Evaluation of the imaging properties of an amorphous selenium-based flat panel detector for digital fluoroscopy

被引:77
作者
Hunt, DC
Tousignant, O
Rowlands, JA
机构
[1] Sunnybrook & Womens Coll Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
[2] ANRAD Corp, St Laurent, PQ H4R 2P1, Canada
关键词
flat-panel detector; amorphous selenium; digital fluoroscopy; MTF; NPS; DQE; anti-aliasing; presampling blur; fill factor;
D O I
10.1118/1.1707755
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The imaging performance of an amorphous selenium (alpha-Se) flat-panel detector for digital fluoroscopy was experimentally evaluated using the spatial frequency dependent modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). These parameters were investigated at beam qualities and exposures within the range typical of gastrointestinal fluoroscopic imaging (similar to0.1-10 muR, 75 kV). The investigation does not take into consideration the detector cover, which in clinical use will lower the DQE measured here by its percent attenuation. The MTF was found to be less than the expected aperture response and the NPS was not white which together indicate presampling blurring. The cause of this blurring was attributed to charge trapping at the interface between two different layers of the a-Se. The effect on the DQE was also consistent with presampling blur, which reduces the aliasing in the NPS and thereby reduces the spatial frequency dependence of the DQE. (The DQE was independent of spatial frequency from 0.12 to 0.73 mm(-1) due to antialiasing of the NPS.) Moreover, the first zero of the measured MTF and the aperture response appeared at the same spatial frequency (6.66 mm(-1) for a pixel of 150 mum). Hence, the geometric fill factor (77%) was increased to an effective fill factor of 99 +/- 1%. A large scale (similar to32 pixels) correlation in the noise due to the configuration of the readout electronics caused increased noise power in the gate line NPS at low spatial frequency (<0.1 mm(-1)). The DQE (f=0) was exposure independent over a large range of exposures but became exposure dependent at low exposures due to the electronic noise. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:1166 / 1175
页数:10
相关论文
共 26 条
[1]   Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications [J].
Antonuk, LE ;
Jee, KW ;
El-Mohri, Y ;
Maolinbay, M ;
Nassif, S ;
Rong, X ;
Zhao, Q ;
Siewerdsen, JH ;
Street, RA ;
Shah, KS .
MEDICAL PHYSICS, 2000, 27 (02) :289-306
[2]   X-ray imaging using amorphous selenium: Determination of Swank factor by pulse height spectroscopy [J].
Blevis, IM ;
Hunt, DC ;
Rowlands, JA .
MEDICAL PHYSICS, 1998, 25 (05) :638-641
[3]   Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV [J].
Boone, JM ;
Seibert, JA .
MEDICAL PHYSICS, 1997, 24 (11) :1661-1670
[4]   Principles of digital radiography with large-area, electronically readable detectors: A review of the basics [J].
Chotas, HG ;
Dobbins, JT ;
Ravin, CE .
RADIOLOGY, 1999, 210 (03) :595-599
[5]   Threshold contrast detail detectability measurement of the fluoroscopic image quality of a dynamic solid-state digital x-ray image detector [J].
Davies, AG ;
Cowen, AR ;
Kengyelics, SM ;
Bury, RF ;
Bruijns, TJ .
MEDICAL PHYSICS, 2001, 28 (01) :11-15
[6]   EFFECTS OF UNDERSAMPLING ON THE PROPER INTERPRETATION OF MODULATION TRANSFER-FUNCTION, NOISE POWER SPECTRA, AND NOISE EQUIVALENT QUANTA OF DIGITAL IMAGING-SYSTEMS [J].
DOBBINS, JT .
MEDICAL PHYSICS, 1995, 22 (02) :171-181
[7]   Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography [J].
Floyd, CE ;
Warp, RJ ;
Dobbins, JT ;
Chotas, HG ;
Baydush, AH ;
Vargas-Voracek, R ;
Ravin, CE .
RADIOLOGY, 2001, 218 (03) :683-688
[8]   Performance of a 41X41 cm2 amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications [J].
Granfors, PR ;
Aufrichtig, R ;
Possin, GE ;
Giambattista, BW ;
Huang, ZS ;
Liu, JQ ;
Ma, B .
MEDICAL PHYSICS, 2003, 30 (10) :2715-2726
[9]   Performance of a 41X41-cm2 amorphous silicon flat panel x-ray detector for radiographic imaging applications [J].
Granfors, PR ;
Aufrichtig, R .
MEDICAL PHYSICS, 2000, 27 (06) :1324-1331
[10]   Evaluation of an algorithm for the assessment of the MTF using an edge method [J].
Greer, PB ;
van Doorn, T .
MEDICAL PHYSICS, 2000, 27 (09) :2048-2059