Cochlear Implant Capacitive Impedance Measurements: A Glimpse Into the Implanted Inner Ear
    Alberto Di Lella, Federico
    De Marco, Diego
    Fernandez, Florencia
    Parreno, Matias
    Mario Boccio, Carlos
    [J]. OTOLOGY & NEUROTOLOGY, 2019, 40 : S18 - S22
  • [4] Clinical investigation of the Nucleus Slim Modiolar Electrode
    Aschendorff, Antje
    Briggs, Robert
    Brademann, Goetz
    Helbig, Silke
    Hornung, Joachim
    Lenarz, Thomas
    Marx, Mathieu
    Ramos, Angel
    Stoever, Timo
    Escude, Bernard
    James, Chris J.
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 2017, 22 (03) : 169 - 179
  • [5] Briggs Robert J S, 2006, Audiol Neurootol, V11 Suppl 1, P42, DOI 10.1159/000095613
  • [6] Factors influencing the efficacy of round window dexamethasone protection of residual hearing post-cochlear implant surgery
    Chang, Andrew
    Eastwood, Hayden
    Sly, David
    James, David
    Richardson, Rachael
    O'Leary, Stephen
    [J]. HEARING RESEARCH, 2009, 255 (1-2) : 67 - 72
  • [7] Cochlear implantation with the nucleus slim modiolar electrode (CI532): a preliminary experience
    Cuda, Domenico
    Murri, Alessandra
    [J]. EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2017, 274 (12) : 4141 - 4148
  • [8] Detection of Extracochlear Electrodes in Cochlear Implants with Electric Field Imaging/Transimpedance Measurements: A Human Cadaver Study
    de Rijk, Simone R.
    Tam, Yu C.
    Carlyon, Robert P.
    Bance, Manohar L.
    [J]. EAR AND HEARING, 2020, 41 (05) : 1196 - 1207
  • [9] Detection of Translocation of Cochlear Implant Electrode Arrays by Intracochlear Impedance Measurements
    Dong, Yu
    Briaire, Jeroen J.
    Siebrecht, Michael
    Stronks, H. Christiaan
    Frijns, Johan H. M.
    [J]. EAR AND HEARING, 2021, 42 (05) : 1397 - 1404
  • [10] Variational Anatomy of the Human Cochlea: Implications for Cochlear Implantation
    Erixon, Elsa
    Hogstorp, Herman
    Wadin, Karin
    Rask-Andersen, Helge
    [J]. OTOLOGY & NEUROTOLOGY, 2009, 30 (01) : 14 - 22