Evaluation of the imaging properties of an amorphous selenium-based flat panel detector for digital fluoroscopy

被引:77
作者
Hunt, DC
Tousignant, O
Rowlands, JA
机构
[1] Sunnybrook & Womens Coll Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
[2] ANRAD Corp, St Laurent, PQ H4R 2P1, Canada
关键词
flat-panel detector; amorphous selenium; digital fluoroscopy; MTF; NPS; DQE; anti-aliasing; presampling blur; fill factor;
D O I
10.1118/1.1707755
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The imaging performance of an amorphous selenium (alpha-Se) flat-panel detector for digital fluoroscopy was experimentally evaluated using the spatial frequency dependent modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). These parameters were investigated at beam qualities and exposures within the range typical of gastrointestinal fluoroscopic imaging (similar to0.1-10 muR, 75 kV). The investigation does not take into consideration the detector cover, which in clinical use will lower the DQE measured here by its percent attenuation. The MTF was found to be less than the expected aperture response and the NPS was not white which together indicate presampling blurring. The cause of this blurring was attributed to charge trapping at the interface between two different layers of the a-Se. The effect on the DQE was also consistent with presampling blur, which reduces the aliasing in the NPS and thereby reduces the spatial frequency dependence of the DQE. (The DQE was independent of spatial frequency from 0.12 to 0.73 mm(-1) due to antialiasing of the NPS.) Moreover, the first zero of the measured MTF and the aperture response appeared at the same spatial frequency (6.66 mm(-1) for a pixel of 150 mum). Hence, the geometric fill factor (77%) was increased to an effective fill factor of 99 +/- 1%. A large scale (similar to32 pixels) correlation in the noise due to the configuration of the readout electronics caused increased noise power in the gate line NPS at low spatial frequency (<0.1 mm(-1)). The DQE (f=0) was exposure independent over a large range of exposures but became exposure dependent at low exposures due to the electronic noise. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:1166 / 1175
页数:10
相关论文
共 26 条
[21]   An experimental comparison of detector performance for direct and indirect digital radiography systems [J].
Samei, E ;
Flynn, MJ .
MEDICAL PHYSICS, 2003, 30 (04) :608-622
[22]  
SCHAFFERT RM, 1975, ELECTROPHOTOGRAPHY, P505
[23]   Comparison of PbI2 and HgI2 for direct detection active matrix x-ray image sensors [J].
Street, RA ;
Ready, SE ;
Van Schuylenbergh, K ;
Ho, J ;
Boyce, JB ;
Nylen, P ;
Shah, K ;
Melekhov, L ;
Hermon, H .
JOURNAL OF APPLIED PHYSICS, 2002, 91 (05) :3345-3355
[24]   Progress report on the performance of real time selenium flat panel detectors for direct X-ray Imaging [J].
Tousignant, O ;
Choquette, M ;
Demers, Y ;
Laperrière, L ;
Leboeuf, J ;
Honda, M ;
Nishiki, M ;
Takahashi, A ;
Tsukamoto, A .
MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 :503-510
[25]   X-RAY-IMAGING USING AMORPHOUS SELENIUM - FEASIBILITY OF A FLAT-PANEL SELF-SCANNED DETECTOR FOR DIGITAL RADIOLOGY [J].
ZHAO, W ;
ROWLANDS, JA .
MEDICAL PHYSICS, 1995, 22 (10) :1595-1604
[26]   Imaging performance of amorphous selenium based flat-panel detectors for digital mammography: Characterization of a small area prototype detector [J].
Zhao, W ;
Ji, WG ;
Debrie, A ;
Rowlands, JA .
MEDICAL PHYSICS, 2003, 30 (02) :254-263